Background

Airway resistance depends not only on an airway's geometry but also on flow rate, and gas density and viscosity. A recent study showed that at clinically relevant concentrations, the mixtures of volatile agents with air and oxygen and oxygen-nitrogen affected the density of the mixture. The goal of the current study was to investigate the effect of different minimum alveolar concentrations (MACs) of three commonly used volatile agents, isoflurane, sevoflurane, and desflurane, on the measurements of airway resistance.

Methods

A two-chamber fixed-resistance test lung was connected to an anesthesia machine using the volume control mode of ventilation. Pulmonary resistance was calculated at baseline (25% oxygen in air); at 1.0, 1.5, and 2.0 MAC; and also at the same concentrations, 1.2% and 4%, of isoflurane, sevoflurane, and desflurane mixtures with 25% oxygen in air. The analysis of variance test for repeated measures and probabilities for post hoc Tukey and least significant difference tests were used.

Results

Isoflurane affected pulmonary resistance only at 2 MAC. Sevoflurane caused a significant increase of pulmonary resistance at 1.5 and 2 MAC, whereas desflurane caused the greatest increase in pulmonary resistance at all MAC values used. At 1.2% concentration, no difference from the baseline resistance was observed, whereas at 4%, the three agents produced similar increases of pulmonary resistance.

Conclusion

High concentrations of volatile agents in 25% oxygen in air increased the density of the gas mixture and the calculated resistance of a test lung model with fixed resistance.

AIRWAY resistance depends on airway geometry, flow rate, and gas density and viscosity.1–3When the flow is laminar, viscosity is the only physical property of the inspired gas that may influence resistance. However, if the gas flow is turbulent, resistance depends mainly on the density of the gas. In a recent study, Habre et al.  4showed that clinical concentrations of volatile agents in mixtures with air, oxygen, and oxygen–nitrogen did not markedly affect the viscosity of the mixture, whereas they affected the density of the mixture at routinely used clinical concentrations, with maximal increase in the density of desflurane in air by 47.7% at 2 minimal alveolar concentration (MAC).4The aim of the current study was to compare the effect of different concentrations of isoflurane, sevoflurane, and desflurane on the measured pulmonary resistance under experimental conditions in a laboratory lung model with fixed resistance. The working hypothesis was that the volatile agents with high density would increase the density of their mixture with 25% oxygen in air and would lead to an increased pulmonary resistance. This effect should have been greater at higher volatile concentrations.

The setup of the experiment is presented in figure 1. A two-chamber test lung with fixed resistance (5600i Dual Adult System; Michigan Instruments, Grand Rapids, MI) was connected to an anesthesia machine (Julian; Draëger Medical, Lübeck, Germany) using the volume control mode of ventilation. The test lung served as a quantitative, calibrated “test load” simulating human pulmonary physiology. The elastance and resistance of the experimental chambers were set at 20 cm H2O · l−1and 15 cm H2O · l−1· s−1, respectively.5 

Fig. 1. Setup of the experiment. The two-chamber test lung (  1 ) was connected to the anesthesia machine (  2 ). The pneumotachograph (  3 ) and the pressure transducer were inserted between the endotracheal tube and the Y piece (  4 ) of the respiratory circuit for the continuous recordings of measured flow, tidal volume, and inspiratory pressures. 

Fig. 1. Setup of the experiment. The two-chamber test lung (  1 ) was connected to the anesthesia machine (  2 ). The pneumotachograph (  3 ) and the pressure transducer were inserted between the endotracheal tube and the Y piece (  4 ) of the respiratory circuit for the continuous recordings of measured flow, tidal volume, and inspiratory pressures. 

Close modal

A screen pneumotachograph, which used a differential pressure–based flow sensor and was used for the measurement of flow rate (RSS100-HR; Hans Rudolph, Kansas City, MO), and a pressure transducer were inserted between the endotracheal tube and the Y piece of the respiratory circuit for the continuous recordings of measured flow, tidal volume, and inspiratory pressures (Ppeakand Pplateau). This particular pneumotachograph provides correction for gas density, viscosity, temperature, and barometric pressure. Moreover, to confirm accuracy, the flows measured at baseline and after the addition of anesthetic gases were compared. Because the lung model was ventilated with volume control mode, it was mandatory that flow remain constant, and only minor breath-to-breath variations should have been recorded. The pulmonary resistance (Rplm) was calculated by the inspiratory pressure method,6where the driving pressure for airflow is the difference between end-inspiratory peak airway pressure and plateau pressure (Ppeak– Pplateau) during occlusion of the expiratory port. The Ppeakand Pplateaupressures were measured from the waveform of the inspiratory pressure. Flow was measured from the waveform of the pneumotachograph (fig. 2). The overall pulmonary resistance (Rplm) was calculated from the equation: pulmonary resistance Rplm= (Ppeak– Pplateau)/(flow) at baseline, when a mixture of 25% oxygen in air was used (without volatile agent) and also at mixtures of 1.0, 1.5, and 2.0 MAC of isoflurane, sevoflurane, and desflurane with 25% oxygen in air. Finally, to discriminate between the effects of the concentration of the anesthetics or the effects of their physical properties, we also measured the resistance at the same low (1.2%) and high (4%) concentrations of isoflurane sevoflurane, and desflurane with 25% oxygen in air. These particular concentrations were chosen because 1.2% was 1 MAC of isoflurane, and 4% is the highest concentration that could be delivered from the isoflurane vaporizer. For each concentration of the volatile agent, five measurements were recorded.

Fig. 2. Recorded waveforms of flow and pressure. Total pulmonary resistance was calculated according to the equation: pulmonary resistance (Rplm) = (Ppeak− Pplateau)/flow. The use of the laboratory model resulted in no time constant inequalities, and P1pressure is not observed. 

Fig. 2. Recorded waveforms of flow and pressure. Total pulmonary resistance was calculated according to the equation: pulmonary resistance (Rplm) = (Ppeak− Pplateau)/flow. The use of the laboratory model resulted in no time constant inequalities, and P1pressure is not observed. 

Close modal

The parameters of the volume control mode of ventilation were the same for all measurements: tidal volume = 600 ml, respiratory rate = 8 breaths/min, positive end-expiratory pressure = 0 cm H2O, TI:TE= 1:1.5, TI:Tplateau= 50%, and fresh gas flow = 8 l/min. The MAC values of each anesthetic agent were calculated by the Agent Analyser (IRIA; Draëger Medical, Lübeck, Germany) of the anesthetic apparatus using the volatile agent end-expiratory concentrations and corrected for age (at a set age of 40 yr), altitude, and gas mixture. IRIA measures the concentration of the anesthetics based on infrared light absorption. The absorption of infrared light at 3 μm is used for the measurement of carbon dioxide–nitrous oxide and at 8 μm for the measurement of end-expiratory volatile anesthetics. Age-corrected MAC values were then calculated according to the Mapleson formula for patients older than 1 yr: MACage= MAC40× 10 [−0.00269 × (age − 40)].7 

Statistical Analysis

All values represent the means from a total of five consecutive respiratory cycles. The increase of pulmonary resistance caused by the volatile agents was also expressed as a percentage of the baseline value. Data were analyzed by analysis of variance for repeated measurements, and post hoc  Tukey and least significant difference tests were  used. All values are expressed as mean ± SD.

The measured end-expiratory concentrations for isoflurane, sevoflurane, and desflurane at the three studied MAC values are shown in table 1. Desflurane was used at greater end-expiratory concentrations at all MAC values (6.6, 9.7, and 13.3% representing 1, 1.5, and 2 MAC, respectively), because it is the least potent of the three volatile agents.

Table 1. End-expiratory Concentrations (%) for Isoflurane, Sevoflurane, and Desflurane 

Table 1. End-expiratory Concentrations (%) for Isoflurane, Sevoflurane, and Desflurane 
Table 1. End-expiratory Concentrations (%) for Isoflurane, Sevoflurane, and Desflurane 

The flow measurements for all gas compositions are shown in table 2. None of the differences recorded were significant. This finding confirmed that the pneumotachograph measures were valid because, with the use of volume control ventilation mode, inspiratory flow was mandatory to remain constant, and only minor breath-to-breath variations were acceptable. Otherwise, the pneumotachograph gave false values because of the altered physical properties of the gas mixtures.

Table 2. Flow Measurements (l/min) at the Different Concentrations of Volatile Agents 

Table 2. Flow Measurements (l/min) at the Different Concentrations of Volatile Agents 
Table 2. Flow Measurements (l/min) at the Different Concentrations of Volatile Agents 

The baseline value of pulmonary resistance of the lung model at 25% oxygen in air was 15.64 ± 0.15 cm H2O · l−1· s−1. The effects of the different concentrations of isoflurane, sevoflurane, and desflurane in 25% oxygen in air on pulmonary resistance are shown in table 3and figure 3.

Table 3. Pulmonary Resistance (cm H2O · l−1· s−1) at the Different Concentrations Studied 

Table 3. Pulmonary Resistance (cm H2O · l−1· s−1) at the Different Concentrations Studied 
Table 3. Pulmonary Resistance (cm H2O · l−1· s−1) at the Different Concentrations Studied 

Fig. 3. Comparison of the effects of isoflurane, sevoflurane, and desflurane at 1.0, 1.5, and 2 minimum alveolar concentration (MAC) on total pulmonary resistance. Isoflurane caused a statistically significant (  P = 0.005) increase of the resistance at 2 MAC compared with baseline. Sevoflurane caused significant increase of the resistance at 1.5 and 2 MAC compared with baseline (  P < 0.001 for both comparisons). Desflurane caused significant increase of the resistance from baseline at the three concentrations studied (  P < 0.001 for all comparisons). 

Fig. 3. Comparison of the effects of isoflurane, sevoflurane, and desflurane at 1.0, 1.5, and 2 minimum alveolar concentration (MAC) on total pulmonary resistance. Isoflurane caused a statistically significant (  P = 0.005) increase of the resistance at 2 MAC compared with baseline. Sevoflurane caused significant increase of the resistance at 1.5 and 2 MAC compared with baseline (  P < 0.001 for both comparisons). Desflurane caused significant increase of the resistance from baseline at the three concentrations studied (  P < 0.001 for all comparisons). 

Close modal

Comparison at Equivalent MAC Values

Compared with the baseline value, isoflurane significantly increased pulmonary resistance only at 2 MAC (P = 0.005). Sevoflurane increased pulmonary resistance at 1.5 and at 2 MAC by 17.8% and 23.3%, respectively (P < 0.001 for both comparisons). Desflurane increased the calculated resistance of the test lung model at all concentrations used (P < 0.001 for all concentrations). At 1 MAC, where the end-expiratory concentration of desflurane was 6.6%, the increase of pulmonary resistance from the baseline value was 44.5%. At 1.5 MAC desflurane, an end-expiratory concentration of 9.7% was used, and the increase of pulmonary resistance was 63.7% from the baseline value. The greatest increase of pulmonary resistance, by 87%, was observed at 2 MAC when an end-expiratory concentration of 13.3% was used.

When comparing the effect of the three volatile agents on the pulmonary resistance at equivalent concentrations, it was shown that at 1 MAC isoflurane and sevoflurane had similar effects on pulmonary resistance, whereas desflurane caused a significant increase compared with the two other agents (P < 0.001; fig. 4). At 1.5 MAC, sevoflurane caused a significant increase of pulmonary resistance compared with isoflurane (P < 0.001), and desflurane caused the greatest increase compared with the two other agents (P < 0.001 for all comparisons). Sevoflurane and desflurane did increase overall pulmonary resistance by 17.8% and 63.7%, respectively. At 2 MAC, a statistically significant increase of pulmonary resistance was observed with all of the three volatile agents. Sevoflurane had a significantly greater effect compared with isoflurane (P < 0.001), and desflurane caused the greatest increase compared with the two other agents (P < 0.001 for both comparisons; table 3and fig. 4).

Fig. 4. Comparison of the effect of different volatile anesthetics at equivalent concentrations on total pulmonary resistance. At 1 minimum alveolar concentration (MAC), only desflurane significantly increased pulmonary resistance compared with isoflurane and sevoflurane (  P < 0.001 for both comparisons). The difference between isoflurane and sevoflurane was not statistically significant (  P = 0.15). At 1.5 MAC, sevoflurane significantly increased (  P = 0.015) total pulmonary resistance compared with isoflurane, whereas desflurane caused a more pronounced increase compared with isoflurane and sevoflurane (  P < 0.001 for all comparisons). The same findings apply at 2 MAC. Sevoflurane significantly increased pulmonary resistance compared with isoflurane (  P < 0.001) and desflurane caused a significant increase compared with the other two volatile agents (  P < 0.001 for both comparisons). * Increased pulmonary resistance compared with seroflurane and isoflurane; ** increased pulmonary resistance compared with isoflurane. 

Fig. 4. Comparison of the effect of different volatile anesthetics at equivalent concentrations on total pulmonary resistance. At 1 minimum alveolar concentration (MAC), only desflurane significantly increased pulmonary resistance compared with isoflurane and sevoflurane (  P < 0.001 for both comparisons). The difference between isoflurane and sevoflurane was not statistically significant (  P = 0.15). At 1.5 MAC, sevoflurane significantly increased (  P = 0.015) total pulmonary resistance compared with isoflurane, whereas desflurane caused a more pronounced increase compared with isoflurane and sevoflurane (  P < 0.001 for all comparisons). The same findings apply at 2 MAC. Sevoflurane significantly increased pulmonary resistance compared with isoflurane (  P < 0.001) and desflurane caused a significant increase compared with the other two volatile agents (  P < 0.001 for both comparisons). * Increased pulmonary resistance compared with seroflurane and isoflurane; ** increased pulmonary resistance compared with isoflurane. 

Close modal

Comparison at the Same Concentrations

There was no significant increase of pulmonary resistance from the baseline when the three anesthetics were delivered at the same low concentration of 1.2%, whereas at 4%, all three produced a significant increase that ranged from 25.8% for desflurane to 28% for isoflurane and 28.9% for sevoflurane (P < 0.001 for the three agents; table 3). The differences between the three anesthetics at the aforementioned concentrations were not significant.

Pulmonary resistance  is defined as the opposition to the flow of gases caused by frictional forces within the respiratory system, and it is calculated according to the equation: Resistance = driving pressure/flow rate.1These frictional pressure losses (ΔP) produced by flow in tubes are a function of flow rate, tube geometry, and fluid physical properties.2 

The inspiratory pressure method6was used to calculate the overall pulmonary resistance according to the equation R = (Ppeak− Pplateau)/flow. During passive ventilation, rapid airway occlusion at the end of inspiration produces an immediate decrease in transpulmonary pressure from its peak (Ppeak) to a lower value (P1), followed by a gradual decrease in pressure to an apparent plateau (Pplateau), which represents the static end-expiratory elastic recoil of the lung. This permits pulmonary flow resistance to be partitioned between the true intrinsic resistance of the airways (Rmin= (Ppeak– P1)/flow) and an additional effective resistance due to time-constant inequalities and tissue stress adaptation (RL= (P1− Pplateau)/flow).8 

This method provides comparable values of resistance with other techniques but does not permit discrimination between patients with airway obstruction and those with infiltrative lung disease. However, the limitations of this method are not considered to affect the measurements during the experiment of the current study.9 

In the current study, the use of a laboratory lung model that simulates the respiratory tract eliminated the effects of lung volume and bronchial tone on airway resistance that largely affect airway resistance in vivo . Therefore, any change to the calculated resistance resulted from changes in the gas mixture physical properties, and changes in the flow pattern that influences the pressure decrease determined by the difference between peak pressure and plateau pressure (ΔP = Ppeak− Pplateau). The test lung served as a quantitative, calibrated “test load” simulating human pulmonary physiology for use in testing respiratory equipment and pulmonary research. It is designed to realistically simulate the mechanics of the adult respiratory system from the upper airway. The lung model is not a detailed model of actual human anatomy, and its use helped in determining whether the physical properties of a gas mixture alter the measurements of a fixed resistance. The adult airway is constructed by using a hose assembly, and the airway resistance exhibits parabolic characteristics in regard to pressure change as a function of flow. This nonlinear parabolic characteristic is similar to that seen in standard endotracheal tubes. Similar models have been previously used for studies examining lung mechanics during different modes of ventilation.5,10Compliance can be independently set for each lung using a precision-calibrated spring, and plug-in resistors are provided to simulate pulmonary resistance. The baseline resistance selected (15 cm H2O · l−1· s−1) has previously been reported to occur during general anesthesia.11The lung model mostly represents the central airways and more specifically the trachea and the main bronchi, whereas it cannot simulate alveoli. Therefore, time constant inequalities did not occur, P1pressure was not observed, and only Ppeakand Pplateaupressures were recorded (fig. 2).

The lung model was ventilated with constant flow inflations, and the inspiratory flow was recorded with screen pneumotachograph that provided correction for gas density, viscosity, temperature, and barometric pressure. This type of pneumotachograph contains a fine-mesh screen that provides a small fixed resistance to airflow. As gas flows through the pneumotachograph, a microprocessor-based system converts the measured differential pressure to volumetric flow rate. The screen is actively heated to prevent condensation, which leads to alterations of the resistance and inaccurate results. Furthermore, it was easy to confirm its accuracy by comparing the flow measured at baseline and after the addition of anesthetic gases. Because the lung model was ventilated with volume control mode, it was mandatory that flow remain constant, and only minor breath-to-breath variations should have been recorded.

The fact that flow measurements remained constant for all gas mixtures delivered confirmed the accuracy of the pneumotachograph.

Volatile anesthetic agents produce dose-dependent decreases in airway resistance after antigen-induced bronchoconstriction in animal models.12Halothane is considered to have the most potent bronchodilatory properties due to decreased vagal tone,13but all volatile anesthetics possess bronchodilatory properties.14,15However, there are no studies regarding the effects of the physical properties of volatile anesthetics on airway resistance. Habre et al.  4estimated the viscosity and density values of the pure component of volatile anesthetics. (isoflurane: viscosity 0.892 Pa s × 10−5, density 5.19 kg/m3; sevoflurane: viscosity 1.276 Pa s × 10−5, density 6.12 kg/m3; desflurane: viscosity 1.452 Pa s × 10−5, density 5.44 kg/m3). They also calculated the viscosity and density of their mixtures with air, 100% oxygen, and 50% oxygen. They concluded that volatile agents in the clinically applied concentrations produced a relatively small decrease of the viscosity values of gas mixtures with maximal decrease in viscosity of 3.3–3.5% at 2 MAC desflurane, whereas they significantly increased the density of the gas mixture with maximal increase of density to 47.7% at 2 MAC desflurane in air.4It should be mentioned that either increased density or decreased viscosity increase pulmonary resistance. The current study clearly demonstrated that isoflurane, sevoflurane, and desflurane at high concentrations increased the calculated overall pulmonary resistance in a test lung model, and the greatest increase was observed at high desflurane concentrations (2 MAC). The significant increase of pulmonary resistance observed with desflurane was considered to be the result of the high concentrations required to achieve 1, 1.5, and 2 MAC values (6.6, 9.7, and 13.3%, respectively; table 1) and not from unusually deviant values of viscosity and density of the pure component of the anesthetic.4This observation was supported by the finding that the three anesthetics produced similar effects on airway resistance when delivered at the same low or high concentration, suggesting that the increase of the resistance is due to the concentration delivered rather than differences of the physical properties of the agents used. Similar effects of increased resistance measurements are reported in previous studies regarding the effects of xenon on respiratory mechanics,16,17which showed that the high density of xenon increased airway pressure and resistance in animal models.

The findings of the current study may explain the discrepancies between in vivo  and in vitro  studies regarding the bronchodilatory properties of desflurane. Mercier et al.  18found that desflurane relaxed proximal isolated human bronchi in a dose-dependent manner. On the contrary, Goff et al.  19found that desflurane exerted no bronchodilation in patients undergoing elective surgery during general anesthesia. In the former study, the relaxation of the bronchi was measured directly, whereas in the latter, the isovolume method was used to assess total airway resistance, which is a product of bronchial tone, lung volume, flow rate, and the physical properties of the inspired gas. The current study may also explain the results of Dikmen et al. ,11who found that desflurane produced bronchodilation at 1 MAC but increased airway resistance at 2 MAC. At 2 MAC, the increased density of the gas mixture may have offset the bronchodilatory effect of the anesthetic, and the overall resistance measured was increased. Theoretically, the increased pulmonary resistance caused by the physical properties of anesthetics might have deleterious effects on patients with chronic obstructive pulmonary disease or asthma. However, desflurane, which had the most pronounced effects of pulmonary resistance, has been widely used without adverse effects even in patients breathing spontaneously,20and there are no studies in humans to support that it should be used with caution in patients with airway hyperreactivity.

In conclusion, the high density of volatile anesthetics significantly increases airway resistance. This phenomenon is more pronounced with less potent agents that are delivered in high concentrations. Studies of the effects of these agents on airway resistance should take into account that a percent of these effects may result from the altered density of the inspired gas mixture.

1.
Tobin MJ, Van de Graaff: Monitoring of lung mechanics and work of breathing, Principles and Practice of Mechanical Ventilation, 1st edition. Edited by Tobin MJ. New York, McGraw-Hill, 1994, pp 967–1004Tobin MJ
New York
,
McGraw-Hill
2.
Wood LD, Engel LA, Griffin P, Despas P, Macklem PT: Effect of gas physical properties and flow on lower pulmonary resistance. J Appl Physiol 1976; 41:234–44
3.
Drazen JM, Loring SH, Ingram RH: Distribution of pulmonary resistance: Effects of gas density, viscosity and flow rate. J Appl Physiol 1976; 41:388–95
4.
Habre W, Asztalos T, Sly PD, Petak F: Viscosity and density of common anesthetic gases: Implications for flow measurements. Br J Anaesth 2001; 87:602–7
5.
Prinianakis G, Kondili E, Georgopoulos D: Effects of the flow waveform method of triggering and cycling on patient-ventilator interaction during pressure support. Intensive Care Med 2003; 29:1950–9
6.
Suter PM, Fairley HB, Isenberg MD: Optimum end-expiratory pressure in patients with acute pulmonary failure. N Engl J Med 1975; 292:284–9
7.
Mapleson WW: Effect of age on MAC in humans: A meta-analysis. Br J Anaesth 1996; 76:179–85
8.
Lavietes MH, Rochester DF: Assessment of airway function during assisted ventilation. Lung 1981; 125:219–29
9.
Bates JH, Ludwig MS, Sly PD, Brown K, Martin JG, Fredberg JJ: Interrupter resistance elucidated by alveolar pressure measurement in open-chest normal dogs. J Appl Physiol 1988; 65:408–14
10.
Gowski DT, Delgado E, Miro AM, Tasota FJ, Hoffman LA, Pinsky MR: Tracheal gas insufflation during pressure-control ventilation: Effect of using a pressure relief valve. Critical Care Med 1997; 25:145–52
11.
Dikmen Y, Eminoglou E, Salihoglou Z, Demiroluk S: Pulmonary mechanics during isoflurane, sevoflurane and desflurane anaesthesia. Anaesthesia 2003; 58:745–8
12.
Hirshman CA, Edelstein G, Peetz S, Wayne R, Downes H: Mechanism of action of inhalational anesthesia on airways. Anesthesiology 1982; 56:107–11
13.
Tobias JD, Hirshman CA: Attenuation of histamine-induced airway constriction by albuterol during halothane anesthesia. Anesthesiology 1990; 72:105–10
14.
Rooke GA, Choi JH, Bishop MJ: The effect of isoflurane, halothane, sevoflurane and thiopental/nitrous oxide on respiratory system resistance after tracheal intubation. Anesthesiology 1997; 86:1294–9
15.
Habre W, Petak F, Sly PD, Hantos Z, Morel DR: Protective effects of volatile agents against methacholine-induced bronchoconstriction in rats. Anesthesiology 2001; 94:348–53
16.
Ping Z, Akitoshi O, Takashi M, Hidemitsu I, Akinori U, Ikutu Y: Pulmonary resistance in dogs: A comparison of xenon with nitrous oxide. Can J Anaesth 1995; 42:547–53
17.
Calzia E, Stahl W, Handschuh T, Marx T, Froba G, Bader S, Georgieff M, Radermacher P: Respiratory mechanics during xenon anesthesia in pigs: Comparison with nitrous oxide. Anesthesiology 1999; 91:1378–86
18.
Mercier FJ, Naline E, Bardou M, Georges O, Denjean A, Benhamou D, Advemier C: Relaxation of proximal and distal isolated bronchi by halothane, isoflurane, and desflurane. Eur Respir J 2002; 20:286–92
19.
Goff MJ, Arain SR, Ficke DJ, Uhrich TD, Ebetr TJ: Absence of bronchodilation during desflurane anesthesia: A comparison to sevoflurane and thiopental. Anesthesiology 2000; 93:404–8
20.
Eshima RW, Mauer A, King T, Lin BK, Heavner JE, Bogezt MS, Kaye AD: A comparison of airway responses during desflurane and sevoflurane administration via  a laryngeal mask airway for maintenance of anesthesia. Anesth Analg 2003; 96:701–5