Background

The design of an endotracheal tube has been shown to influence the passage of the tube through the glottis during fiberoptic intubation. Difficulty in passing the endotracheal tube can occur if the aryepiglottic folds obstruct the passage of the bevel. The relevant aspects of endotracheal tube design include the shape of the bevel, the material used by the manufacturer, and the ability of the tube to conform to the shape of the fiberscope. The aim of the current study was to compare the ease of passage through the glottis of two different tubes. One tube was a wire reinforced polyvinyl chloride tube with a standard bevel and the other was a newly designed tube with a bevel of different shape and made of silicone rubber. The new design is for use with the a commerical intubating laryngeal mask.

Methods

The authors studied a population of 30 patients who received a standard anesthetic. In all cases, oral fiberoptic intubation was attempted. Anesthetic was administered to each patient using both tubes, and before the study the order of the tubes was randomized. The difficulty in passing the tube was assessed by a blinded observer and graded using a three-point scale (grade 1: no difficulty passing the tube; grade 2: obstruction to passing the tube relieved by withdrawal and a 90 degrees anticlockwise rotation; grade 3: obstruction necessitating more than one manipulation or external laryngeal manipulation).

Results

In 27 patients, no difficulty was shown by use of the silicone-tipped tube. In only three patients was there difficulty that necessitated a 90 degrees anticlockwise twist. With the wire-reinforced tube, no difficulty was experienced on 14 occasions. Grade 1 difficulty was experienced eight times and difficulty necessitating more than one maneuver, head movement, or external laryngeal manipulation was seen on eight occasions. Statistical significance was achieved at P = 0.0002 (Wilcoxon signed rank test).

Conclusions

The authors conclude that the use of the silicone-tipped tube with the new bevel design may provide an advantage in the clinical situation of fiberoptic intubation.

THE technique of fiberoptic intubation was first described by Murphy in 1967. 1It is now a well-established technique in anesthesia in which difficult intubation is anticipated or encountered unexpectedly.

Obstruction to passing the tube over the fiberscope and into the trachea may be encountered at the laryngeal inlet. It has been suggested that various aspects of tube design may influence the passage of the tube through the glottis. The Intubating Laryngeal Mask Airway (Intavent, Maidenhead, Berkshire, United Kingdom) is supplied with its own tube. 2The bevel of this tube is made from silicone rubber and is softer than the standard polyvinyl chloride design and the bevel is hemispherical, with the leading edge in the midline; the tube itself is also less rigid than the standard design (fig. 1). The aim of the current study was to compare the ease of passage through the glottis of a wire-reinforced tube with a standard bevel with that of a silicone-tipped (ILMA) tube.

Fig. 1. The bevel design of the two tubes compared in the study. (Left ) The standard wire-reinforced tube. (Right ) The silicone-tipped bevel of the Intubation Laryngeal Mask Airway tube.

Fig. 1. The bevel design of the two tubes compared in the study. (Left ) The standard wire-reinforced tube. (Right ) The silicone-tipped bevel of the Intubation Laryngeal Mask Airway tube.

Close modal

The study was approved by the local ethics committee and we received fully informed written consent from the patients. We studied 30 adult patients who were classified as American Society of Anesthesiologists physical status I or II and who were undergoing elective surgery. Patients were excluded from the study if they were had difficult airway. We anesthetized the patients in a standard manner using 2 to 3 mg/kg propofol and 1 to 2 μg/kg fentanyl followed by 0.6 mg/kg atracurium. We performed manual ventilation for 3 min to allow full muscle relaxation to develop; this was considered to be adequate if the patient did not cough after the passage of the fiberscope through the vocal cords. In all cases, oral fiberoptic intubation was attempted. Each patient was to undergo intubation using both tubes and, before the study, the patients were randomized to undergo intubation with either the wire-reinforced tube or the silicone-tipped tube in the first instance. We compared a size 7.0-mm ID wire-reinforced tube (10.6 mm OD; Mallinkrodt Medical, Athlone, Ireland) with a size 7.0-mm ID silicone-tipped tube (10.0 mm OD). In all cases, we used the Olympus LF1 intubating laryngoscope (4 mm OD; Keymed Ltd, Stock Road, Southend on Sea, Essex, England).

A 10-cm William airway was placed in the patient’s mouth, and jaw thrust was applied to hold the oral airway in place. The first tube was passed over the fiberscope. The fiberscope was advanced through the glottis. At this stage, the fiberscope was in the trachea; however, the tip of the tube was above the vocal cords. The fiberscope, the tube, and patient’s face were covered with a green towel. An investigator blind to the study then entered the room and passed the tube into the trachea. The observer wore a pair of gloves and passed the tube covered with the green towel into the trachea. The investigator then left the room, and the tube and the fiberscope were removed. The patient underwent manual ventilation for a brief period to ensure that adequate oxygenation was maintained and the maneuver was repeated using the second tube. The tube was oriented on the fiberscope such that for each tube the bevel faced to the left. The observer graded the intubation using a three-point scale (table 1). We analyzed the results on the three-point scale using the Wilcoxon signed rank test.

Table 1. Grading of Difficulty of Passage of the Tube

Table 1. Grading of Difficulty of Passage of the Tube
Table 1. Grading of Difficulty of Passage of the Tube

We studied 30 patients. We graded the difficulty of intubation as grade 0, grade 1, or grade 2. The silicone-tipped tube was passed without difficulty in 27 patients and only necessitated a 90° anticlockwise rotation in 3. Using the wire-reinforced tube, no difficulty was encountered in 14 patients, in 8 patients the passage of the tube necessitated a 90° anticlockwise rotation, and in another 8 patients there was difficulty intubating, necessitating more than one manuever, external laryngeal manipulation, or movement of the head.

We analyzed our data using the Wilcoxon signed rank test and demonstrated statistical significance (P = 0.0002). We did not demonstrate carry-over effect (table 2).

Table 2. Results of the Difficulty of Passage of the Tube through the Glottis

ILMA = Intubating Laryngeal Mask Airway (Intavent, Maidenhead, Berkshire, United Kingdom).

Table 2. Results of the Difficulty of Passage of the Tube through the Glottis
Table 2. Results of the Difficulty of Passage of the Tube through the Glottis

It has been shown that the passage of an endotracheal tube over a fiberscope and into the trachea may be difficult. 3The incidence of this difficulty has been described as between 23%4and 46%. 3The factors implicated in causing difficulty in passing the tube into the glottis include the diameter of the fiberscope and the design of the tube and the tube tip.

Hakala and Randell 4compared the effect of differing fiberscope diameters on the ease of passage of an endotracheal tube. A fiberscope with a thicker insertion cord promoted greater ease of passage through the glottis. It was thought that the excess free distance between the inside of the tube and the insertion cord may lead to snagging on the laryngeal inlet. The use of a pediatric tube will decrease the incidence of difficult fiberoptic intubation by decreasing the distance between the fiberscope and the tube. 5 

Other aspects of tube design may affect its progression. It has been suggested that a wire-reinforced tube may be easier to pass than a preformed polyvinyl chloride tube. 6This tube has a less natural curve and may cause less displacement of the fiberscope if rotation of the tube is necessary. It may also follow the bronchoscope better and, being soft, may cause less damage to the soft tissues of the airway.

The influence of endotracheal tube tip design has previously been described as having an effect on the passage of a tube through the vocal cords. A tube was designed and described in 1983 by Dr. P. Moore. 7This tube has no bevel and tapers in a conical fashion. When this tube is positioned on the fiberscope, the gap between the inside of the tube and the fiberscope is minimized. It was shown that this design caused less obstruction during passage through the vocal cords. This tube was only described as a prototype.

The ILMA was described in 1997 as an adjunct to difficult intubation. The ILMA is supplied with its own tube. The design of this tube is such that it facilitates passage through the intubating laryngeal mask and into the trachea. The bevel is made of silicone rubber and is softer than the standard polyvinyl chloride bevel. The bevel is hemispherical and has a leading edge in the midline (fig. 1). The ILMA tube is relatively less stiff than the wire-reinforced tube, which conforms better to the fiberoptic scope and may allow easier passage. The ILMA tube is reusable.

In our study we compared a size 7.0 flexometallic tube to a size 7.0 ILMA tube. This is the smallest size of ILMA tube available and we thought this decision afforded the best comparison available. We chose to use the wire-reinforced tube because it is the one commonly used in our hospital practice (fig. 1).

External manipulation of the tube or larynx can affect passage. During fiberoptic intubation, the tip of the tube will catch most frequently on the right aryepiglottic fold or the arytenoid cartilage. This has also been described with the passage of an endotracheal tube over a gum elastic bougie. The passage of the tube was found to be easier after a 90° anticlockwise rotation was applied to the tube. 8,9The effect of this maneuver was to rotate the tube into a position in which the tip lies anterior to the bougie and is impeded by nothing.

We have demonstrated that the use of the silicone-tipped tube (ILMA) is advantageous during fiberoptic intubation. We believe that the passage of the tube is made easier primarily by the bevel design. This bevel design has previously been shown to be advantageous in the passage of this same tube through the glottis. A study by Murashima et al . 10compared a tube with a wedge-shaped end with the tube we used in our study when passing the tube through the ILMA. This article states that the midline bevel guided the tube away from the arytenoids and toward the center of the glottis.

Improved flexibility of the tube may be a factor in the ease of fiberoptic intubation. A study by Brull et al . 11suggested that a more flexible tube can change direction more easily and thereby lead to easier fiberoptic intubation. The silicone-tipped tube has been shown to be more flexible than the polyvinyl chloride wire-reinforced tube in vitro . 12 

Fiberoptic intubation can be difficult, requiring considerable operator skill, and the consequences of failure are great. Repeated attempts at passage of the tube through the glottis may lead to laryngeal trauma or bleeding in the airway, and this may confound further attempts at airway control. We propose that the ease of passage of an ILMA tube with a hemispherical bevel with the leading edge in the midline in comparison with a more rigid flexometallic tube of standard bevel design confers a clinical advantage during fiberoptic intubation. This tube may be considered as a first choice for fiberoptic intubation or may be considered as an adjunct to fiberoptic intubation when a polyvinyl chloride tube proves to be difficult to pass on the first attempt.

The authors thank Sally Hollis, Department of Medical Statistics, Lancaster University, Lancaster, United Kingdom.

1.
Murphy P: A fibreoptic scope used for nasal intubation. Anaesthesia 1967; 22: 489–91
2.
Brain AIJ, Varghese C: The intubating laryngeal mask airway. Instruction manual, 1st edition. 1998.
3.
Coe PA, King TA, Towey RM: Teaching guided fibreoptic nasotracheal intubation. An assessment of an anaesthetic technique to aid training. Anaesthesia 1988; 43: 410–3
4.
Hakala P, Randell T: Comparison between two fibrescopes with different diameter insertion cords for fibreoptic intubation. Anaesthesia 1995; 50: 735–7
5.
Marsh NJ: Easier Fiberoptic intubations. A nesthesiology 1992; 76: 860–1
6.
Calder I: When the endotracheal tube will not pass over the flexible fiberoptic bronchoscope (letter). A nesthesiology 1992; 77: 398
7.
Jones HE, Pearce AC, Moore P: Fibreoptic intubation. Influence of tracheal tube tip design. Anaesthesia 1993; 48: 672–4
8.
Cossham PS: Difficult intubation. Br J Anaesth 1985; 55: 187–8
9.
Dogra S, Falconer R, Latto IP: Successful difficult intubation. Tracheal tube placement over a gum-elastic bougie. Anaesthesia 1990; 45: 774–6
10.
Murashima K, Fukutome T, Brimaombe J: A comparison of two silicone-reinforced tracheal tubes with different bevels for use with the intubating laryngeal mask. Anaesthesia 1999; 54: 1198–200
11.
Brull SJ, Wiklund R, Ferris C, Connelly NR, Ehrenwerth J, Silvermann DG: Facilitation of fibreoptic oral intubation with a flexible tracheal tube. Anaesth Aanalg 1994; 78: 746–8
12.
Brain AIJ, Verghese C, Addy EV, Kapila A: The intubating laryngeal mask: I. Development of a new device for intubation of the trachea. Br J Anaesth 1997; 79: 699–703