IMPLANTABLE pumps with intrathecal delivery catheters are increasingly used for management of chronic pain conditions. [1-4]Originally, these devices were used mostly in patients with cancer-related pain. [5,6]Now, because of implantation in patients with nonmalignant pain and a longer life expectancy, new complications are seen. These include catheter or pump problems, infections, intrathecal granuloma, lower extremity paralysis, accidental overdose, postdural puncture headache, amenorrhea, arthralgia, decreased libido, chronic headache, and others. [7]We report a case of self-administration of illicit substances into an implanted intrathecal pump.

A 43-yr-old man had a history of lower back pain since age 30 when he underwent the first instance of back surgery-a one-level discectomy and fusion. He underwent a total of five lumbar operations, including decompression with intertransverse process fusions at L5 and S1, pedicle screws with fixation plates, and ultimately hardware removal at age 38. At age 40, he underwent implantation of a Medtronic (Medtronic Neurological, Minneapolis, MN) intrathecal pump for the delivery of morphine to treat chronic back pain. This was implanted at an outside institution, and according to records we obtained, he was psychologically evaluated and thought to be an appropriate candidate for an intrathecal pump device.

The patient was incarcerated in the Texas Criminal Justice System at age 42 and was cared for in the Texas Department of Criminal Justice Hospital System. Our decision was to maintain intrathecal morphine because the pump was already in place. The patient complained of marginal pain relief at every visit, and over the course of 6 months his dose was increased from 2.5 mg/day to 5.0 mg/day. On at least two occasions before the incident described herein, the aspirated amount from the patient's pump was 2 to 4 ml more than the anticipated amount. This led to a suspicion of catheter blockage or pump dysfunction. The patency of the catheter into the intrathecal space was confirmed during fluoroscopy approximately 6 months before this incident. At that same time, pump function was confirmed by filling the pump with saline and using a code supplied by the manufacturer to run the pump at a rapid rate while visually inspecting the pump mechanism during fluoroscopy.

The patient was seen in the pain clinic for a pump refill. A cloudy liquid (8 ml) was aspirated from the pump. The expected volume was less than 2 ml, according to data from the pump-programming computer. The fluid was sent for analysis, and the results were positive for the presence of opiates (as expected), phencyclidine, methamphetamine, and propoxyphene.

The patient was confronted with the results, but denied self-administration of medication through the pump. During the ensuing coversation, the patient stated the only drug available in his prison unit was heroin; therefore, he could not possibly have injected this medication into his pump. A detailed neurologic examination of the patient revealed no abnormalities. The patient denied having a history of drug or alcohol abuse. He consented to a psychologic evaluation at this time, including a Minnesota Multiphasic Personality Inventory evaluation. It revealed a significant affective component to his pain.

For the patient's safety, he consented to and underwent pump and catheter removal. The pump appeared quite scored near the refill aperture, presumably because of the patient's access attempts. We sent the pump to Medtronic for analysis; it was noted to be undamaged and functioning normally. The manufacturer has seen other cases of suspected patient tampering in which the pump was not functioning because of septum damage (personal communication, Andrea Hutton-Lau, March 3, 1998). We are maintaining propoxyphene treatment dispensed daily and have referred the patient for substance abuse counseling.

Intrathecal drug administration by infusion is used increasingly as an option in the treatment of nonmalignant chronic pain. [1-4]New complications are being seen, especially in this patient population with chronic pain. [7] 

Schuchard et al. [7]reported that early complications often were catheter related, with 5 of 50 patients requiring reoperation because of a kink in the intrathecal catheter. The same authors reported pump flipping in 2 of 50 patients. Medical device adverse events reported to the Food and Drug Administration regarding implantable pumps (of all types, including intrathecal) have increased from 28 reports during 1984-1989 to 537 during 1992-1994. Of those 537 reports during the 1992-1994 interval, 11% were high-output errors, 11% were no-output errors, 9% were caused by leakage, and 9% were other malfunctions. In this report, chemotherapy pumps and intrathecal pumps were classified in the same category; therefore, it is difficult to interpret these data. [8]Schuchard et al. [7]also reported a 4% incidence of superficial infection and a 6% incidence of deep infection necessitating pump removal. A case of meningitis has also been reported. [9] 

Intrathecal granuloma formation around the tip of the catheter has been reported in six patients. [9-12]This granuloma is a mass of inflammatory tissue that develops around the catheter tip. Cultures for four of six patients cases were negative. These patients had an insidious increase in back pain and varying degrees of neurologic deficits. The time of granuloma development varied from 2 months after catheter implantation to 26 months (mean, 17 months). Five of six patients were administered morphine (10-19 mg/day), whereas one was administered hydromorphone (10 mg/day). A high index of suspicion for intrathecal granuloma must be maintained when a patient with an intrathecal catheter reports a change in the severity of back pain unrelated to injury, especially if accompanied by new neurologic deficit. The optimal imaging to rule out intrathecal granuloma is gadolinium-enhanced magnetic resonance imaging, because in one patient, the mass did not show up during magnetic resonance imaging without gadolinium. [7] 

One case of permanent paraplegia has been reported. [13]The patient had an intrathecal catheter and a spinal cord stimulator (epidural space) in place and had acute myelophaty at the level of the intrathecal catheter. At the time of emergent exploration, the spinal cord near the catheter tip appeared to be necrotic and liquefied. During pathologic evaluation, arachnoidal fibrosis and meningothelial hyperplasia was seen. All cultures were negative. The paraplegia persisted after exploratory open laminectomy and removal of both the catheter and the spinal cord stimulator. The cause of the paraplegia is unclear.

Accidental massive intrathecal morphine overdose has been reported in two patients. [14,15]In one patient the dose administered was 250 mg, whereas in the other a 450-mg bolus of intrathecal morphine was administered. In one patient, the catheter port was injected directly instead of in the pump refill port. The possibility of this error occurring again has been minimized by a manufacturer modification to the newer intrathecal pumps. Newer intrathecal pumps have a screen over the direct intrathecal catheter access port that limits needle size to 25 gauge or smaller. Both of these patients had severe seizures that were difficult to treat and hemodynamic instability, including hypotension and hypertension.

Postdural puncture headache has been reported after intrathecal catheter placement. Crul and Delhaas [16]reported an incidence of 7% of postdural puncture headache necessitating blood patch after intrathecal catheter placement.

Illicit substance abuse via an implanted intrathecal pump is a previously unreported complication. Although it has not been reported in the literature, Medtronic has seen other cases in which pump tampering has been suspected (personal communication, Andrea Hutton-Lau, March 3, 1998). Various histophatologic and neurotoxic side effects can occur with administration of various intrathecal substances. In an animal study, motor weakness and axonal changes associated with neurotoxicity were seen with low- and high-dose butorphanol administration. [17]The same study revealed motor weakness and mild inflammatory histophatologic changes with high-dose sufentanil only, whereas both low- and high-dose nalbuphine were shown to be relatively free of side effects. Karpinsky et al. [18]reported administering intrathecal ketamine by infusion to a cancer patient with good control of intratable neuropathic pain. However, motor weakness developed in the patient, and at autopsy there was evidence of neurotoxicity in the lumbosacral nerve roots (subpial vacuolar myelopathy). Amazingly, our patient suffered no apparent neurotoxic or infectious sequelae from grinding up of oral medications and self-administration into the pump reservoir.

This patient was seen 6 months before our discovery of the illicit use of his pump and was misdiagnosed with a pump malfunction. We reviewed an algorithm (Figure 1) for checking apparent pump malfunction. [19]Now added to the list of possibilities is “patient tampering with device” based on this case report. During pump refill, we recommend sending samples for analysis whenever the amount aspirated is significantly higher than expected.

Figure 1. Recommended algorithm for checking intrathecal pump dysfunction. Modified and reprinted with permission. [19] 

Figure 1. Recommended algorithm for checking intrathecal pump dysfunction. Modified and reprinted with permission. [19] 

Close modal
1.
Winkelmuller M, Winkelmuller W: Long-term effects of continuous intrathecal opioid treatment in chronic pain of nonmalignant etiology. J Neurosurg 1996; 85:458-67
2.
de Lissovoy G, Brown RE, Halpern M, Hassenbusch SJ, Ross E: Cost effectiveness of long-term intrathecal morphine therapy for pain associated with failed back surgery syndrome. Clin Ther 1997; 19:96-112
3.
Onofrio BM, Yaksh TL: Long-term pain relief produced by intrathecal morphine infusion in 53 patients. J Neurosurg 1990; 72:200-9
4.
Becker WJ, Ablett DP, Harris CJ, Dold ON: Long term treatment of intractable reflex sympathetic dystrophy with intrathecal morphine. Can J Neurol Sci 1995; 22:153-9
5.
Cousins MJ, Mather LE: Intrathecal and epidural administration of opioids. Anesthesiology 1984; 61:276-310
6.
Krames ES, Gershow J, Glassberg A, Kenefick T, Lyons A, Taylor P, Wilkie D: Continuous infusion of spinally administered narcotics for the relief of pain due to malignant disorders. Cancer 1985; 56:696-702
7.
Schuchard M, Krames ES, Lanning R: Intraspinal analgesia for nonmalignant pain. A retrospective analysis for efficacy, safety, and feasibility in 50 patients. Neuromodulation 1998; 1:46-56
8.
Brown SL, Morrison AE, Parmentier CM, Woo EK: Infusion pump adverse events: Experience from medical device reports. J Intravenous Nurs 1997; 20:41-9
9.
Bennett MI, Tai YM, Symonds JM: Staphylococcal Meningitis following Synchromed intrathecal pump implant: A case report. Pain 1994; 56:243-4
10.
Bejjani GK, Karim NO, Tzortzidis F: Intrathecal granuloma after implantation of a morphine pump: Case report and review of the literature. Surg Neurol 1997; 48:288-91
11.
Blount JP, Remley KB, Yue SK, Erickson DL: Intrathecal granuloma complicating chronic spinal infusion of morphine. Report of three cases. J Neurosurg 1996; 84:272-6
12.
North RB, Cutchis PN, Epstein JA, Long DM: Spinal cord compression complicating subarachnoid infusion of morphine: Case report and laboratory experience. Neurosurgery 1991; 29:778-84
13.
Aldrete JA, Vascello LA, Ghaly R, Tomlin D: Paraplegia in a patient with an intrathecal catheter and a spinal cord stimulator. Anesthesiology 1994; 81:1542-5
14.
Groudine SB, Cresanti-Daknis C, Lumb PD: Successful treatment of a massive intrathecal morphine overdose. Anesthesiology 1995; 82:292-5
15.
Sauter K, Kaufman HH, Bloomfield SM, Cline S, Banks D: Treatment of high-dose intrathecal morphine overdose. J Neurosurg 1994; 81:143-6
16.
Crul BJ, Delhaas EM: Technical complications during long-term subarachnoid or epidural administration of morphine in terminally ill cancer patients: A review of 140 cases. Reg Anesth 1991; 16:209-13
17.
Rawal N, Nuutinen L, Raj PP, Lovering SL, Gobuty AH, Hargardine J, Lehmkuhl J, Herva R, Abouleish El: Behavioral and histophatologic effects following intrathecal administration of butorphanol, sufentanil, and nalbuphine in sheep. Anesthesiology 1991; 75:1025-34
18.
Karpinski N, Dunn J, Hansen L, Masliah E: Subpial vacuolar myelopathy after intrathecal ketamine: Report of a case. Pain 1997; 73:103-5
19.
Schurch B: Errors and limitations of the multimodality checking methods of defective spinal intrathecal pump systems. Case report. Paraplegia 1993; 31:611-5