Fig. 1. Schematic representation of the situation in which a force applied to the plunger contorts the rubber piston when a prefilled syringe is used with a syringe pump. The gap (arrow ) breaks the seal between the barrel of the syringe and the rubber piston, resulting in drug leakage as air entered through the gap. This plunger/piston misassembly caused an incident during drug administration.

Fig. 1. Schematic representation of the situation in which a force applied to the plunger contorts the rubber piston when a prefilled syringe is used with a syringe pump. The gap (arrow ) breaks the seal between the barrel of the syringe and the rubber piston, resulting in drug leakage as air entered through the gap. This plunger/piston misassembly caused an incident during drug administration.