Background

Electromyographic activity has previously been reported to elevate the Bispectral Index (BIS) in patients not receiving neuromuscular blockade while under sedation in the intensive care unit. This study aimed to investigate the magnitude of the decrease of BIS following administration of muscle relaxant in sedated intensive care unit patients.

Methods

The authors prospectively investigated 45 patients who were continuously sedated with midazolam and sufentanil to achieve a Sedation-Agitation Scale value equal to 1 and who required administration of muscle relaxant. BIS (BIS version 2.10), electromyography, and acceleromyography at the adductor pollicis muscle were recorded simultaneously before and after neuromuscular blockade. Sixteen of these 45 patients were also studied simultaneously with the new BIS XP.

Results

After administration of a muscle relaxant, BIS (67 +/- 19 vs. 43 +/- 10, P < 0.001) and electromyographic activity (37 +/- 9 vs. 27 +/- 3 dB, P < 0.001) significantly decreased. Multiple regression analysis showed that the decrease of BIS following administration of myorelaxant was significantly correlated to BIS and electromyographic baseline values. Using standard BIS range guidelines, the number of patients under light or deep sedation versus general anesthesia or deep hypnotic state was markedly overestimated before administration of myorelaxant (53 vs. 2%, P < 0.001).

Conclusions

The BIS in sedated intensive care unit patients may be lower with paralysis for an equivalent degree of sedation because of high muscular activity. The magnitude of BIS overestimation is significantly correlated to both BIS and electromyographic activity before neuromuscular blockade. The authors conclude that clinicians who determine the amount of sedation in intensive care unit patients only from BIS monitoring may expose them to unnecessary oversedation.

FOR several years, the Bispectral Index (BIS), a variable derived from the electroencephalogram, has been reported as a quantifiable measure of the depth of sedation and degree of awareness during and after general anesthesia 1,2and of the depth of sedation in the intensive care unit (ICU). 3–10Good correlations have been reported between the BIS and several sedation scores in sedated ICU patients. 3–6,8,9,11Elsewhere, electromyographic activity, which had previously been reported to falsely elevate BIS values in anesthetized patients without neuromuscular blockade, 12may also elevate BIS values in sedated ICU patients not receiving neuromuscular blockade. 5,8,9,13Simmons et al.  4reported that patients with neuromuscular blockade had a higher suppression ratio than patients without neuromuscular blockade. Against this, Greif et al.  14recently reported that the BIS level during propofol anesthesia was unchanged by mivacurium administration. However, this study was performed in volunteers, who were highly sedated only with propofol, at a target effect site concentration of 3.8 ± 0.4 μg/ml, which represents clinically irrelevant conditions.

High electromyographic activity interference, which may elevate the BIS, is acknowledged as one of the most important limitations and pitfalls to BIS monitoring in ICU patients. 5,11,13,15Nevertheless, to the best of our knowledge, the range of potential overestimation of BIS in sedated ICU patients without neuromuscular blockade remained unknown. Therefore, the aim of this study was to investigate the magnitude of the decrease in BIS value following administration of muscle relaxant in sedated ICU patients.

This prospective study was conducted in accordance with the Helsinki declaration, and the protocol was approved by our local ethical committee (Comité Consultatif pour la Protection des Personnes se prêtant aux Recherches Biomédicales, Pitié-Salpêtrière, Paris). Care of the patients conformed to standard procedures currently used in our ICU, including standard equipment monitoring with continuous electrocardiography, invasive arterial blood pressure monitoring, pulse oximetry and capnography, and the additional use of noninvasive BIS monitoring. Patients were eligible for this study if they were mechanically ventilated and continuously sedated with midazolam and sufentanil to achieve a Sedation-Agitation Scale value equal to 1 (table 1). 16We prospectively included patients who required administration of muscle relaxant either for adaptation to mechanical ventilation or to facilitate invasive clinical procedures such as transesophageal echocardiography or bronchoscopy. Exclusion criteria were hemodynamic instability, administration of sedative drugs other than midazolam and sufentanil, change in midazolam or sufentanil administration regimen during the study period, and recent (<3 h) administration of muscle relaxant before the study period. Finally, because brain injury and hypothermia may decrease BIS values, we also excluded patients with neurologic injury, head trauma, or hypothermia (tympanic temperature < 36.0°C). 9,15,17,18 

Table 1. Riker Sedation-Agitation Scale

Adapted from Riker et al.  16 

Table 1. Riker Sedation-Agitation Scale
Table 1. Riker Sedation-Agitation Scale

The BIS was measured with an Aspect A-2000 monitor (Aspect Medical Systems, Newton, MA; software version 2.10). In every patient, the skin was carefully cleaned before positioning the BIS sensor (ZipPrep® three-electrode sensor; Aspect Medical Systems) on the forehead as specified by the manufacturer. Before starting recording, we verified that electrode impedance was below 5 kΩ. The BIS, electromyographic activity, total power of the signal (TOTPOW), and signal quality index (SQI) were continuously recorded and stored on a computer for off-line analysis. Because it has been shown that many devices such as forced-air warming systems can interfere with BIS monitoring, all of those artifacts were systematically noted and removed from computer file recordings during off-line analysis. 19,20Similarly, physical examination and care of the patient, such as tracheal suction, 7were carefully noted because they could markedly interfere with BIS value either through neurologic stimulation of the patient or through artifact production.

Elsewhere, because a new version of the BIS® monitor had been recently commercialized (BIS® XP, software version 3.12, four-electrode sensor; Aspect Medical Systems), which should better discriminate artifacts such as electromyographic activity, we decided to investigate some of the patients simultaneously with the old and the new BIS® monitors, as previously described. 21For these patients, the two BIS® sensors were simultaneously placed on opposite sides of the head.

The effect of neuromuscular blocking agents on muscle activity was monitored through stimulation of the ulnar nerve and acceleromyography at the adductor pollicis muscle. 22,23The stimulation surface electrodes were placed over the ulnar nerve on the volar side of the wrist. The acceleration transducer was placed at the distal part of the thumb, with its largest flat side against the thumb. The train-of-four (TOF) sequence (2 Hz) was recorded after supramaximal stimuli (200-μs duration) repeated every 15 s (TOF-Watch SX Monitor 1.2; Organon, Dublin, Ireland). The first response in the TOF sequence (T1) was used as the baseline value to which all subsequent T1 responses were compared. A preliminary study (data not shown) reported that TOF watch monitoring at the adductor pollicis did not interfere with BIS monitoring.

Experimental Protocol

Before administration of myorelaxant, BIS, electromyographic activity, TOTPOW, and SQI were recorded for a 20-min period without clinical stimulation to the patient (baseline period) to determine the value of these variables in the absence of muscle relaxant. Immediately after this 20-min recording, control twitch height of the adductor pollicis was calibrated to set the T1 100% baseline level. Then, intravenous administration of myorelaxant (0.5 mg/kg atracurium) was performed. As soon as T1 decreased below 5% of baseline value, BIS, electromyographic activity, TOTPOW, and SQI were determined from a second 20-min recording period without clinical stimulation to the patient (myorelaxation period).

For patients who required myorelaxation for adaptation to mechanical ventilation, the study period was then finished. For the other patients who transitorily required myorelaxation to facilitate transesophageal echocardiography or bronchoscopy, the following procedure was subsequently performed: After completion of this investigation, as soon as T1 recovered 90% of baseline value, BIS, electromyographic activity, TOTPOW, and SQI were determined from a last, third 20-min recording period without clinical stimulation to the patient (recovery period). We also studied the shape of both BIS and electromyographic activity recovery curves and arbitrarily defined two pathways of recovery: (1) smooth recovery when the shapes were smooth and regular; and (2) stepped recovery when there were sudden increases in BIS (>20 units) and/or electromyographic (>5 dB) values occurring in less than 1 min.

For the baseline and myorelaxation periods, patients were conventionally classified in four levels of sedation, according to standard BIS range guidelines 11: light sedation (BIS ≥ 85), deep sedation (65 ≥ BIS < 85), general anesthesia (40 ≥ BIS < 65), and deep hypnotic state (BIS < 40). Elsewhere, as previously described, we defined a threshold of 42 dB, above which electromyographic activity is considered to artifactually increase BIS values. 9,13 

Statistical Analysis

Data are expressed as mean ± SD. Means ± SDs of BIS, electromyographic activity, TOTPOW, and SQI were calculated from each 20-min recording period. Comparison of two means was performed using the unpaired and paired Student t  tests when appropriate. Comparison of several means was performed using repeated-measures analysis of variance and Newman-Keuls test. Correlation between two variables was performed using the least squares method. Multiple regression analysis was performed to research predictor baseline variables of the overestimation of BIS. All P  values were two-tailed, and a P  value of less than 0.05 was required to reject the null hypothesis. All statistical analysis were run on a computer using NCSS 6.0 software (Statistical Solutions Ltd., Cork, Ireland).

Forty-five patients, aged 48 ± 22 yr (35 men, 10 women; weight, 71 ± 9 kg), were investigated prospectively. All of them required sedation for more than 24 h in the ICU for multiple trauma (n = 18) or postoperative care after major vascular (n = 13), orthopedic (n = 6), abdominal (n = 4), or maxillofacial (n = 4) surgery. Fifteen of these patients required long-period myorelaxation for adaptation to mechanical ventilation, whereas 30 patients required transitory administration of myorelaxant for facilitation of transesophageal echocardiography or bronchoscopy (transitory group).

No change in midazolam (6.5 ± 4.0 mg/h) or sufentanil (25 ± 13 μg/h) administration regimen occurred in any of the 45 patients during the study periods. For the 45 patients, hemodynamic variables remained unchanged between the baseline and myorelaxation study periods (table 2). For the 30 patients of the transitory group, hemodynamic variables remained unchanged among the baseline, myorelaxation, and recovery study periods (table 3).

Table 2. Hemodynamics and Bispectral Monitoring Variables in the 45 Studied Patients

Data are mean ± SD.

*P < 0.05 versus  baseline values.

Baseline = study period (20 min) before administration of myorelaxant; myorelaxation = study period (20 min) after administration of myorelaxant; Spo2= pulse oximetry; BIS = bispectral index; TOTPOW = total power of the signal; EMG = electromyographic activity; SQI = signal quality index.

Table 2. Hemodynamics and Bispectral Monitoring Variables in the 45 Studied Patients
Table 2. Hemodynamics and Bispectral Monitoring Variables in the 45 Studied Patients

Table 3. Hemodynamics and Bispectral Monitoring Variables in the 30 Patients in the Transitory Group

Data are mean ± SD. P  values refer to between-period differences (analysis of variance).

*P < 0.05 versus  baseline values.

P < 0.05 versus  recovery values.

Transitory group = patients who required transitory administration of myorelaxant for facilitation of invasive clinical procedures such as transesophageal echocardiography or bronchoscopy; baseline = study period (20 min) before administration of myorelaxant; myorelaxation = study period (20 min) after administration of myorelaxant; recovery = study period (20 min) after recovery of muscular activity; NS = not significant; Spo2= pulse oximetry; BIS = bispectral index; TOTPOW = total power of the signal; EMG = electromyographic activity; SQI = signal quality index.

Table 3. Hemodynamics and Bispectral Monitoring Variables in the 30 Patients in the Transitory Group
Table 3. Hemodynamics and Bispectral Monitoring Variables in the 30 Patients in the Transitory Group

After administration of myorelaxant, BIS (67 ± 19 vs.  43 ± 10, P < 0.001) and electromyographic activity (37 ± 9 vs.  27 ± 3 dB, P < 0.001) significantly decreased, whereas SQI significantly increased (90 ± 8 vs.  94 ± 7%, P < 0.001) and TOTPOW was unchanged (53.1 ± 3.2 vs.  52.7 ± 2.9 μV2, NS ). Figure 1shows two representative examples of the effect of administration of neuromuscular blocker on BIS and electromyographic activity, respectively, in one patient who required long-period myorelaxation for adaptation to mechanical ventilation and in another patient who required transitory myorelaxation for facilitation of bronchoscopy.

Fig. 1. (A ) Bispectral Index (BIS), electromyographic activity (EMG), total power of the signal (TOTPOW), and signal quality index (SQI) trend recordings in a patient who required long-period myorelaxation for adaptation to mechanical ventilation. Administration of myorelaxant induced a significant decrease in both BIS and electromyographic activity. The regular peaks in electromyographic activity and SQI signal activity are linked to regular resets of the BIS® monitor (automatic ground verification). (B ) BIS, electromyographic activity, and SQI trend recordings in a patient who required transitory myorelaxation for facilitation of bronchoscopy. Administration of myorelaxant induced a significant decrease in both BIS and electromyographic activity, which began to recover while the effect of the myorelaxant lessened. (TOTPOW trend recording is not plotted to avoid superimposition of curves.)

Fig. 1. (A ) Bispectral Index (BIS), electromyographic activity (EMG), total power of the signal (TOTPOW), and signal quality index (SQI) trend recordings in a patient who required long-period myorelaxation for adaptation to mechanical ventilation. Administration of myorelaxant induced a significant decrease in both BIS and electromyographic activity. The regular peaks in electromyographic activity and SQI signal activity are linked to regular resets of the BIS® monitor (automatic ground verification). (B ) BIS, electromyographic activity, and SQI trend recordings in a patient who required transitory myorelaxation for facilitation of bronchoscopy. Administration of myorelaxant induced a significant decrease in both BIS and electromyographic activity, which began to recover while the effect of the myorelaxant lessened. (TOTPOW trend recording is not plotted to avoid superimposition of curves.)

Close modal

Nevertheless, the effect of administration of neuromuscular blocker on BIS and electromyographic activity was highly variable among the 45 patients studied (fig. 2). In 13 patients, there was no change in BIS range assessment, according to the four standard BIS range guidelines; 19 patients showed a decrease of one step in BIS range assessment, 11 patients showed a decrease of two steps, and 2 patients showed a decrease of three steps (fig. 2, A ). These findings imply that BIS range assessment was overestimated in 32 of the 45 studied patients (71%). On the other hand, before administration of myorelaxant, only 10 of the 45 patients (22%) showed electromyographic activity greater than the threshold of 42 dB, above which electromyographic activity is considered to artifactually increase BIS values (fig. 2, B ). Figure 3presents trend recordings of two patients who respectively showed a nonsignificant change and a major decrease in both BIS and electromyographic activity following myorelaxant administration.

Fig. 2. Effect of myorelaxant administration on Bispectral Index (BIS;A ) and electromyographic activity (EMG;B ) (n = 45 patients for baseline and myorelaxation periods, n = 30 patients for recovery period). (A ) Dashed lines represent the conventional limits of the four levels of sedation, according to standard BIS range guidelines 11: light sedation (BIS ≥ 85), deep sedation (65 ≥ BIS < 85), general anesthesia (40 ≥ BIS < 65), and deep hypnotic state (BIS < 40). After administration of myorelaxant, 13 patients did not change in terms of BIS range assessment, 19 patients decreased by one step of BIS range assessment, 11 patients decreased by two steps, and 2 patients decreased by three steps. (B ) Dashed line represents the threshold of 42 dB, above which electromyographic activity is considered to artifactually increase BIS values. 9,13Before administration of myorelaxant, only 10 patients were above this threshold.

Fig. 2. Effect of myorelaxant administration on Bispectral Index (BIS;A ) and electromyographic activity (EMG;B ) (n = 45 patients for baseline and myorelaxation periods, n = 30 patients for recovery period). (A ) Dashed lines represent the conventional limits of the four levels of sedation, according to standard BIS range guidelines 11: light sedation (BIS ≥ 85), deep sedation (65 ≥ BIS < 85), general anesthesia (40 ≥ BIS < 65), and deep hypnotic state (BIS < 40). After administration of myorelaxant, 13 patients did not change in terms of BIS range assessment, 19 patients decreased by one step of BIS range assessment, 11 patients decreased by two steps, and 2 patients decreased by three steps. (B ) Dashed line represents the threshold of 42 dB, above which electromyographic activity is considered to artifactually increase BIS values. 9,13Before administration of myorelaxant, only 10 patients were above this threshold.

Close modal

Fig. 3. (A ) Bispectral Index (BIS), electromyographic activity (EMG), and twitch height of the first response of train-of-four sequence recorded at the adductor pollicis muscle (twitch) trend recordings in a patient who required transitory myorelaxation for facilitation of bronchoscopy. Administration of myorelaxant induced a minimal decrease in both BIS and electromyographic activity, whereas twitch decreased to 0, assessing complete myorelaxation. No change in BIS or electromyographic activity where observed while twitch recovered. (B ) BIS, electromyographic activity, and twitch height of the first response of train-of-four sequence recorded at the adductor pollicis muscle (twitch) trend recordings in a patient who required transitory myorelaxation for facilitation of transoesophageal echocardiography. Administration of myorelaxant induced a major decrease in both BIS and electromyographic activity, whereas twitch decreased to 0, assessing complete myorelaxation. During recovery of myorelaxation, BIS and electromyographic activity recovered smoothly and earlier than twitch recovery.

Fig. 3. (A ) Bispectral Index (BIS), electromyographic activity (EMG), and twitch height of the first response of train-of-four sequence recorded at the adductor pollicis muscle (twitch) trend recordings in a patient who required transitory myorelaxation for facilitation of bronchoscopy. Administration of myorelaxant induced a minimal decrease in both BIS and electromyographic activity, whereas twitch decreased to 0, assessing complete myorelaxation. No change in BIS or electromyographic activity where observed while twitch recovered. (B ) BIS, electromyographic activity, and twitch height of the first response of train-of-four sequence recorded at the adductor pollicis muscle (twitch) trend recordings in a patient who required transitory myorelaxation for facilitation of transoesophageal echocardiography. Administration of myorelaxant induced a major decrease in both BIS and electromyographic activity, whereas twitch decreased to 0, assessing complete myorelaxation. During recovery of myorelaxation, BIS and electromyographic activity recovered smoothly and earlier than twitch recovery.

Close modal

Figure 4shows the distribution of patients during the baseline and myorelaxation periods according to the four standard BIS range assessments. The number of patients under light or deep sedation versus  general anesthesia or deep hypnotic state was markedly overestimated before administration of myorelaxant (53 vs.  2%, P < 0.001).

Fig. 4. Distribution of the 45 patients during the baseline and myorelaxation study periods according to the four standard Bispectral Index (BIS) range assessments 11: light sedation (BIS ≥ 85), deep sedation (65 ≥ BIS < 85), general anesthesia (40 ≥ BIS < 65), and deep hypnotic state (BIS < 40). The number of patients under light or deep sedation versus  general anesthesia or deep hypnotic state was markedly overestimated before administration of myorelaxant (53 vs.  2%, P < 0.001).

Fig. 4. Distribution of the 45 patients during the baseline and myorelaxation study periods according to the four standard Bispectral Index (BIS) range assessments 11: light sedation (BIS ≥ 85), deep sedation (65 ≥ BIS < 85), general anesthesia (40 ≥ BIS < 65), and deep hypnotic state (BIS < 40). The number of patients under light or deep sedation versus  general anesthesia or deep hypnotic state was markedly overestimated before administration of myorelaxant (53 vs.  2%, P < 0.001).

Close modal

We observed a significant correlation between BIS and electromyographic activity during the baseline period (R = 0.78, P < 0.001;fig. 5, A ). We also observed a significant correlation between ΔBIS (= baseline BIS value − myorelaxation BIS value) and both baseline BIS (R = 0.85, P < 0.001;fig. 5, B ) and baseline electromyographic activity (R = 0.74, P < 0.001;fig. 5, C ). Multiple regression analysis performed with ΔBIS as a dependent variable, and baseline variables as independent variables showed that ΔBIS was significantly correlated only to both baseline BIS and electromyographic values (table 4).

Fig. 5. (A ) Correlation between the Bispectral Index (BIS) and electromyographic activity (EMG) during the baseline period. (B ) Correlation between the decrease of BIS (ΔBIS = BIS baseline − BIS myorelaxation) following administration of myorelaxant and BIS baseline value (BIS baseline). (C ) Correlation between the decrease of BIS (ΔBIS = BIS baseline − BIS myorelaxation) following administration of myorelaxant and electromyographic baseline value (EMG baseline; n = 45).

Fig. 5. (A ) Correlation between the Bispectral Index (BIS) and electromyographic activity (EMG) during the baseline period. (B ) Correlation between the decrease of BIS (ΔBIS = BIS baseline − BIS myorelaxation) following administration of myorelaxant and BIS baseline value (BIS baseline). (C ) Correlation between the decrease of BIS (ΔBIS = BIS baseline − BIS myorelaxation) following administration of myorelaxant and electromyographic baseline value (EMG baseline; n = 45).

Close modal

Table 4. Result of Multiple Regression Analysis with ΔBIS = BIS Baseline − BIS Myorelaxation as Dependent Variable (n = 45)

Baseline = study period (20 min) before administration of myorelaxant; myorelaxation = study period (20 min) after administration of myorelaxant; BIS = bispectral index; TOTPOW = total power of the signal; EMG = electromyographic activity; SQI = signal quality index.

Table 4. Result of Multiple Regression Analysis with ΔBIS = BIS Baseline − BIS Myorelaxation as Dependent Variable (n = 45)
Table 4. Result of Multiple Regression Analysis with ΔBIS = BIS Baseline − BIS Myorelaxation as Dependent Variable (n = 45)

For the 30 patients in the transitory group, during the recovery period, the variables BIS, electromyographic activity, TOTPOW, and SQI recovered values, which were not significantly different from baseline values (table 3). Recovery of BIS and electromyographic activity occurred generally earlier than recovery of acceleromyography at the adductor pollicis (fig. 3, B ). According to our definition of recovery curve shape, recovery of BIS and electromyographic activity was smooth in 13 patients and parallel with acceleromyography. On the contrary, in the 17 remaining patients, we observed a stepped recovery of BIS and electromyographic curves, whereas recovery of acceleromyography at the adductor pollicis occurred progressively and smoothly (fig. 6).

Fig. 6. Bispectral Index (BIS), electromyographic activity (EMG), and twitch height of the first response of train-of-four sequence recorded at the adductor pollicis muscle (twitch) trend recordings in a patient who required transitory myorelaxation for facilitation of transesophageal echocardiography. Administration of myorelaxant induced a major decrease in both BIS and electromyographic activity, whereas twitch decreased to 0, assessing complete myorelaxation. BIS and electromyographic activity recovered irregularly with several steps (arbitrarily defined as increase in BIS >20 units and/or increase in electromyographic activity >5 dB occurring in less than 1 min), whereas twitch recovered progressively and smoothly.

Fig. 6. Bispectral Index (BIS), electromyographic activity (EMG), and twitch height of the first response of train-of-four sequence recorded at the adductor pollicis muscle (twitch) trend recordings in a patient who required transitory myorelaxation for facilitation of transesophageal echocardiography. Administration of myorelaxant induced a major decrease in both BIS and electromyographic activity, whereas twitch decreased to 0, assessing complete myorelaxation. BIS and electromyographic activity recovered irregularly with several steps (arbitrarily defined as increase in BIS >20 units and/or increase in electromyographic activity >5 dB occurring in less than 1 min), whereas twitch recovered progressively and smoothly.

Close modal

Finally, among the 45 patients of this study, 16 had been studied using the old and new versions of the BIS® monitor simultaneously. We observed that the overestimation of BIS value because of high electromyographic activity was still present with the new BIS® XP monitor (fig. 7). Indeed, administration of neuromuscular blockade significantly decreased BIS values with both the old (−25 ± 15%, P < 0.001) and the new (−21 ± 14%, P < 0.001) monitors (table 5). In addition, there was no significant difference in these BIS decreases after myorelaxant administration between the two monitors.

Fig. 7. Bispectral Index (BIS) and electromyographic activity (EMG) recorded simultaneously with the old BIS® monitor (BIS® and EMG, software version 2.10, three-electrode sensor) and the new BIS® monitor (BIS® XP and EMG XP, software version 3.12, four-electrode sensor). Whatever the monitor, administration of myorelaxant induced a major decrease in both bispectral and electromyographic signals. The regular peaks in electromyographic activity are linked to regular resets of the BIS® monitors and may occasionally induce change in baseline level of BIS and electromyographic activity. Note the arrows marked “(1),” which show a synchronous change in BIS and electromyographic values on both monitors, whereas the arrows marked “(2)” show an asynchronous change in BIS and electromyographic values on both monitors.

Fig. 7. Bispectral Index (BIS) and electromyographic activity (EMG) recorded simultaneously with the old BIS® monitor (BIS® and EMG, software version 2.10, three-electrode sensor) and the new BIS® monitor (BIS® XP and EMG XP, software version 3.12, four-electrode sensor). Whatever the monitor, administration of myorelaxant induced a major decrease in both bispectral and electromyographic signals. The regular peaks in electromyographic activity are linked to regular resets of the BIS® monitors and may occasionally induce change in baseline level of BIS and electromyographic activity. Note the arrows marked “(1),” which show a synchronous change in BIS and electromyographic values on both monitors, whereas the arrows marked “(2)” show an asynchronous change in BIS and electromyographic values on both monitors.

Close modal

Table 5. Comparison of Bispectral Monitoring Variables in 16 Patients Examined Simultaneously with the Old and New Monitors

Data are mean ± SD.

*P < 0.05 versus  A2000 Monitor values.

P < 0.05 versus  baseline values.

Baseline = study period (20 min) before administration of myorelaxant; myorelaxation = study period (20 min) after administration of myorelaxant; BIS = bispectral index; TOTPOW = total power of the signal; EMG = electromyographic activity; SQI = signal quality index.

Table 5. Comparison of Bispectral Monitoring Variables in 16 Patients Examined Simultaneously with the Old and New Monitors
Table 5. Comparison of Bispectral Monitoring Variables in 16 Patients Examined Simultaneously with the Old and New Monitors

In this study, we have shown that in sedated ICU patients, (1) BIS markedly decreased after administration of myorelaxant; (2) the decrease in BIS value following neuromuscular blockade was significantly correlated with BIS and electromyographic values before administration of myorelaxant; (3) using standard BIS range guidelines, the number of patients under light or deep sedation versus  general anesthesia or deep hypnotic state was markedly overestimated before administration of myorelaxant; and (4) overestimation of the BIS value because of high electromyographic activity was still present with the new BIS® XP monitor.

Sedation is an important issue in ICU patients, and assessing the level of unconsciousness of patients is mandatory to determine the amount of sedative drugs to be administered. 24Whereas adequate sedation reduces the stress response and improves the tolerance of routine ICU procedures, oversedation may be deleterious by increasing the duration of mechanical ventilation and the incidence of pneumonia. 10,25,26Several clinical scores such as the Ramsay scale and the Sedation-Agitation Scale (table 1) are subjective but reliable methods to assess the level of sedation in sedated ICU patients and have been used widely both clinically and in many comparative sedation trials. 16,24,27In the current study, we used the Sedation-Agitation Scale instead of the Ramsay scale. Indeed, although the Ramsay scale has existed for much longer, it is actually considered to be less reliable than the Sedation-Agitation Scale to assess sedation in ICU patients. 13,16Elsewhere, we have chosen to investigate deeply sedated patients with a Sedation-Agitation Scale of 1. Indeed, it should be kept in mind that this is below the sensitive range of this score and may represent a large potential source of patient variability. Nevertheless, one of the potential uses of BIS monitoring in the ICU should be to detect oversedation. Therefore, we have voluntary chosen to investigate BIS monitoring in patients with a Sedation-Agitation Scale score of 1, i.e. , those who were exposed to a risk of oversedation much more than others.

For several years, BIS monitoring has been used in sedated ICU patients as a tool for an objective assessment of sedation or hypnotic drug effect trials. 24Indeed, good correlations were reported between clinical sedation scores and BIS values in sedated ICU patients. 3–6,8,9However, among these studies, some of them pointed out that BIS was correctly correlated with sedation scores in only some sedated ICU patients. 4,5,8,9On the other hand, following the cases reported by Bruhn et al.  12in anesthetized patients, several authors have recently noted that BIS reliability could be severely impaired because of excessive electromyographic activity, which could contaminate BIS monitoring and therefore lead to overestimated and irrelevant BIS values in sedated ICU patients. 5,8,9,13Nevertheless, to our knowledge, the magnitude of BIS overestimation in sedated ICU patients without neuromuscular blockade has not previously been established. Our study clearly demonstrated that BIS overestimation in sedated ICU patients without neuromuscular blockade was significant and, in addition, showed that this overestimation was correlated both with BIS and electromyographic values before administration of myorelaxant (fig. 5).

In the current study, in sedated ICU patients, we used the BIS range guidelines, which have been initially developed for anesthetized patients. 11This point should be kept in mind because the relevance of these anesthetic guidelines for sedated ICU patients had not yet been proven. Nevertheless, to the best of our knowledge, no BIS range guidelines that describe appropriate and inappropriate levels of sedation when assessed by such a monitor have been widely accepted for the ICU.

After administration of myorelaxant, 32 of the 45 studied patients had a decrease of at least one step in BIS range assessment (fig. 2, A ), which means that BIS range assessment was overestimated in approximately 71% of our patients before administration of myorelaxant. This finding is important because only 10 of the 45 studied patients (approximately 22%) showed a baseline electromyographic value greater than the threshold of 42 dB, above which electromyographic activity is considered to artifactually increase BIS values 9,13(fig. 2, B ).

The number of patients under light or deep sedation versus  general anesthesia or deep hypnotic state was markedly overestimated before administration of myorelaxant (53 vs.  2%, P < 0.001;fig. 4). Therefore, clinicians who wish to determine the amount of sedation in ICU patients only from BIS monitoring would expose them to an unnecessary oversedation with potential morbidity, mortality, and increased cost. 10,25,26Elsewhere, because we found that on one hand, baseline BIS value was significantly correlated to baseline electromyographic activity, and on the other hand, that the magnitude of decrease in BIS after myorelaxation was significantly correlated both with BIS and with electromyographic values before administration of neuromuscular blocker, the inference is that the “true” BIS value could be roughly approximated from both BIS and electromyographic values before neuromuscular blockade. It should be remembered that before administration of neuromuscular blockers, overestimation of BIS was present, although SQI showed high values (90 ± 8%), indicating that BIS monitoring should have been reliable. Even with SQI near 100%, we still observed a significant decrease in BIS value after administration of myorelaxant drug (fig. 1). Finally, even if myorelaxation significantly increased SQI, we did not find any significant correlation between the decrease in BIS value after neuromuscular blockade and the SQI before myorelaxation.

The BIS was initially developed as an objective measurement of the depth of anesthesia in the operating room, where patients are highly sedated and often paralyzed. 1In the ICU, patients are generally less profoundly sedated than in the operating room and, most of the time, do not receive neuromuscular blockers. 23Also, the data used to develop the BIS algorithm where obtained from a large database of anesthetized patients and may not necessary always apply to sedated ICU patients. 28In addition, the sources of interference during BIS monitoring are various and common in ICU. 13,29Among those artifacts, high electromyographic activity is probably the most common limitation to BIS monitoring in ICU patients. 5,8,9,13Nonetheless, the contamination of BIS by electromyographic activity is inherent in the calculation of BIS by the monitor: BIS® software uses electroencephalographic signals up to 47 Hz, whereas electroencephalographic and electromyographic signals are conventionally considered to exist in the 0.5- to 30-Hz and 30- to 300-Hz bands, respectively. 11This overlap of electroencephalographic and electromyographic signals will undoubtedly remain the main pitfall of BIS monitoring in the ICU or in the operating room in unparalyzed patients.

Recovery of electromyographic activity and thus of BIS occurred earlier than recovery of acceleromyography at the adductor pollicis (fig. 3, B ). Because the BIS® sensor was applied on the forehead, this could be explained by the known delayed recovery after myorelaxation of adductor pollicis as compared to facial muscles such as the orbicularis oculi. 30On the other hand, in 17 patients, recovery of BIS and electromyographic activity was unexpectedly abrupt and irregular, with several steps, whereas recovery of acceleromyography at the adductor pollicis occurred progressively and smoothly. This is probably linked to the ICU environment, where various stimuli are unavoidable. 29,31Indeed, Johansen et al.  32recently showed that administration of stimulus to sedated ICU patients could induce increases in both electromyographic activity and BIS and that these increases may also depend on the initial level of hypnosis. Moreover, noise alone, which is generally high in the ICU, could be seen as a noxious stimulus, causing sympathetic activation and therefore increasing BIS in lightly sedated patients. 33,34 

Most of the previous studies investigating the accuracy of BIS monitoring in ICU patients were performed without myorelaxation (which was even an exclusion criterion for some of them) and consequently involved exposure to the highly contaminating effect of electromyographic activity, which should limit the interpretation of their conclusions. 3–6,8,9In the extreme, BIS values up to 97, related to very high electromyographic activity, were recently reported in brain-dead patients, which demonstrated that BIS could be highly contaminated by electromyographic activity. 15Because BIS monitoring has been recently proposed as a monitor of neurologic function in unsedated brain-injured patients, physicians should keep in mind the potential major interference from high electromyographic activity. 18,35–43 

Finally, there are some limits in the interpretation of the results of our study. First, we only investigated patients who were deeply sedated with sufentanil and midazolam, and therefore, we could not extrapolate the magnitude of overestimation of BIS to other sedation protocols. Nevertheless, this was done purposely to investigate a homogeneously sedated group of patients under a common sedative drug protocol administration, in contrast to most previous studies of BIS in the ICU. 4–6,9,13Second, this study was performed with the monitor Aspect 2000 running software version 2.10 with a three-electrode sensor. BIS monitoring is a recent technology, and improvements in hardware, software, and sensor are continuously available (BIS® XP, software version 3.12, four-electrode sensor). From the point of view of AspectMS®, this new equipment should be better at detecting interference from electromyographic activity and other artifact conditions often found in the ICU. Indeed, Coluzzi et al.  21recently reported an improvement in sedation monitoring of ICU patients with the new BIS® XP monitor as compared to the old standard BIS® monitor. Nevertheless, our results from the 16 patients investigated simultaneously with the two monitors showed that neuromuscular blockade still induced a significant decrease in BIS values as assessed by the new XP monitor, which was similar to that observed with the old monitor (table 5).

In summary, BIS monitoring in the ICU may be dramatically overestimated because of high muscular activity. The magnitude of the BIS decrease following neuromuscular blockade is significantly correlated to both BIS and electromyographic values before myorelaxation. The corollary of this overestimation could be a BIS-induced oversedation, which is unnecessary and may be deleterious for ICU patients.

The authors thank David J. Baker, M.D., F.R.C.A. (Département d'Anesthésie-Réanimation, Hôpital Necker, Paris, France), for reviewing the manuscript.

1.
Gan TJ, Glass PS, Windsor A, Payne F, Rosow C, Sebel P, Manberg P: Bispectral index monitoring allows faster emergence and improved recovery from propofol, alfentanil, and nitrous oxide anesthesia. BIS Utility Study Group. A nesthesiology 1997; 87: 808–15
2.
Rampil IJ: A primer for EEG signal processing in anesthesia. A nesthesiology 1998; 89: 980–1002
3.
De Deyne C, Struys M, Decruyenaere J, Creupelandt J, Hoste E, Colardyn F: Use of continuous bispectral EEG monitoring to assess depth of sedation in ICU patients. Intensive Care Med 1998; 24: 1294–8
4.
Simmons LE, Riker RR, Prato BS, Fraser GL: Assessing sedation during intensive care unit mechanical ventilation with the bispectral index and the Sedation-Agitation Scale. Crit Care Med 1999; 27: 1499–504
5.
Riker RR, Fraser GL, Simmons LE, Wilkins ML: Validating the Sedation-Agitation Scale with the Bispectral Index and Visual Analog Scale in adult ICU patients after cardiac surgery. Intensive Care Med 2001; 27: 853–8
6.
Berkenbosch JW, Fichter CR, Tobias JD: The correlation of the bispectral index monitor with clinical sedation scores during mechanical ventilation in the pediatric intensive care unit. Anesth Analg 2002; 94: 506–11
7.
Brocas E, Dupont H, Paugam-Burtz C, Servin F, Mantz J, Desmonts JM: Bispectral index variations during tracheal suction in mechanically ventilated critically ill patients: Effect of an alfentanil bolus. Intensive Care Med 2002; 28: 211–3
8.
Frenzel D, Greim CA, Sommer C, Bauerle K, Roewer N: Is the bispectral index appropriate for monitoring the sedation level of mechanically ventilated surgical ICU patients? Intensive Care Med 2002; 28: 178–83
9.
Riess ML, Graefe UA, Goeters C, Van Aken H, Bone HG: Sedation assessment in critically ill patients with bispectral index. Eur J Anaesthesiol 2002; 19: 18–22
10.
Triltsch AE, Welte M, Von Homeyer P, Grosse J, Genahr A, Moshirzadeh M, Sidiropoulos A, Konertz W, Kox WJ, Spies CD: Bispectral index-guided sedation with dexmedetomidine in intensive care: A prospective, randomized, double blind, placebo-controlled phase II study. Crit Care Med 2002; 30: 1007–14
11.
Johansen JW, Sebel PS: Development and clinical application of electroencephalographic bispectrum monitoring. A nesthesiology 2000; 93: 1336–44
12.
Bruhn J, Bouillon TW, Shafer SL: Electromyographic activity falsely elevates the bispectral index. A nesthesiology 2000; 92: 1485–7
13.
Nasraway SA Jr, Wu EC, Kelleher RM, Yasuda CM, Donnelly AM: How reliable is the Bispectral Index in critically ill patients? A prospective, comparative, single-blinded observer study. Crit Care Med 2002; 30: 1483–7
14.
Greif R, Greenwald S, Schweitzer E, Laciny S, Rajek A, Caldwell JE, Sessler DI: Muscle relaxation does not alter hypnotic level during propofol anesthesia. Anesth Analg 2002; 94: 604–8
15.
Vivien B, Paqueron X, Le Cosquer P, Langeron O, Coriat P, Riou B: Detection of brain death onset using the bispectral index in severely comatose patients. Intensive Care Med 2002; 28: 419–25
16.
Riker RR, Picard JT, Fraser GL: Prospective evaluation of the Sedation-Agitation Scale for adult critically ill patients. Crit Care Med 1999; 27: 1325–9
17.
Doi M, Gajraj RJ, Mantzaridis H, Kenny GN: Effects of cardiopulmonary bypass and hypothermia on electroencephalographic variables. Anaesthesia 1997; 52: 1048–55
18.
Gilbert TT, Wagner MR, Halukurike V, Paz HL, Garland A: Use of bispectral electroencephalogram monitoring to assess neurologic status in unsedated, critically ill patients. Crit Care Med 2001; 29: 1996–2000
19.
Guignard B, Chauvin M: Bispectral index increases and decreases are not always signs of inadequate anesthesia. A nesthesiology 2000; 92: 903
20.
Hemmerling TM, Fortier JD: Falsely increased bispectral index values in a series of patients undergoing cardiac surgery using forced-air-warming therapy of the head. Anesth Analg 2002; 95: 322–3
21.
Coluzzi F, Di Filippo C, Rossetti E, Summonti D, Mattia C: BIS monitoring in ICU: Advantages of the new XP generation (abstract). Crit Care 2002; 6 (suppl 1): P68.
22.
Pearson AJ, Harper NJ, Pollard BJ: The infusion requirements and recovery characteristics of cisatracurium or atracurium in intensive care patients. Intensive Care Med 1996; 22: 694–8
23.
Murray MJ, Cowen J, DeBlock H, Erstad B, Gray AW Jr, Tescher AN, McGee WT, Prielipp RC, Susla G, Jacobi J, Nasraway SA Jr, Lumb PD: Clinical practice guidelines for sustained neuromuscular blockade in the adult critically ill patient. Crit Care Med 2002; 30: 142–56
24.
Jacobi J, Fraser GL, Coursin DB, Riker RR, Fontaine D, Wittbrodt ET, Chalfin DB, Masica MF, Bjerke HS, Coplin WM, Crippen DW, Fuchs BD, Kelleher RM, Marik PE, Nasraway SA Jr, Murray MJ, Peruzzi WT, Lumb PD: Clinical practice guidelines for the sustained use of sedatives and analgesics in the critically ill adult. Crit Care Med 2002; 30: 119–41
25.
Kollef MH, Levy NT, Ahrens TS, Schaiff R, Prentice D, Sherman G: The use of continuous i.v. sedation is associated with prolongation of mechanical ventilation. Chest 1998; 114: 541–8
26.
Sydow M, Neumann P: Sedation for the critically ill. Intensive Care Med 1999; 25: 634–6
27.
Ramsay MA, Savege TM, Simpson BR, Goodwin R: Controlled sedation with alphaxalone-alphadolone. BMJ 1974; 2: 656–9
28.
Drummond JC: Monitoring depth of anesthesia: With emphasis on the application of the bispectral index and the middle latency auditory evoked response to the prevention of recall. A nesthesiology 2000; 93: 876–82
29.
Johansen JW: Logistics of 25 hour bispectral index monitoring in the ICU (abstract). Anesth Analg 2000; 90 (suppl): S212
30.
Donati F, Meistelman C, Plaud B: Vecuronium neuromuscular blockade at the diaphragm, the orbicularis oculi, and adductor pollicis muscles. A nesthesiology 1990; 73: 870–5
31.
Chassard D, Nakamura V, Bouletreau P, Allaouchiche B, Duflo F: Result of an anti-noise campaign in an intensive care unit (abstract). A nesthesiology 2002; 96 (suppl): A407
32.
Johansen JW, Ouchi T: Bispectral index and electromyographic activity change in response to sedation assessment (OAA/S) in the surgical intensive care unit (abstract). A nesthesiology 2001; 95 (suppl): A387
33.
Kim DW, Kil HY, White PF: The effect of noise on the bispectral index during propofol sedation. Anesth Analg 2001; 93: 1170–3
34.
Allaouchiche B, Duflo F, Debon R, Bergeret A, Chassard D: Noise in the postanaesthesia care unit. Br J Anaesth 2002; 88: 369–73
35.
England MR: The change in bispectral index during a hypovolemic cardiac arrest. A nesthesiology 1999; 91: 1947–9
36.
Billard V: Brain injury under general anesthesia: Is monitoring of the EEG helpful? Can J Anaesth 2001; 48: 1055–60
37.
Mérat S, Lévecque JP, Le Gulluche Y, Diraison Y, Brinquin L, Hoffmann JJ: BIS monitoring may allow the detection of severe cerebral ischemia. Can J Anaesth 2001; 48: 1066–9
38.
Valero R, Gambús P, Zavala B, Guix E, Fábregas N: BIS monitoring as an outcome predictor in severely brain damaged nonsedated critically ill patients (abstract). Eur J Anaesthesiol 2001; 18 (suppl): A245
39.
Añez Simóna C, Camps Vidal M, Puig Bitria R, Rull Bartomeu M: La monitorización del índece biespectral del EEG podría ser un buen monitor de la evolución neurológica de los pacientes en coma de origen metabólico? Rev Esp Anestesiol Reanim 2001; 48: 344–5
40.
O'Connor MF, Woo MC, Shahul S, Kress JP, Hall J: Poor correlation of BIS™ value with neurologic status in metabolically deranged or neurologically injured patients (abstract). A nesthesiology 2002; 96 (suppl): A378
41.
Pandit JJ, Schmelzle-Lubiecki B, Goodwin M, Saeed N: Bispectral index-guided management of anaesthesia in permanent vegetative state. Anaesthesia 2002; 57: 1190–4
42.
Szekely B, Saint-Marc T, Degremont AC, Castelain MH, Fischler M: Value of bispectral index monitoring during cardiopulmonary resuscitation. Br J Anaesth 2002; 88: 443–4
43.
Wu CC, Lin CS, Mok MS: Bispectral index monitoring during hypoglycemic coma. J Clin Anesth 2002; 14: 305–6