Editor’s Perspective
What We Already Know about This Topic

To date, there are not clinically practical, accurate, and precise noninvasive methods for measuring cardiac output in small children

What This Article Tells Us That Is New

This study describes a noninvasive method by which ultrasound can be used in small children to determine cardiac output with good precision

After surgery in 43 small children for repair of atrial or ventricular septal defects, cardiac output measurements performed using saline bolus injections and ultrasound detection of the expected blood dilution showed similar precision for measuring cardiac output as a cardiac outputs measured using periaortic flow probe

Background

Technology for cardiac output (CO) and blood volume measurements has been developed based on blood dilution with a small bolus of physiologic body temperature saline, which, after transcardiopulmonary mixing, is detected with ultrasound sensors attached to an extracorporeal arteriovenous loop using existing central venous and peripheral arterial catheters. This study aims to compare the precision and agreement of this technology to measure cardiac output with a reference method, a perivascular flow probe placed around the aorta, in young children. The null hypothesis is that the methods are equivalent in precision, and there is no bias in the cardiac output measurements.

Methods

Forty-three children scheduled for cardiac surgery were included in this prospective single-center comparison study. After corrective cardiac surgery, five consecutive repeated cardiac output measurements were performed simultaneously by both methods.

Results

A total of 215 cardiac output measurements were compared in 43 children. The mean age of the children was 354 days (range, 30 to 1,303 days), and the mean weight was 7.1 kg (range, 2.7 to 13.6 kg). The precision assessed as two times the coefficient of error was 3.6% for the ultrasound method and 5.0% for the flow probe. Bias (mean COultrasound 1.28 l/min − mean COflow probe 1.20 l/min) was 0.08 l/min, limits of agreement was ±0.32 l/min, and the percentage error was 26.6%.

Conclusions

The technology to measure cardiac output with ultrasound detection of blood dilution after a bolus injection of saline yields comparable precision as cardiac output measurements by a periaortic flow probe. The difference in accuracy in the measured cardiac output between the methods can be explained by the coronary blood flow, which is excluded in the cardiac output measurements by the periaortic flow probe.

The clinical examination of a severely hemodynamically unstable child has obvious limitations and may give unreliable estimates of the overall hemodynamic status. The cause of arterial desaturation, degree of hypovolemia, and systemic hypoperfusion are parameters that may be particular difficult to evaluate because of compensatory mechanisms.1–5  Easily adapted and safe monitoring devices with good reliability and reproducibility are needed to address these parameters, especially in children.6–8 

An available invasive cardiac output (CO) measuring method, such as thermodilution technique, although applicable for adults, has size restrictions for use in young children and may cause a substantial risk of complications (bleeding, thrombosis, embolism, and arrythmia). Noninvasive methods are less accurate and precise in measuring CO in small children (electrical bioimpedance, carbon dioxide rebreathing, and Doppler methods), require high level of training (echocardiography), or are clinically impractical and time consuming (magnetic resonance imaging).9,10 

A technology that uses ultrasound detection of blood dilution by means of a saline bolus injection and an extracorporeal arteriovenous loop between existing central venous and peripheral arterial lines has been developed for hemodynamic assessment in young children and neonates.11  It is minimally invasive, because it uses existing catheters and does not require additional invasive procedures. It measures total CO, including coronary blood flow.

The hypothesis of the study is that blood dilution detected by ultrasound is comparable with the reference method, perivascular flow probe around aorta to determine cardiac output. The aim of the study was to estimate the precision of this technology in young children compared with CO measured by a perivascular flow probe around the aorta (reference method) through five simultaneously replicated CO measurements and to assess the agreement by quantifying the difference (bias) between the two methods.

Study Design and Subjects

Forty-seven children undergoing elective cardiac surgery for correction of atrial septal defect and/or ventricular septal defect were enrolled in the study. Inclusion criteria were informed written parental consent, weight of less than 15 kg, and atrial septal and/or ventricular septal defects. Because lack of repeatability of the CO measurement can interfere with the comparison of the two methods, it was important that the CO measurement and the stroke volume were fairly constant during the measurement. Exclusion criteria were, therefore, shunts (undiagnosed extracardiac or significant residual shunt after the surgical correction), perioperative arrythmias (supraventricular, nodal tachycardia, and atrioventricular heart block), and/or significant valvular regurgitations (aortic, mitral, tricuspid, and pulmonary valvular leaks). This study was approved and registered by the Ethics Committee of Lund University, Lund, Sweden (Dnr 2013636).

Cardiac Output Measurements

Calculation of CO by Use of Saline Blood Dilution Detected by Ultrasound Sensors (COUD).

The technology is based on the premise that the ultrasound velocity of blood changes linearly with the level of blood dilution caused by injection of a specified bolus volume of body-temperature isotonic saline. The ultrasound velocity in blood is 1,560 to 1,585 m/s and decreases toward the ultrasound velocity of saline (1,530 m/s) after a bolus injection of saline. The device, developed by Transonic Systems Inc. (USA), uses an extracorporeal arteriovenous loop connected between existing arterial and central venous lines. The loop is connected to an external roller pump, which maintains a stable blood flow in the loop of 10 to 12 ml/min. The loop contains specific venous and arterial segments to which external ultrasound sensors fit. The sensors measure the ultrasound velocity and the blood flow at the out-flow and in-flow parts of the loop circuit.

A measurement session begins by entering the patient’s weight, length, arterial blood pressure, central venous pressure, and heart rate into the device. The connection stopcock to the out- and in-flow segments of the arteriovenous loop, primed with 5 ml of body-temperature, heparinized isotonic saline, is opened. The roller pump starts, and a small amount of body-temperature physiologic isotonic saline (0.5 to 1.0 ml/kg) is injected at the out-flow segment of the loop on the venous side before the venous ultrasound sensor. The saline is warmed to 37°C body temperature by a bag warmer that is connected to the device. The volume and time of the injected saline is determined by the venous ultrasound sensor. The saline is completely mixed as the blood passes through the cardiopulmonary circulation and gives rise to a homogenous blood dilution on the arterial side. The final blood dilution that occurs in the systemic arterial circulation is detected by the arterial ultrasound sensor at the arterial in-flow segment of the loop, and an ultrasound velocity curve is generated (fig. 1).

Fig. 1.

Schematics (A) and monitor display (B) of the tested cardiac output (CO) measurement device. The y axis C (%) represents the percentage concentration of saline in the arterial blood while the x axis is time (seconds). ACVI, active circulation volume index; BSA, body surface area; CBVI, central blood volume index; CI, cardiac index; CVP, central venous pressure; HR, heart rate; SVI, stroke volume index; SVRI, systemic vascular resistance index; TEDVI, total end diastolic volume index; TEF, total ejection fraction.

Fig. 1.

Schematics (A) and monitor display (B) of the tested cardiac output (CO) measurement device. The y axis C (%) represents the percentage concentration of saline in the arterial blood while the x axis is time (seconds). ACVI, active circulation volume index; BSA, body surface area; CBVI, central blood volume index; CI, cardiac index; CVP, central venous pressure; HR, heart rate; SVI, stroke volume index; SVRI, systemic vascular resistance index; TEDVI, total end diastolic volume index; TEF, total ejection fraction.

Close modal

Because the technology records the ultrasound velocity simultaneously at both the out-flow and in-flow segments of the loop at a constant blood flow rate, not only can the area under the curve be analyzed, but the time of occurrence and form of the dilution curve after it passes through the lungs and heart can be used to calculate total end-diastolic cardiac volume, central blood volume, and active blood volume and to determine and detect cardiac shunts.12–17  CO is calculated by analyzing the area under curve based on the Stewart–Hamilton indicator dilution principle.18–20 

CO Measurement with Perivascular Flow Probes (COPVFP).

AU-series confidence perivascular flow probes (Transonic Systems Inc.) were used in this study. The flow probe is custom-designed to fit around vessels to measure blood flow in real time by ultrasound transit-time technology. Transit-time technology uses four crystals and wide-beam illumination to send ultrasonic signals back and forth across the vessel, alternately intersecting the blood in upstream and downstream directions. The flowmeter derives an accurate measure of the “transit time” it takes for the wave of ultrasound to travel from one transducer to the other. The difference between the upstream and downstream integrated transit times is a measure of true volume flow, not velocity.

The flow probes are available in different diameter sizes (8 to 24 mm) and can be used multiple times because they can undergo standard sterilization. The probes come ready to use and calibrated from the manufacturer with a certified length of use for more than 1 yr.

In our study, the flow probe was applied to the aorta approximately 1 cm distal to the origin of the coronary arteries (fig. 2). The flow probe was then connected to an Optima dual-channel HT363 Flow-QC meter (Transonic Systems Inc.). AureFlo diagnostic software (Transonic Systems Inc.) was used to visualize a good signal of pulsating aortic blood flow waveform and record CO. Transit-time ultrasound perivascular flow probes are considered the standard reference method for cardiac output estimation and have been verified in number of studies.21,22 

Fig. 2.

Placement of the perivascular flow probe around aorta.

Fig. 2.

Placement of the perivascular flow probe around aorta.

Close modal

Experimental Protocol

Anesthesia was induced using fentanyl (5 μg/kg) and penthothal (5 mg/kg) and maintained with isoflurane (0.5 to 1.0%). Pancuronium (0.2 mg/kg) was given to facilitate intubation with a cuffed endotracheal tube. As is routine in children undergoing cardiac surgery, all subjects had a peripheral arterial catheter placed in the radial artery and a central venous catheter placed in the right internal jugular vein. The catheters were connected to the arteriovenous loop of the CO device and were primed with heparinized (2 units/ml) 37°C physiologic saline. Ultrasound sensors were placed on the venous and arterial segments of the arteriovenous loop before surgery. After surgical correction and weaning from cardiopulmonary bypass, transesophageal echocardiography was performed to exclude residual intracardiac shunts or valve regurgitations. When a stable sinus rhythm and normal body temperature were achieved, the surgeons applied the perivascular flow probe around the aorta, and measurements were initiated. Each measurement session consisted of five consecutively repeated CO measurements with injections of body temperature physiologic saline boluses and, simultaneously, five readings of aortic blood flow measured with the transit-time ultrasound periaortic flow probe.

Statistical Analysis

Statistical analysis was performed using Statistica version 12 (Statsoft, USA). No statistical power calculation was conducted before the study, because the bias and SD of the bias between the two methods were unknown, and the CI for the 95% limits of agreement was not possible to estimate. The sample size was based on previous experience with this design. All data are expressed as mean ± SD unless indicated otherwise.

The degree of variation of each technique was presented as the coefficient of error (CE) of average repeated measurements, calculated as ratio of the coefficient of variation (CV) divided with the square root of the number (n) of repeated measurements (CE = CV/√n). Precision of a technique is considered to be two times the coefficient of error as suggested by Cecconi et al.23 

Bland–Altman analysis was used to estimate bias between the different techniques while accounting for the repeated measurements within each individual.24  The mean difference (bias) between cardiac output with saline dilution and ultrasound detection (COUD) minus cardiac output with perivascular flow probe around aorta (COPVFP) was calculated and plotted against the average of the comparison (COUD + COPVFP)/2. The 95% limits of agreement were calculated as mean bias ± 1.96 × SD (SD of the bias). Limit of agreement analysis was performed to determine whether the two methods agreed sufficiently with each other so that one could replace the other. The 95% CI of the bias and the limits of agreement were determined after testing for normal distribution using Levene’s test.25 

According to Critchley and Critchley,26  the percentage error (PE) was calculated as 1.96 × SD of the bias/mean cardiac output of the reference method × 100%.

formula

A total of 47 children were enrolled in the study. Four children were excluded before surgery: one because of hemodynamic instability; another because an ultrasound sensor came loose underneath the surgical drape; and two because the internal calibration date for the tested device had expired, and the monitor refused to accept data. This resulted in 43 children being included in the study, in which a total of 215 paired COUD and COPVFP measurements were performed (table 1). Mean age was 356 days (range, 30 to 1,303 days); mean weight was 7.1 kg (range, 2.7 to 13.6 kg); and mean body surface area was 0.36 m2 (range, 0.18 to 0.59 m2). There were no missing data from the included children.

Table 1.

Simultaneously Measured Mean Cardiac Outputs by COUD and COPVFP (n = 5), Heart Rates, Blood Pressures, Central Venous Pressures, and Ejection Fractions in All 43 Children after Surgical Correction

Simultaneously Measured Mean Cardiac Outputs by COUD and COPVFP (n = 5), Heart Rates, Blood Pressures, Central Venous Pressures, and Ejection Fractions in All 43 Children after Surgical Correction
Simultaneously Measured Mean Cardiac Outputs by COUD and COPVFP (n = 5), Heart Rates, Blood Pressures, Central Venous Pressures, and Ejection Fractions in All 43 Children after Surgical Correction

The mean COUD was 1.28 l/min (range, 0.46 to 2.98 l/min) and COPVFP 1.20 l/min (range, 0.42 to 2.70 l/min). Means and SDs were normally distributed. The coefficient of error for repeated COUD measurements was 1.8%, resulting in precision of 3.6% for COUD. Coefficient of error for COPVFP was 2.5%, resulting in precision of 5.0% for COPVFP.

Bland–Altman analysis (fig. 3) showed that the bias between COUD and COPVFP was 0.08 l/min (95% CI, 0.05 to 0.10), and the limits of agreement were −0.24 l/min (95% CI, −0.17 to −0.32) and 0.40 l/min (95% CI, 0.33 to 0.47). The percentage error between COUD and COPVFP was 26.6%.

Fig. 3.

Bland–Altman plot comparing different methods for cardiac output (CO) measurement, saline dilution and ultrasound detection (COUD), and perivascular flow probe (COPVFP). LOA, limits of agreement.

Fig. 3.

Bland–Altman plot comparing different methods for cardiac output (CO) measurement, saline dilution and ultrasound detection (COUD), and perivascular flow probe (COPVFP). LOA, limits of agreement.

Close modal

Three patients had insignificant residual shunts after the surgical correction. No valvular regurgitations were observed that affected the results. There were no adverse effects in any patient related to measurements.

In the present study, we found that repeated CO measurements by the tested method, which uses saline bolus injections and ultrasound detection of the expected blood dilution (COUD), showed a similar precision as our reference method, a perivascular flow probe around the aorta (COPVFP). We believe that CO measurement with a periaortic flow probe is the most accurate technique to be used for comparison studies using repeated CO measurements in vivo. There was a small but significant bias of the measured CO between COUD and COPVFP according to the Bland–Altman analysis of 0.08 l/min. This finding of a lower COPVFP compared with COUD is consistent with a coronary blood flow of approximately 4 to 10% of the measured CO, which agrees with suggestions from earlier publications.27 

Earlier studies using perivascular flow probes in animals have applied the flow probe around the pulmonary artery as a measurement of total CO, which is also measured by COUD. This had been our intention, but we discovered that the peripulmonary flow probe tended to compress the right coronary vessels, resulting in a reduction of CO. This led us to abandon its use for the simultaneously repeated CO measurements and rely solely on the blood flow probe around the aorta, although the total CO missed the drain by the coronary blood flow.

There will always be a physiologic variability in CO caused by ventilation and cardiac filling, which result in various degree of imprecision during the CO measurements. The COPVFP analysis recorded beat-to-beat fluctuations in CO caused by the ventilation, changes in cardiac filling, and changes in coronary blood flow. This can be the reason for a slightly but not significantly higher coefficient of error (2.5%) compared with CO measured with COUD (1.8%). In addition, the dilution curve, which is used by the COUD to calculate CO, reflects a mean of several heart beats, which may give a more stable value. We suspected that five consecutive repeated injections of 0.5 to 1.0 ml/kg of saline could decrease the precision of the COUD analysis, because of a potential dilutional effect on the blood by multiple saline injections, but this did not affect the precision of our analysis. Because we conducted a trial with a high number (n = 5) of repeated measurements and did not remove any measurements, we believe that our analysis is accurate and indicates that the dilution of the blood with bolus doses of saline and detection by ultrasound sensors in a constant extracorporeal arteriovenous loop flow gives stable CO measurements compared with, for example, thermodilution, which still is regarded to be the gold standard for CO measurement in children and a reference method in validation studies.28,29 

Although our analysis was negatively influenced by the bias caused by the coronary blood flow between COUD and COPVFP, our results show a percentage error of 26.6%, which is less than the 30% limit, that has been concluded by Critchley and Crithchley26  as an acceptable limit of a new technique to be equivalent to the reference method.

One limitation to the tested technology is that the arterial blood pressure monitoring has to be closed during the CO measurement. The device, in its present form, only allows for intermittent but not continuous CO measurements as would be preferable for a monitoring device. In addition, one central venous line is occupied for the injection and the loop circulation during the measurement. However, there is no patient blood loss because all blood in the arteriovenous loop is flushed back into the circulation after measurements. Only small-volume (0.5 to 1.0 ml/kg) body-temperature saline boluses are needed to do the measurements, and there is no drop in heart rate as often is seen during measurements with a thermodilution catheter because of cold saline boluses.

In summary, the tested technology, which uses a saline bolus into the venous side and ultrasound sensors on an extracorporeal loop to detect the blood dilution both on the venous and arterial side for calculation of CO, has a precision comparable with the precision obtained by continuous CO measurement by a periaortic flow probe in young children. A Bland–Altman plot shows that the tested method detects a small expected bias between COUD and COPVFP caused by the coronary blood flow. It is easy to apply clinically in children with arterial and central venous catheters in place. The technology could potentially be a promising alternative as a reference method for comparison studies of CO in young children.

The authors thank their pediatric cardiac surgical team at the Children’s Hospital, University Hospital of Lund, Lund, Sweden, for help, patience, and understanding; and Nikolai Krivitski, Ph.D., DSc, Transonic, Ithaca, New York, for technical assistance.

Supported by project grants from the Swedish Children Heart Association, Stockholm, Sweden, and the Anna and Edwin Bergen Foundation, Lidingö, Sweden.

The authors declare no competing interests.

1.
Lemson
J
,
Nusmeier
A
,
van der Hoeven
JG
:
Advanced hemodynamic monitoring in critically ill children.
Pediatrics
2011
;
128
:
560
71
2.
Tibby
SM
,
Hatherill
M
,
Marsh
MJ
,
Murdoch
IA
:
Clinicians’ abilities to estimate cardiac index in ventilated children and infants.
Arch Dis Child
1997
;
77
:
516
8
3.
Egan
JR
,
Festa
M
,
Cole
AD
,
Nunn
GR
,
Gillis
J
,
Winlaw
DS
:
Clinical assessment of cardiac performance in infants and children following cardiac surgery.
Intensive Care Med
2005
;
31
:
568
73
4.
Skowno
JJ
,
Broadhead
M
:
Cardiac output measurement in pediatric anesthesia.
Paediatr Anaesth
2008
;
18
:
1019
28
5.
de Boode
WP
:
Cardiac output monitoring in newborns.
Early Hum Dev
2010
;
86
:
143
8
6.
Shephard
JN
,
Brecker
SJ
,
Evans
TW
:
Bedside assessment of myocardial performance in the critically ill.
Intensive Care Med
1994
;
20
:
513
21
7.
Le Manach
Y
,
Collins
GS
:
Disagreement between cardiac output measurement devices: Which device is the gold standard?
Br J Anaesth
2016
;
116
:
451
3
8.
Suehiro
K
,
Joosten
A
,
Murphy
LS
,
Desebbe
O
,
Alexander
B
,
Kim
SH
,
Cannesson
M
:
Accuracy and precision of minimally-invasive cardiac output monitoring in children: A systematic review and meta-analysis.
J Clin Monit Comput
2016
;
30
:
603
20
9.
Darling
E
,
Thuramalla
N
,
Searles
B
:
Validation of cardiac output measurement by ultrasound dilution technique with pulmonary artery thermodilution in a pediatric animal model.
Pediatr Cardiol
2011
;
32
:
585
9
10.
Peyton
PJ
,
Chong
SW
:
Minimally invasive measurement of cardiac output during surgery and critical care: A meta-analysis of accuracy and precision.
Anesthesiology
2010
;
113
:
1220
35
11.
Krivitski
NM
,
Kislukhin
VV
,
Thuramalla
NV
:
Theory and in vitro validation of a new extracorporeal arteriovenous loop approach for hemodynamic assessment in pediatric and neonatal intensive care unit patients.
Pediatr Crit Care Med
2008
;
9
:
423
8
12.
Crittendon
I
3rd
,
Dreyer
WJ
,
Decker
JA
,
Kim
JJ
:
Ultrasound dilution: An accurate means of determining cardiac output in children.
Pediatr Crit Care Med
2012
;
13
:
42
6
13.
Vrancken
SL
,
de Boode
WP
,
Hopman
JC
,
Looijen-Salamon
MG
,
Liem
KD
,
van Heijst
AF
:
Influence of lung injury on cardiac output measurement using transpulmonary ultrasound dilution: A validation study in neonatal lambs.
Br J Anaesth
2012
;
109
:
870
8
14.
Lindberg
L
,
Johansson
S
,
Perez-de-Sa
V
:
Validation of an ultrasound dilution technology for cardiac output measurement and shunt detection in infants and children.
Pediatr Crit Care Med
2014
;
15
:
139
47
15.
Saxena
R
,
Krivitski
N
,
Peacock
K
,
Durward
A
,
Simpson
JM
,
Tibby
SM
:
Accuracy of the transpulmonary ultrasound dilution method for detection of small anatomic shunts.
J Clin Monit Comput
2015
;
29
:
407
14
16.
Vrancken
SL
,
de Boode
WP
,
Hopman
JC
,
Singh
SK
,
Liem
KD
,
van Heijst
AF
:
Cardiac output measurement with transpulmonary ultrasound dilution is feasible in the presence of a left-to-right shunt: A validation study in lambs.
Br J Anaesth
2012
;
108
:
409
16
17.
Vrancken
SL
,
van Heijst
AF
,
Hopman
JC
,
Liem
KD
,
van der Hoeven
JG
,
de Boode
WP
:
Hemodynamic volumetry using transpulmonary ultrasound dilution (TPUD) technology in a neonatal animal model.
J Clin Monit Comput
2015
;
29
:
643
52
18.
Stewart
GN
:
Researches on the circulation time and on the influences which affect it.
J Physiol
1897
;
22
:
159
83
19.
Hamilton
WF
,
Moore
JW
,
Kinsman
JM
,
Spurling
RG
:
Studies on the circulation.
Am J Physiol
1932
;
99
:
534
51
20.
Reuter
DA
,
Huang
C
,
Edrich
T
,
Shernan
SK
,
Eltzschig
HK
:
Cardiac output monitoring using indicator-dilution techniques: Basics, limits, and perspectives.
Anesth Analg
2010
;
110
:
799
811
21.
Dean
DA
,
Jia
CX
,
Cabreriza
SE
,
D’Alessandro
DA
,
Dickstein
ML
,
Sardo
MJ
,
Chalik
N
,
Spotnitz
HM
:
Validation study of a new transit time ultrasonic flow probe for continuous great vessel measurements.
ASAIO J
1996
;
42
:
M671
6
22.
Botero
M
,
Kirby
D
,
Lobato
EB
,
Staples
ED
,
Gravenstein
N
:
Measurement of cardiac output before and after cardiopulmonary bypass: Comparison among aortic transit-time ultrasound, thermodilution and noninvasive partial CO2 rebreathing.
J Cardiothorac Vasc Anesth
2004
;
18
:
563
72
23.
Cecconi
M
,
Rhodes
A
,
Poloniecki
J
,
Della Rocca
G
,
Grounds
RM
:
Bench-to-bedside review: The importance of the precision of the reference technique in method comparison studies–with specific reference to the measurement of cardiac output.
Crit Care
2009
;
13
:
201
24.
Bland
JM
,
Altman
DG
:
Statistical methods for assessing agreement between two methods of clinical measurement.
Lancet
1986
;
1
:
307
10
25.
Zou
GY
:
Confidence interval estimation for the Bland–Altman limits of agreement with multiple observations per individual.
Stat Methods Med Res
2013
;
22
:
630
42
26.
Critchley
LA
,
Critchley
JA
:
A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques.
J Clin Monit Comput
1999
;
15
:
85
91
27.
The circulation during exercise: The coronary circulation and ischemic heart disease
in
Textbook of Medical Physiology
, 9th edition. Edited by
Guyton
AC
,
Hall
JE
.
Philadelphia
,
W.B. Saunders Company
,
1996
, pp
253
63
28.
Stetz
CW
,
Miller
RG
,
Kelly
GE
,
Raffin
TA
:
Reliability of the thermodilution method in the determination of cardiac output in clinical practice.
Am Rev Respir Dis
1982
;
126
:
1001
4
29.
Tibby
S
:
Transpulmonary thermodilution: Finally, a gold standard for pediatric cardiac output measurement.
Pediatr Crit Care Med
2008
;
9
:
341
2