Background

Arterial oxygenation is often impaired during one-lung ventilation, due to both pulmonary shunt and atelectasis. The use of low tidal volume (VT) (5 ml/kg predicted body weight) in the context of a lung-protective approach exacerbates atelectasis. This study sought to determine the combined physiologic effects of positive end-expiratory pressure and low VT during one-lung ventilation.

Methods

Data from 41 patients studied during general anesthesia for thoracic surgery were collected and analyzed. Shunt fraction, high V/Q and respiratory mechanics were measured at positive end-expiratory pressure 0 cm H2O during bilateral lung ventilation and one-lung ventilation and, subsequently, during one-lung ventilation at 5 or 10 cm H2O of positive end-expiratory pressure. Shunt fraction and high V/Q were measured using variation of inspired oxygen fraction and measurement of respiratory gas concentration and arterial blood gas. The level of positive end-expiratory pressure was applied in random order and maintained for 15 min before measurements.

Results

During one-lung ventilation, increasing positive end-expiratory pressure from 0 cm H2O to 5 cm H2O and 10 cm H2O resulted in a shunt fraction decrease of 5% (0 to 11) and 11% (5 to 16), respectively (P < 0.001). The Pao2/Fio2 ratio increased significantly only at a positive end-expiratory pressure of 10 cm H2O (P < 0.001). Driving pressure decreased from 16 ± 3 cm H2O at a positive end-expiratory pressure of 0 cm H2O to 12 ± 3 cm H2O at a positive end-expiratory pressure of 10 cm H2O (P < 0.001). The high V/Q ratio did not change.

Conclusions

During low VT one-lung ventilation, high positive end-expiratory pressure levels improve pulmonary function without increasing high V/Q and reduce driving pressure.

What We Already Know about This Topic
  • Low tidal volume is necessary to prevent lung injury during one-lung anesthesia but can potentiate development of atelectasis. Although positive end-expiratory pressure can reduce atelectasis, it may cause hyperinflation and increased alveolar dead space.

What This Article Tells Us That Is New
  • Shunt fraction and ventilation/perfusion ratio ( ) distribution was measured in 41 patients undergoing one-lung anesthesia. Positive end-expiratory pressure at 5 or 10 cm H2O, applied in random order, resulted in lower shunt fraction and driving pressure, without increasing dead space.

ARTERIAL oxygenation is impaired during one-lung ventilation (OLV) in lateral decubitus due to the obligatory shunt through the nondependent lung.1,2  The generation of atelectasis in the dependent, ventilated lung further decreases oxygenation by reducing the aerated lung volume and inducing ventilation–perfusion mismatch.3,4  Applying a positive end expiratory pressure (PEEP) to the dependent lung could ameliorate intrapulmonary shunt.2,4–6  However, studies have shown conflicting results, with some showing sustained improvement5–8  and others showing no effects9  or even worsening of oxygenation.10,11  These conflicting results might be at least partially explained by the different interplay between PEEP and tidal volume (VT) in the different studies. Indeed, the previous quoted studies report the use of different VT during OLV, ranging between 6 and 10 ml/kg. High VT have per se the potential of decreasing shunt by recruiting the atelectatic lung areas, but this strategy may be deleterious, both causing lung injury12,13  and augmenting cytokine production.12,13  Thus, considerable attention has been paid to identify the correct VT to be used during OLV,13–16  and recent evidence suggests that a lung protective tidal volume of 4 to 5 ml/kg predicted body weight (PBW) should be applied during OLV.16 

Because atelectasis more likely occurs with low VT,17  the aim of our study was to investigate whether higher PEEP during low VT OLV can improve both oxygenation through reduction in shunt and lung mechanics through reduced driving pressure. Thus, we applied different PEEP levels (0, 5, and 10 cm H2O) and measured ventilation/perfusion matching and respiratory mechanics in patients undergoing thoracoscopic surgery ventilated with a VT of 4 to 5 ml/kg PBW during OLV. We therefore sought to describe the physiologic effects of increased PEEP.

This study was performed in the Department of Anesthesia and Intensive Care at the University Hospital of Ferrara (Italy) from January to November 2016. Our trial was approved by the Ethics Committee of our institution (protocol No. 140495) and registered in Clinicaltrial.gov (trial No. NCT02968550). Written informed consent was obtained from each patient before surgery.

Patients scheduled for elective lobectomy or lung resection through video-assisted thoracoscopic surgery requiring OLV and lateral position were enrolled if they were more than 18 yr of age and with an American Society of Anesthesiologists Physical Status I to III. Patients were excluded in case of hemodynamic instability (defined as a decrease in systolic arterial pressure of more than 20% from baseline), severe chronic respiratory failure (chronic obstructive pulmonary disease patients with Global Initiative for Obstructive Lung Disease stage 3 or 4,18  preoperative hemoglobin less than 10 g ml−1), or procedures requiring unplanned conversion to thoracotomy surgery or planned to be shorter than 30 min.

As a routine practice in our institution, patients underwent preoperative spirometry performed in sitting position according to the American Thoracic Society’s standards, using SpiroPro spirometer (SpiroPro; Jaeger, Germany). Spirometry measurement of vital capacity, forced expiratory volume in the 1st second, forced vital capacity (FVC), expiratory reserve volume (ERV), and transfer coefficient (KCO) was performed. Before anesthesia induction, a thoracic epidural catheter (Tuohy; Braun Laboratories, Germany) was placed between T3 and T6 and 3 ml of bupivacaine 0.25% was administered.

All patients breathed an inspiratory oxygen fraction (Fio2) of 0.8 during the induction of general anesthesia to maintain adequate oxygenation while reducing the risk of absorption atelectasis.19  Anesthesia was induced with propofol (1.5 to 2 mg/kg), fentanyl (3 μg/kg), and rocuronium (0.6 mg/kg) to facilitate tracheal intubation. The trachea was intubated with an appropriately sized and side double lumen tube (Broncho-part; Rush, Germany). Tube position was confirmed by bronchoscopy in the supine and lateral positions. Anesthesia was maintained with a continuous infusion of propofol (150 to 200 μg·kg −1·min–1), remifentanil (0.1 to 0.2 μg·kg−1·min–1), and cis-atracurium (2 μg·kg−1·min–1). Balanced crystalloid solutions20  were infused at a rate of 3 ml·kg−1·h−1.

Patients were ventilated with a square flow waveform using Dräger Primus ventilator (Drägerwerk AG and Co. KGaA, Germany). During two-lung (bilateral) ventilation, VT was set to 6 to 8 ml/kg PBW and zero PEEP. During OLV, VT was reduced to 4 to 5 ml/kg PBW, and PEEP varied from 0 to 10 cm H2O, according to the experimental protocol (see Study Protocol). Fio2 was set to maintain peripheral oxygen saturation (SpO2) equal to or greater than 92%. The inspiratory-to-expiratory ratio was set to 1:2, and frequency was adjusted to maintain an arterial Paco2 between 40 and 60 mmHg.

Respiratory mechanics were assessed by the constant V´/rapid occlusion method previously described in detail.21  End-inspiratory occlusion was obtained by increasing end-inspiratory pause to 40%. Driving pressure (ΔP) was calculated as plateau pressure – PEEP, whereas respiratory system compliance (CRS) was calculated as VT/(end-inspiratory plateau pressure – PEEP).

Patients were monitored using a Dräger Infinity C700 monitor (Dräger Medical, Germany) with an electrocardiogram, pulse oximetry, end-tidal carbon dioxide, and continuous arterial pressure monitoring via a catheter inserted into the radial artery. The latter was placed under local anesthesia before induction of general anesthesia, in line with the standard practice of our institution, for invasive blood pressure and to obtain samples for blood gas monitoring. Analysis of arterial blood gases were performed within 3 min from sampling (Cobas 123 POC; Roche Diagnostics, Switzerland). Depth of anesthesia was monitored using bispectral index (Aspect A-2000; Aspect Medical System, USA).

Shunt and ventilation/perfusion ratio (V/Q) matching were assessed by the ALPE system (ALPE Integrated, Mermaid Care A/S, Denmark). To assess V/Q matching, the ALPE system instructs the user to modify Fio2 in three or four steps. At each Fio2 level, the ALPE system identifies steady state and measures ventilation, SpO2, oxygen consumption, CO2 production, and inspiratory and expiratory fractions of O2 and CO2. These measurements are taken automatically by inserting a sampling tube in the respiratory circuit for measurement of flow, O2, and CO2 and by placing the pulse oximeter on a finger. In addition, the system estimates the acid–base and oxygenation status including arterial Paco2, taking into account the results of an arterial blood gas sample. These parameters are then used to identify the fractions of ventilation and perfusion in a three-compartment model of the lung, including two ventilated and perfused compartments and a further perfused-only compartment, describing pulmonary shunt. The model takes into account also some extrapulmonary factors including acid–base status, hemoglobin concentration, the nonlinearity of hemoglobin oxygen binding, cardiac output, and the measured oxygen consumption. The system assumes a cardiac index of 3.7 l · min–1 · m–2, as previously reported in intensive care patients.22  Body surface area was calculated from height and weight as previously performed by Gehan and George.23  The estimation of ventilation and perfusion parameters is performed as follows. It is well known that variation in Fio2 can be used to identify shunt, with oxygenation problems at Fio2 = 1 being due to shunt alone. Because Fio2 values of 1 may increase the risk of absorption atelectasis24  and may therefore be undesirable, the ALPE algorithm applies the principle that in the case of true pulmonary shunt, SpO2 will change little when changing Fio2. This is in contrast to areas with low V/Q, where SpO2 will change greatly with Fio2. Accordingly, through variation of Fio2 in three or four steps, the system mathematically estimates shunt and low V/Q ratios. Further, the ALPE algorithm takes into account the end-tidal to arterial CO2 gradient to account for the part of this gradient due to shunt and low V/Q and the one due to high V/Q ratio. For ease of understanding, the estimates of ventilation and perfusion obtained from ALPE analysis are converted into indices describing low and high V/Q regions. Low V/Q mismatch is represented as the difference in O2 partial pressure between end-tidal gas and blood leaving lung capillaries in the low V/Q areas. As an example, a low V/Q index of 10 kPa indicates the need for an increase in Fio2 of approximately 10% to counter the effect of low V/Q on oxygenation of nonshunted blood. High V/Q mismatch is represented as an index constituting the difference in CO2 partial pressure between end-tidal gas and blood leaving lung capillaries. A high V/Q index of more than 0 kPa can be interpreted as insufficient removal of CO2 due to high V/Q. The ALPE technique has been validated and applied in varied patient populations.25–28 

Study Protocol

Measurements were made (1) before surgery, when patients were ventilated at zero PEEP during bilateral lung ventilation in the supine position (TLV baseline), and (2) during OLV in the lateral decubitus, after collapse of the nondependent lung. OLV (fig. 1) immediately opened the lumen of the endotracheal tube of the nonventilated lung to room air. After the assessment of shunt, respiratory mechanics, and gas exchange at ZEEP, we applied in random order 5 and 10 cm H2O of PEEP. Randomization was obtained by using a computer-generated number. Each level of PEEP was maintained for 15 min, allowing the effects of PEEP to reach an equilibrium.29  Parameters describing respiratory mechanics, hemodynamics, and gas exchange were measured at each PEEP step. The design of the study is summarized in figure 1.

Fig. 1.

Flowchart of the study. CPAP = continuous positive airway pressure; OLV = one-lung ventilation; PEEP = positive end-expiratory pressure; TLV = two-lung ventilation ZEEP = zero positive end-expiratory pressure.

Fig. 1.

Flowchart of the study. CPAP = continuous positive airway pressure; OLV = one-lung ventilation; PEEP = positive end-expiratory pressure; TLV = two-lung ventilation ZEEP = zero positive end-expiratory pressure.

Close modal

Statistical Analysis

Normal distribution of data was tested by the Shapiro–Wilk normality test. The data are reported as means ± SD or median (interquartile range) as appropriate. Differences between measurements at different PEEP levels were analyzed using repeated measure ANOVA or Friedman’s rank analysis for normally or not normally distributed variables, respectively. When multiple comparisons were made, p values were adjusted by the Bonferroni post hoc procedure. The treatment effect is expressed as mean difference and 95% CI or median difference (interquartile range). Pearson correlation with R square was used to analyze the correlation. Correlation strength was considered based on the absolute value of the r (0.20 to 0.39 “weak,” 0.40 to 0.59 “moderate,” and 0.60 to 0.79 “strong”).30  All the analyses performed for the primary and secondary outcomes were preplanned; furthermore, a post hoc, subgroup analysis was performed to describe the behavior of physiologic variables, after identification of a subgroup of patients where ΔP did not decrease with increased PEEP. Two-tailed statistical hypothesis testing was performed with P values of ≤ 0.05 considered statistically significant. Statistical analysis was performed with using SPSS Statistics for Windows, version 20.0 (IBM, USA). This is the primary analysis of these data.

Sample Size Calculation

An a priori sample size was calculated according to the primary endpoint: the improvement in shunt fraction by increasing PEEP levels in patients undergoing OLV in lateral decubitus. Based on at least 90% power, 40 patients were required to detect a mean difference in shunt fraction from 38 ± 5 to 34 ± 7% after the application of 5 cm H2O of PEEP using paired t tests with an α = 0.05. This is consistent with the observed difference in shunt fraction seen previously when investigating the effects of PEEP during OLV at a VT of 10 ml/kg.8  Finally, 50 patients were required to account for an anticipated dropout of 20% due to declining participation, interruption of intervention, and unplanned thoracotomic conversion. Sample size analysis was performed using MedCalc software (version 9.3.6.0; MedCalc, Belgium).

Study Population

Among the 50 patients assessed for eligibility, 41 completed the study (figure 1); their clinical and demographic characteristics are described in table 1. There were no missing data in the data set. The median shunt during TLV at zero PEEP was 19% (9 to 23) with a CRS of 36.2 ± 10 ml/cm H2O and a ΔP of 13 ± 4 cm H2O. The average shunt raised to 33% (27 to 45) during OLV at ZEEP, whereas CRS to 22 ± 5 ml/cm H2O and ΔP increased to 16 ± 3 cm H2O. Hemodynamic parameters did not change throughout the protocol, irrespective of the applied PEEP level (table 2).

Table 1.

Patient Characteristics

Patient Characteristics
Patient Characteristics
Table 2.

Intraoperative Variables

Intraoperative Variables
Intraoperative Variables

Effects of PEEP on Ventilation/Perfusion and Respiratory Mechanics (Table 2)

The median decrease in shunt fraction was 5% (0 to 11) at PEEP 5 cm H2O and 11% (5 to 16) at PEEP 10 cm H20 (P < 0.001), whereas the CRS increased by 3 ml/cm H2O (CI 1.4 to 4.6) at PEEP 5 cm H2O and 6.7 ml/cm H2O (CI 4.7 to 8.5) at PEEP 10 cm H2O (P < 0.001). Similarly, ΔP decreased from 16 ± 3 cm H2O to 14 ± 3 cm H2O at PEEP 5 and to 12 ± 3 cm H2O at PEEP 10; P < 0.001 (fig. 2). High V/Q ratio was not significantly different between TLV and OLV, regardless of the PEEP level (table 2). We found a tendency for high V/Q to increase at PEEP 10 cm H2O in those patients where ΔP increased with PEEP (Supplemental Digital Content 1, https://links.lww.com/ALN/B572). The Pao2/Fio2 ratio increased significantly only at PEEP 10 cm H2O compared to zero PEEP (281 (129 to 243) mmHg versus 142 (96 to 168) mmHg; P < 0.001).

Fig. 2.

Individual differences in driving pressure at different levels of positive end-expiratory pressure (PEEP). The horizontal dashed line shows the cutoff of 14 cm H2O. The continuous horizontal lines show the means in each group.

Fig. 2.

Individual differences in driving pressure at different levels of positive end-expiratory pressure (PEEP). The horizontal dashed line shows the cutoff of 14 cm H2O. The continuous horizontal lines show the means in each group.

Close modal

Predictors of Shunt Severity during OLV

There was a strong inverse correlation between ERV and the amount of shunt developed during OLV at ZEEP (r = −0.79; r2 = 0.62; P < 0.001) (fig. 3). A similar but weaker correlation was found at PEEP 5 (r = −0.72; r2=0.52; P < 0.001) (Supplemental Digital Content 2, https://links.lww.com/ALN/B573) and PEEP 10 (r = −0.58; r2 = 0.40; Supplemental Digital Content 3, https://links.lww.com/ALN/B574). Furthermore, there was a moderate correlation between KCO and shunt (r = −0.47; r2 = 0.23; P = 0.04) and a weak correlation between body mass index (r = 0.33; r2 = 0.12; P = 0.03) and shunt.

Fig. 3.

Correlation between expiratory reserve volume (ERV) and intrapulmonary shunt measured at zero positive end-expiratory pressure.

Fig. 3.

Correlation between expiratory reserve volume (ERV) and intrapulmonary shunt measured at zero positive end-expiratory pressure.

Close modal

The main finding of this study is that a PEEP of 10 cm H2O is needed to decrease the shunt fraction and the driving pressure while increasing oxygenation in patients ventilated with “protective” low VT during OLV. In patients undergoing general anesthesia and muscle paralysis, the decrease of FRC associated with the development of atelectasis impairs the matching of ventilation and perfusion. During OLV, the absence of ventilation in the nondependent lung and the atelectasis induced by anesthesia in the dependent lung result in further ventilation/perfusion mismatch and hypoxia. However, no conclusive data are available on the correct amount of PEEP that should be applied during OLV to ameliorate oxygenation. This probably reflects the fact that shunt is highly influenced by the ventilatory pattern and in particular by the interplay between VT and PEEP.15  The recent extension of the “lung protective ventilation” concept from the ARDS to the anesthesia field underlines the need for minimizing both atelectasis and overdistension,31  suggesting the use of low VT and a “adequate” PEEP levels. However, OLV might deserve even lower VT as compared to those recommended for protective ventilation during TLV.13,16  In thoracic surgery, a VT of 5 ml/kg was shown to decrease postoperative levels of tumor necrosis factor-α, interleukin-8, and interleukin-10 as compared to 10 ml/kg.13  Of note, in an animal study, a VT of 10 ml/kg compared to one of 5 ml/kg resulted in inhomogeneous distribution of aeration predisposing to postoperative lung injury.14  The role played by low VT during OLV is further supported by a study from Qutub et al.32  demonstrating higher extravascular lung water with a VT of 8 or even 6 ml/kg as compared to a VT of 4 ml/kg. Hence, as suggested by Lohser and Slinger,16  the adequate VT during OLV should be around 4 or 5 ml/kg PBW. However, the use of low VT may exacerbate the atelectasis in the dependent, ventilated lung. In patients with acute lung injury, Cereda et al.33  demonstrated that low VT could induce a progressive decrease in compliance, which could be prevented by setting an adequate PEEP level. Indeed, the use of low VT without setting an appropriate PEEP level could likely exacerbate atelectasis and favor postoperative pulmonary complications (PPC).3,34–36 

Our results suggest that 10 cm H2O of PEEP are needed when a VT of 4 to 5 ml/kg is used. Indeed, 5 cm H2O of PEEP were not able to improve oxygenation or to reduce both shunt and driving pressure (table 2). Recently, Neto et al.37  demonstrated that the higher the intraoperative driving pressure, the greater the incidence of PPC, and this is likely true also in patients undergoing thoracic surgery.36  Of note, a relatively high percentage of our patients (65%) had a value of ΔP higher than 14 cm H2O during OLV at zero PEEP, and recent studies described a significant association between this ΔP cutoff and mortality in patients with ARDS.38  Because in our patients 10 cm H2O of PEEP applied during OLV decreased ΔP from 16 ± 3 cm H2O to 12 ± 3 cm H2O (P < 0.001) and decreased the percentage of patients with a ΔP level of more than 14 cm H2O (29%), we speculate that the combination of low VT and relatively high PEEP levels during OLV could be beneficial in reducing PPC. However, our physiologic study was not designed to investigate the impact of the ventilator strategy on clinical outcomes, and further studies are needed to confirm this hypothesis.

One may argue that, despite the low VT, the application of PEEP can overdistend the lung parenchyma during OLV. In our study, we measured the high V/Q, as a marker of hyperinflation39  and found that it did not change neither at PEEP 5 or 10 cm H2O (table 2). This indicates that PEEP 10 cm H2O, when associated with low VT, does not result in an increase in dead space ventilation. Based on these data, we speculate that PEEP did not cause alveolar hyperinflation in our patients. However, hyperinflation, as commonly defined in the ARDS literature, is usually assessed using CT Hounsfield units and the relationship between hyperinflation and physiologic dead space ventilation (West zone 1).40  Interestingly, we recorded a nonsignificant trend for PEEP–induced increase in high V/Q ratio only in the few patients (6/41; 15%) in which the driving pressure did not decrease by increasing PEEP (Supplemental Digital Content 1, https://links.lww.com/ALN/B572). The lack of positive physiologic response in patients where driving pressure did not decrease by increasing PEEP was also seen in shunt, where the median value changed little on increasing PEEP (PEEP 0: 32% [29 to 45]; PEEP 5: 33% [22 to 40]; PEEP 10: 28% [22 to 34]).

Because patients undergoing thoracic surgery can have very different levels of shunt, usually ranging between 20 and 30%,41,42  we investigated the possible preoperative determinants of shunt to predict a higher risk of intraoperative hypoxia. Interestingly, we found a strong negative correlation (r = −0.79; r2 = 0.62) between the preoperative ERV and the shunt fraction (fig. 3). This was not true for other spirometry parameters, such as forced expiratory volume in the 1st second, FVC, and Tiffenau Index, whereas other clinical or spirometry variables showed only weak (body mass index and FVC/vital capacity) to moderate (KCO) predicting values for intraoperative shunt. The relationship between preoperative ERV and intraoperative shunt can be explained by two factors. First, it is known that FRC and hence ERV are reduced during induction of anesthesia.43  Second, a preexisting low ERV would therefore be reduced further during anesthesia and may result in an FRC below closing volume. Rothen et al.44  previously demonstrated that pulmonary shunt is increased when the closing volume is greater than FRC. PEEP should increase ERV above closing volume reducing shunt and as a consequence weaken the relationship between preoperative ERV and perioperative shunt as observed in this study (Supplemental Digital Content 2 [https://links.lww.com/ALN/B573] and 3 [https://links.lww.com/ALN/B574] show the relationship between ERV and shunt at PEEP 5 and 10 cm H2O, respectively).

Our study has some limitations. First, although the overall shunt levels were similar to those previously reported in the literature,42  this was a single-center study, and thus our results may be dependent on local surgical and anesthesiological practice. Second, the technique for assessing ventilation perfusion matching used in the present study employs a three-compartment lung model. This model has been shown to be a substantial improvement in describing data when compared to oxygenation indices such as the Pao2/Fio245  but does not include the complexity of the 50 compartments model used in the multiple inert gas elimination technique (MIGET), the reference method for assessing gas exchange.46  Though this technique is simpler than the reference one,46  it has been shown to provide a good fit to MIGET data47,48  and to simulate arterial oxygenation with accuracy comparable to the MIGET model.48  Accordingly, considering that the MIGET technique is costly for routine clinical use,49  the presented technique could be regarded as suitable for bedside estimation of the V/Q ratio. Although the model used here accounts for several extrapulmonary parameters, cardiac output (CO) was not measured, and the system assumes a fixed cardiac index. This may be a potential source of errors in the calculation of pulmonary shunt in our patients, because PEEP may impact on CO with several mechanisms, for example by decreasing the cardiac preload or by increasing right ventricular afterload. However, previous studies have showed no significant changes in CO after the application of PEEP in the dependent lung during OLV,8,50,51  and even aggressive recruitment maneuvers have been shown to have slight and transient effects on CO in this context.7,52  Furthermore, a previous validation study of this model showed that its estimate of shunt varies by an average of 2% per liter of CO change.53  It should also be noted that the high V/Q values reported in table 2 represent a functional description of the gas exchange at the lung level rather than an anatomical description, which is usually derived from CT measurements.40  Finally, although our results showed positive short-term physiologic effects of increasing PEEP, further studies are required to see whether the application of protective OLV combined with a PEEP of 10 cm H2O would translate to improved postoperative outcome or whether an even higher PEEP level might be beneficial in some patients.

In conclusion, this study has shown that when using low VT during one-lung ventilation, it is important to apply a proper amount of PEEP to prevent intraoperative increases in driving pressure and intrapulmonary shunt. It is likely that a PEEP of 10 cm H2O is required. Our results indicate that this level of PEEP could be applied without compromising high V/Q. These results are of potential clinical interest for designing lung-protective ventilatory protocols to be applied during OLV.

Support was provided solely from institutional and/or departmental sources.

None of the authors received compensation to perform this study. Dr. Rees is a board member and minor shareholder of Mermaid Care A/S (Nørresundby, Denmark), who commercially produces the ALPE system. Dr. Karbing has performed consultancy work for Mermaid Care A/S. The remaining authors declare no competing interests.

1.
Benumof
JL
:
One-lung ventilation and hypoxic pulmonary vasoconstriction: Implications for anesthetic management.
Anesth Analg
1985
;
64
:
821
33
2.
Hedenstierna
G
,
Tenling
A
:
The lung during and after thoracic anaesthesia.
Curr Opin Anaesthesiol
2005
;
18
:
23
8
3.
Michelet
P
,
Roch
A
,
Brousse
D
,
D’Journo
XB
,
Bregeon
F
,
Lambert
D
,
Perrin
G
,
Papazian
L
,
Thomas
P
,
Carpentier
JP
,
Auffray
JP
:
Effects of PEEP on oxygenation and respiratory mechanics during one-lung ventilation.
Br J Anaesth
2005
;
95
:
267
73
4.
Duggan
M
,
Kavanagh
BP
:
Pulmonary atelectasis: A pathogenic perioperative entity.
Anesthesiology
2005
;
102
:
838
54
5.
Tusman
G
,
Böhm
SH
,
Melkun
F
,
Staltari
D
,
Quinzio
C
,
Nador
C
,
Turchetto
E
:
Alveolar recruitment strategy increases arterial oxygenation during one-lung ventilation.
Ann Thorac Surg
2002
;
73
:
1204
9
6.
Tusman
G
,
Böhm
SH
,
Sipmann
FS
,
Maisch
S
:
Lung recruitment improves the efficiency of ventilation and gas exchange during one-lung ventilation anesthesia.
Anesth Analg
2004
;
98
:
1604
9
7.
Cinnella
G
,
Grasso
S
,
Natale
C
,
Sollitto
F
,
Cacciapaglia
M
,
Angiolillo
M
,
Pavone
G
,
Mirabella
L
,
Dambrosio
M
:
Physiological effects of a lung-recruiting strategy applied during one-lung ventilation.
Acta Anaesthesiol Scand
2008
;
52
:
766
75
8.
Abe
K
,
Mashimo
T
,
Yoshiya
I
:
Arterial oxygenation and shunt fraction during one-lung ventilation: A comparison of isoflurane and sevoflurane.
Anesth Analg
1998
;
86
:
1266
70
9.
Mascotto
G
,
Bizzarri
M
,
Messina
M
,
Cerchierini
E
,
Torri
G
,
Carozzo
A
,
Casati
A
:
Prospective, randomized, controlled evaluation of the preventive effects of positive end-expiratory pressure on patient oxygenation during one-lung ventilation.
Eur J Anaesthesiol
2003
;
20
:
704
10
10.
Slinger
PD
,
Kruger
M
,
McRae
K
,
Winton
T
:
Relation of the static compliance curve and positive end-expiratory pressure to oxygenation during one-lung ventilation.
Anesthesiology
2001
;
95
:
1096
102
11.
Leong
LM
,
Chatterjee
S
,
Gao
F
:
The effect of positive end expiratory pressure on the respiratory profile during one-lung ventilation for thoracotomy.
Anaesthesia
2007
;
62
:
23
6
12.
Shen
Y
,
Zhong
M
,
Wu
W
,
Wang
H
,
Feng
M
,
Tan
L
,
Wang
Q
:
The impact of tidal volume on pulmonary complications following minimally invasive esophagectomy: A randomized and controlled study.
J Thorac Cardiovasc Surg
2013
;
146
:
1267
74
13.
Schilling
T
,
Kozian
A
,
Huth
C
,
Bühling
F
,
Kretzschmar
M
,
Welte
T
,
Hachenberg
T
:
The pulmonary immune effects of mechanical ventilation in patients undergoing thoracic surgery.
Anesth Analg
2005
;
101
:
957
65
14.
Kozian
A
,
Schilling
T
,
Schütze
H
,
Senturk
M
,
Hachenberg
T
,
Hedenstierna
G
:
Ventilatory protective strategies during thoracic surgery: Effects of alveolar recruitment maneuver and low-tidal volume ventilation on lung density distribution.
Anesthesiology
2011
;
114
:
1025
35
15.
Rozé
H
,
Lafargue
M
,
Perez
P
,
Tafer
N
,
Batoz
H
,
Germain
C
,
Janvier
G
,
Ouattara
A
:
Reducing tidal volume and increasing positive end-expiratory pressure with constant plateau pressure during one-lung ventilation: Effect on oxygenation.
Br J Anaesth
2012
;
108
:
1022
7
16.
Lohser
J
,
Slinger
P
:
Lung injury after one-lung ventilation: A review of the pathophysiologic mechanisms affecting the ventilated and the collapsed lung.
Anesth Analg
2015
;
121
:
302
18
17.
Bendixen
HH
,
Hedley-Whyte
J
,
Laver
MB
:
Impaired oxygenation in surgical patients during general anesthesia with controlled ventilation: A concept of atelectasis.
N Engl J Med
1963
;
269
:
991
6
18.
Vestbo
J
,
Hurd
SS
,
Agustí
AG
,
Jones
PW
,
Vogelmeier
C
,
Anzueto
A
,
Barnes
PJ
,
Fabbri
LM
,
Martinez
FJ
,
Nishimura
M
,
Stockley
RA
,
Sin
DD
,
Rodriguez-Roisin
R
:
Global strategy for the diagnosis, management, and prevention of chronic obstructive pulmonary disease: GOLD executive summary.
Am J Respir Crit Care Med
2013
;
187
:
347
65
19.
Tusman
G
,
Böhm
SH
,
Warner
DO
,
Sprung
J
:
Atelectasis and perioperative pulmonary complications in high-risk patients.
Curr Opin Anaesthesiol
2012
;
25
:
1
10
20.
Volta
CA
,
Trentini
A
,
Farabegoli
L
,
Manfrinato
MC
,
Alvisi
V
,
Dallocchio
F
,
Marangoni
E
,
Alvisi
R
,
Bellini
T
:
Effects of two different strategies of fluid administration on inflammatory mediators, plasma electrolytes and acid/base disorders in patients undergoing major abdominal surgery: A randomized double blind study.
J Inflamm (Lond)
2013
;
10
:
29
21.
Volta
CA
,
Alvisi
V
,
Petrini
S
,
Zardi
S
,
Marangoni
E
,
Ragazzi
R
,
Capuzzo
M
,
Alvisi
R
:
The effect of volatile anesthetics on respiratory system resistance in patients with chronic obstructive pulmonary disease.
Anesth Analg
2005
;
100
:
348
53
22.
Gattinoni
L
,
Brazzi
L
,
Pelosi
P
,
Latini
R
,
Tognoni
G
,
Pesenti
A
,
Fumagalli
R
:
Svo2 Collaborative Group: A trial of goal-oriented hemodynamic therapy in critically ill patients.
N Engl J Med
1995
;
333
:
1025
32
23.
Gehan
EA
,
George
SL
:
Estimation of human body surface area from height and weight.
Cancer Chemother Rep
1970
;
54
:
225
35
24.
Edmark
L
,
Auner
U
,
Enlund
M
,
Ostberg
E
,
Hedenstierna
G
:
Oxygen concentration and characteristics of progressive atelectasis formation during anaesthesia.
Acta Anaesthesiol Scand
2011
;
55
:
75
81
25.
Karbing
DS
,
Kjærgaard
S
,
Andreassen
S
,
Espersen
K
,
Rees
SE
:
Minimal model quantification of pulmonary gas exchange in intensive care patients.
Med Eng Phys
2011
;
33
:
240
8
26.
Spadaro
S
,
Karbing
DS
,
Mauri
T
,
Marangoni
E
,
Mojoli
F
,
Valpiani
G
,
Carrieri
C
,
Ragazzi
R
,
Verri
M
,
Rees
SE
,
Volta
CA
:
Effect of positive end-expiratory pressure on pulmonary shunt and dynamic compliance during abdominal surgery.
Br J Anaesth
2016
;
116
:
855
61
27.
Kjaergaard
S
,
Rees
SE
,
Grønlund
J
,
Nielsen
EM
,
Lambert
P
,
Thorgaard
P
,
Toft
E
,
Andreassen
S
:
Hypoxaemia after cardiac surgery: Clinical application of a model of pulmonary gas exchange.
Eur J Anaesthesiol
2004
;
21
:
296
301
28.
Kjaergaard
S
,
Rees
S
,
Malczynski
J
,
Nielsen
JA
,
Thorgaard
P
,
Toft
E
,
Andreassen
S
:
Non-invasive estimation of shunt and ventilation-perfusion mismatch.
Intensive Care Med
2003
;
29
:
727
34
29.
Chiumello
D
,
Coppola
S
,
Froio
S
,
Mietto
C
,
Brazzi
L
,
Carlesso
E
,
Gattinoni
L
:
Time to reach a new steady state after changes of positive end expiratory pressure.
Intensive Care Med
2013
;
39
:
1377
85
30.
Evans
JD
:
Straightforward statistics for the behavioral sciences
.
Pacific Grove, CA
,
Brooks/Cole Publishing
,
1996
31.
Richard
JC
,
Maggiore
SM
,
Jonson
B
,
Mancebo
J
,
Lemaire
F
,
Brochard
L
:
Influence of tidal volume on alveolar recruitment: Respective role of PEEP and a recruitment maneuver.
Am J Respir Crit Care Med
2001
;
163
:
1609
13
32.
Qutub
H
,
El-Tahan
MR
,
Mowafi
HA
,
El Ghoneimy
YF
,
Regal
MA
,
Al Saflan
AA
:
Effect of tidal volume on extravascular lung water content during one-lung ventilation for video-assisted thoracoscopic surgery: A randomised, controlled trial.
Eur J Anaesthesiol
2014
;
31
:
466
73
33.
Cereda
M
,
Foti
G
,
Musch
G
,
Sparacino
ME
,
Pesenti
A
:
Positive end-expiratory pressure prevents the loss of respiratory compliance during low tidal volume ventilation in acute lung injury patients.
Chest
1996
;
109
:
480
5
34.
Levin
MA
,
McCormick
PJ
,
Lin
HM
,
Hosseinian
L
,
Fischer
GW
:
Low intraoperative tidal volume ventilation with minimal PEEP is associated with increased mortality.
Br J Anaesth
2014
;
113
:
97
108
35.
Yang
M
,
Ahn
HJ
,
Kim
K
,
Kim
JA
,
Yi
CA
,
Kim
MJ
,
Kim
HJ
:
Does a protective ventilation strategy reduce the risk of pulmonary complications after lung cancer surgery?: A randomized controlled trial.
Chest
2011
;
139
:
530
7
36.
Blank
RS
,
Colquhoun
DA
,
Durieux
ME
,
Kozower
BD
,
McMurry
TL
,
Bender
SP
,
Naik
BI
:
Management of one-lung ventilation: Impact of tidal volume on complications after thoracic surgery.
Anesthesiology
2016
;
124
:
1286
95
37.
Neto
AS
,
Barbas
CSV
,
Simonis
FD
,
Artigas-Raventós
A
,
Canet
J
,
Determann
RM
,
Anstey
J
,
Hedenstierna
G
,
Hemmes
SNT
,
Hermans
G
,
Hiesmayr
M
,
Hollmann
MW
,
Jaber
S
,
Martin-Loeches
I
,
Mills
GH
,
Pearse
RM
,
Putensen
C
,
Schmid
W
,
Severgnini
P
,
Smith
R
,
Treschan
TA
,
Tschernko
EM
,
Melo
MFV
,
Wrigge
H
,
de Abreu
MG
,
Pelosi
P
,
Schultz
MJ
;
PRoVENT; PROVE Network investigators
:
Epidemiological characteristics, practice of ventilation, and clinical outcome in patients at risk of acute respiratory distress syndrome in intensive care units from 16 countries (PRoVENT): An international, multicentre, prospective study.
Lancet Respir Med
2016
;
4
:
882
93
38.
Bellani
G
,
Laffey
JG
,
Pham
T
,
Fan
E
,
Brochard
L
,
Esteban
A
,
Gattinoni
L
,
van Haren
F
,
Larsson
A
,
McAuley
DF
,
Ranieri
M
,
Rubenfeld
G
,
Thompson
BT
,
Wrigge
H
,
Slutsky
AS
,
Pesenti
A
;
LUNG SAFE Investigators; ESICM Trials Group
:
Epidemiology, patterns of care, and mortality for patients with acute respiratory distress syndrome in intensive care units in 50 countries.
JAMA
2016
;
315
:
788
800
39.
Murias
G
,
Blanch
L
,
Lucangelo
U
:
The physiology of ventilation.
Respir Care
2014
;
59
:
1795
807
40.
Gattinoni
L
,
Caironi
P
,
Pelosi
P
,
Goodman
LR
:
What has computed tomography taught us about the acute respiratory distress syndrome?
Am J Respir Crit Care Med
2001
;
164
:
1701
11
41.
Karzai
W
,
Schwarzkopf
K
:
Hypoxemia during one-lung ventilation: Prediction, prevention, and treatment.
Anesthesiology
2009
;
110
:
1402
11
42.
Beck
DH
,
Doepfmer
UR
,
Sinemus
C
,
Bloch
A
,
Schenk
MR
,
Kox
WJ
:
Effects of sevoflurane and propofol on pulmonary shunt fraction during one-lung ventilation for thoracic surgery.
Br J Anaesth
2001
;
86
:
38
43
43.
Juno
J
,
Marsh
HM
,
Knopp
TJ
,
Rehder
K
:
Closing capacity in awake and anesthetized-paralyzed man.
J Appl Physiol Respir Environ Exerc Physiol
1978
;
44
:
238
44
44.
Rothen
HU
,
Sporre
B
,
Engberg
G
,
Wegenius
G
,
Hedenstierna
G
:
Airway closure, atelectasis and gas exchange during general anaesthesia.
Br J Anaesth
1998
;
81
:
681
6
45.
Karbing
DS
,
Kjaergaard
S
,
Smith
BW
,
Espersen
K
,
Allerød
C
,
Andreassen
S
,
Rees
SE
:
Variation in the Pao2/Fio2 ratio with Fio2: Mathematical and experimental description, and clinical relevance.
Crit Care
2007
;
11
:
R118
46.
Wagner
PD
,
Saltzman
HA
,
West
JB
:
Measurement of continuous distributions of ventilation-perfusion ratios: Theory.
J Appl Physiol
1974
;
36
:
588
99
47.
Rees
SE
,
Kjaergaard
S
,
Andreassen
S
,
Hedenstierna
G
:
Reproduction of MIGET retention and excretion data using a simple mathematical model of gas exchange in lung damage caused by oleic acid infusion.
J Appl Physiol (1985)
2006
;
101
:
826
32
48.
Rees
SE
,
Kjaergaard
S
,
Andreassen
S
,
Hedenstierna
G
:
Reproduction of inert gas and oxygenation data: A comparison of the MIGET and a simple model of pulmonary gas exchange.
Intensive Care Med
2010
;
36
:
2117
24
49.
Wagner
PD
:
Assessment of gas exchange in lung disease: Balancing accuracy against feasibility.
Crit Care
2007
;
11
:
182
50.
Fujiwara
M
,
Abe
K
,
Mashimo
T
:
The effect of positive end-expiratory pressure and continuous positive airway pressure on the oxygenation and shunt fraction during one-lung ventilation with propofol anesthesia.
J Clin Anesth
2001
;
13
:
473
7
51.
Ferrando
C
,
Mugarra
A
,
Gutierrez
A
,
Carbonell
JA
,
García
M
,
Soro
M
,
Tusman
G
,
Belda
FJ
:
Setting individualized positive end-expiratory pressure level with a positive end-expiratory pressure decrement trial after a recruitment maneuver improves oxygenation and lung mechanics during one-lung ventilation.
Anesth Analg
2014
;
118
:
657
65
52.
Garutti
I
,
Martinez
G
,
Cruz
P
,
Piñeiro
P
,
Olmedilla
L
,
de la Gala
F
:
The impact of lung recruitment on hemodynamics during one-lung ventilation.
J Cardiothorac Vasc Anesth
2009
;
23
:
506
8
53.
Karbing
DS
,
Allerød
C
,
Thomsen
LP
,
Espersen
K
,
Thorgaard
P
,
Andreassen
S
,
Kjærgaard
S
,
Rees
SE
:
Retrospective evaluation of a decision support system for controlled mechanical ventilation.
Med Biol Eng Comput
2012
;
50
:
43
51