Background

High pressures exerted by balloons and cuffs of conventional endotracheal tubes, the Combitube (Tyco Healthcare Nellcor Mallinckrodt, Pleasanton, CA), the EasyTube (Teleflex Ruesch, Kernen, Germany), the Laryngeal Mask Airway (LMA North America, San Diego, CA), the Intubating Laryngeal Mask Airway (Fastrach; LMA North America), the ProSeal (LMA North America), and the Laryngeal Tube (LT; VBM Medizintechnik, Sulz, Germany) may traumatize the pharyngeal mucosa. The aim of this study was to compare pressures exerted on the pharyngeal, tracheal, and esophageal mucosa by different devices designed for securing the patient's airways.

Methods

Nineteen fresh cadavers were included. To measure mucosal pressures, microchip sensors were fixed on the anterior, lateral, and posterior surfaces of the proximal balloon and the distal cuff of the investigated devices. Depending on the respective airway device, the cuff volume was increased in 10-ml increments at the proximal balloon starting from 0 to a maximum of 100 ml, and in 2-ml increments at the distal cuff starting from 0 up to 12 ml.

Results

Tracheal mucosal pressures were significantly higher using the Combitube compared with the endotracheal tube and the EasyTube. Maximal esophageal pressures were significantly higher using the EasyTube compared with the Combitube. Using cuff volumes according to the manufacturers' guidelines, we found the highest pharyngeal pressures with the Intubating Laryngeal Mask Airway versus all other devices. At maximal volumes, the Laryngeal Mask Airway, the Intubating Laryngeal Mask Airway, and the ProSeal induced significantly higher pharyngeal pressures compared with all other devices. Using a pharyngeal cuff volume of 40 ml, the Intubating Laryngeal Mask Airway followed by the Laryngeal Mask Airway exerted significantly higher pressures compared with the other devices.

Conclusions

Although some devices exhibit a somewhat higher mucosal pressure when compared with others, the authors believe that the observed differences of the cuff pressures do not suggest a clinically relevant danger, because the investigated devices, except the endotracheal tubes, are not intended for prolonged use.

SOFT tissue injury is an important issue in airway management. Potential hazards are direct trauma, ranging from minor superficial laceration to perforations of trachea or esophagus, or indirect trauma by impaired mucosal blood flow. Mechanisms of injury are either the insertion process per se  or high pressures exerted by the artificial airways on the mucosa of surrounding anatomical structures.1The endotracheal tube (ETT) is acknowledged as the standard for tracheal intubation.2Nevertheless, under difficult airway conditions, other devices may be helpful to handle live threatening situations. In these cases, knowledge of the cuff volume and mucosal pressure relation can help to avoid severe mucosal damage of the upper airway region. Brimacombe et al.  3showed that inflation of a cuffed oropharyngeal airway with 40–60 ml air prevented blood flow to the posterior pharyngeal mucosa. The pharyngeal cuffs of other devices such as the esophageal tracheal Combitube® (ETC) are inflated with even higher volumes up to 85–100 ml air.1The primary aim of airway management is to provide a patent airway followed by ventilation of the lungs. The aim of this experimental cadaver investigation was to determine and to compare the pressures exerted on the pharyngeal, tracheal, and esophageal mucosa induced by seven airway devices to assess the clinical importance.

Research and ethical committee approval (Medical University Vienna, Vienna, Austria) was obtained. All patients, or their relatives, consented to postmortem research before the cadavers were used. In the morgue, we studied 19 fresh cadavers less than 24 h postmortem. The causes of death were sepsis (n = 4), bronchial carcinoma (n = 3), carcinoma of the colon (n = 3), myocardial infarction (n = 3), pneumonia (n = 2), pulmonary embolism (n = 2), carcinoma of the breast (n = 1), and cerebral hemorrhage (n = 1). None of the patients were intubated or resuscitated before death. Cadavers underwent laryngoscopic examination, and cadavers presenting with upper esophageal or laryngopharyngeal pathology were excluded. The following devices were investigated:

  1. Conventional endotracheal tubes (ETT; Tyco Healthcare Nellcor Mallinckrodt, Pleasanton, CA)

  2. Esophageal tracheal Combitube® (ETC; Tyco Healthcare Nellcor Mallinckrodt)

  3. EasyTube® (EzT; Teleflex Ruesch, Kernen, Germany)

  4. Laryngeal Mask Airway ™ (LMA ™; LMA North America, San Diego, CA)

  5. Intubating Laryngeal Mask ™ (ILMA ™, Fastrach ®; Lma North America)

  6. Proseal ™ laryngeal mask airway (PLMA ™; Lma North America)

  7. Laryngeal Tube Airway (LT; Vbm Medizintechnik, Sulz, Germany)

Each device was inserted in each cadaver. The insertion was randomized by a computer-generated code. Two experienced participants of the study (F.H., P.K.) with experience with all devices in at least 10 patients inserted the devices. We left the devices in place for 3 min for each measurement. The airway tips were placed in either the trachea (ETT, ETC, EzT) or the esophagus (ETC, EzT, LT) without inducing mucosal damage. Maximal esophageal pressures of ETC and EzT at a cuff inflation of 12 ml were measured. Mucosal pressures of proximal cuffs (ETC, EzT, LT) as well as of cuffs of the LMA ™, ILMA ™, and PLMA ™ were evaluated by inflation in 10-ml increments from 0 to 60 ml (LMA ™, ILMA ™) and 0 to 100 ml (ETC, EzT, PLMA ™, LT), respectively.

Mucosal pressure was assessed using a 1.2-mm-diameter strain gauge silicon microchip sensor (Codman MicroSensor; Johnson and Johnson Medical Ltd., Bracknell, United Kingdom), which was calibrated before its use. For each position of the respective device, a sensor was fixed to the surface of the proximal balloon and/or distal cuff with a clear adhesive dressing (thickness of 0.45 mm; Tegaderm; 3M, Ontario, Canada). The sensors were attached to the locations described in table 1. The sensing element of the sensor was orientated toward the mucosal surface. The ETT, ETC, EzT, and LT were inserted into the cadavers using a standard Macintosh laryngoscope No. 3. The correct position of the sensors and the integrity of the mucosa were checked before and after measurement by laryngoscopy. The LMA ™, ILMA ™, and PLMA ™ were inserted using the standard methods described by Brain et al.  4The airway sizes as well as the references are shown in table 2.

Table 1. Attachment of the Sensors to the Following Locations (Corresponding Mucosal Areas) 

Table 1. Attachment of the Sensors to the Following Locations (Corresponding Mucosal Areas) 
Table 1. Attachment of the Sensors to the Following Locations (Corresponding Mucosal Areas) 

Table 2. Characteristics and References of the Airway Devices Used 

Table 2. Characteristics and References of the Airway Devices Used 
Table 2. Characteristics and References of the Airway Devices Used 

Tracheal mucosal pressures exerted by the ETT cuff and the distal cuffs of the ETC and EzT were evaluated by cuff inflation from 0 to 12 ml in 2-ml increments. In addition, the mucosal pressures exerted by proximal cuffs were compared with respect to the amount of air recommended for emergency situations: 85 ml with the ETC,5–1080 ml with the EzT,1140 ml with the LMA ™,12–1540 ml with the ILMA ™,16,1740 ml with the PLMA ™,18,19and 80 ml with the LT.20In addition, a cuff volume of 40 ml was chosen for the ETC and EzT because there is a difference in the recommended filling volume between emergency intubation and elective use during general anesthesia: In elective anesthesia, the volume of the upper balloon can be reduced according to the minimal volume technique from 40 to 85 ml.21We compared the maximal tracheal pressures of the ETT, ETC, and EzT at 12 ml; maximal esophageal pressures of the ETC and EzT at 12 ml; and maximal pharyngeal pressures of the ETC, EzT, LMA ™, ILMA ™, PLMA ™, and LT at the respective recommended and highest volumes as well as at 40 ml.

Statistics

The sample size calculation was based on a type I error α of 0.008 corrected for multiple testing, a power of 0.8, a mean difference of 25%, and an SD within the treatment groups of σ= 0.2, resulting in a sample size of 17. The data were tested for normal distribution using the Shapiro-Wilk statistic (in case of a sample size less than 51) and the usual Kolmogorov D statistic otherwise.22For normally distributed data, we used the general linear model or otherwise the Kruskal-Wallis test to test for differences between various tubes and regions, respectively. The statistical analysis system was SAS 8.01 (SAS, Cary, North Carolina).

We compared the means of the highest pressures of the three locations (anterior, left lateral, and posterior) of all respective 19 measurements. For each comparison, we took out the maximal pressures of the three locations (anterior, lateral, posterior) of the respective airways. The box plots show the median and the quartiles of the respective pressures. A P  value of less than 0.05 was considered to be significant.

The mean ± SD (range) age, height, and weight of cadavers were 75 ± 11 (48–90) yr, 177 ± 8 (169–191) cm, and 85 ± 9 (72–97) kg, respectively. The male-to-female ratio was 14:5. Tracheal as well as esophageal placement of the airway tips was achieved in all cadavers on the first attempt.

Maximal tracheal pressures at a cuff volume of 12 ml were higher using the ETC compared with the ETT and EzT (fig. 1). Maximal esophageal pressures at a cuff volume of 12 ml were significantly higher using the EzT compared with the ETC (fig. 2). Assessing pharyngeal pressures using the ETC (with 85 ml), EzT (with 80 ml), LMA ™ (with 40 ml), ILMA ™ (with 40 ml), PLMA ™ (with 40 ml), and LT (with 80 ml) at those volumes recommended by the manufacturers, we found significantly higher pressures with the ILMA ™versus  all other devices (fig. 3). At maximal volumes (ETC [with 100 ml], EzT [with 100 ml], LMA ™[with 60 ml], ILMA ™[with 60 ml], PLMA ™[with 100 ml], and LT [with 100 ml]), the LMA ™, ILMA ™, and PLMA ™ induced higher pharyngeal pressures compared with all other devices (fig. 3). At a cuff volume of 40 ml, the ILMA ™ and LMA ™ induced higher pressures compared with all other devices (fig. 3).

Fig. 1. Box plots of maximal pressures in the trachea with 12 ml filling. x =  P < 0.05, ETC  versus EasyTube® (EzT) and endotracheal tube (ETT). ETC = Combitube®. 

Fig. 1. Box plots of maximal pressures in the trachea with 12 ml filling. x =  P < 0.05, ETC  versus EasyTube® (EzT) and endotracheal tube (ETT). ETC = Combitube®. 

Close modal

Fig. 2. Box plots of maximal pressures in the esophagus with 12 ml filling. x =  P < 0.05, EasyTube® (EzT)  versus Combitube® (ETC). 

Fig. 2. Box plots of maximal pressures in the esophagus with 12 ml filling. x =  P < 0.05, EasyTube® (EzT)  versus Combitube® (ETC). 

Close modal

Fig. 3. Box plots of pressures in the pharynx. (  A ) Box plots of pressures at the recommended cuff volumes in the pharynx. w =  P < 0.05,  Fastrach ™ intubating laryngeal mask airway (  ILMA ™)  versus all other devices. x =  P < 0.05, EasyTube® (EzT) and  Laryngeal Mask Airway ™ (  LMA ™)  versus Combitube® (ETC), Laryngeal Tube (LT), and  ProSeal ™ (  PLMA ™). y =  P < 0.05, ETC  versus LT and  PLMA ™. z =  P < 0.05, LT  versus PLMA ™. (  B ) Box plots of maximal pressures in the pharynx. x =  P < 0.05,  ILMA ™,  LMA ™, and  PLMA   versus EzT, ETC, and LT. (  C ) Box plots of pressures at 40-ml cuff volume in the pharynx. v =  P < 0.05,  ILMA   versus all other devices. w =  P < 0.05,  LMA   versus EzT, ETC, LT, and  PLMA ™. x =  P < 0.05, ETC  versus EzT, LT, and  PLMA ™. y =  P < 0.05, EzT  versus LT and  PLMA ™. z =  P < 0.05,  PLMA   versus  LT. 

Fig. 3. Box plots of pressures in the pharynx. (  A ) Box plots of pressures at the recommended cuff volumes in the pharynx. w =  P < 0.05,  Fastrach ™ intubating laryngeal mask airway (  ILMA ™)  versus all other devices. x =  P < 0.05, EasyTube® (EzT) and  Laryngeal Mask Airway ™ (  LMA ™)  versus Combitube® (ETC), Laryngeal Tube (LT), and  ProSeal ™ (  PLMA ™). y =  P < 0.05, ETC  versus LT and  PLMA ™. z =  P < 0.05, LT  versus PLMA ™. (  B ) Box plots of maximal pressures in the pharynx. x =  P < 0.05,  ILMA ™,  LMA ™, and  PLMA   versus EzT, ETC, and LT. (  C ) Box plots of pressures at 40-ml cuff volume in the pharynx. v =  P < 0.05,  ILMA   versus all other devices. w =  P < 0.05,  LMA   versus EzT, ETC, LT, and  PLMA ™. x =  P < 0.05, ETC  versus EzT, LT, and  PLMA ™. y =  P < 0.05, EzT  versus LT and  PLMA ™. z =  P < 0.05,  PLMA   versus  LT. 

Close modal

Other data are not shown for clearness; however, all results of the respective pressures are available on request. With a few exceptions, directly measured mucosal pressures showed significant differences (P < 0.0001) for the ETC, EzT, LMA ™, ILMA ™, PLMA ™, and LT, respectively, in all pharyngeal locations.

Directly measured mucosal pressures exerted by the EzT (cuff volume = 12 ml) on the esophageal mucosa were significantly different (P < 0.0001) in the anterior versus  lateral as well as anterior versus  posterior region. No differences were found for the ETC.

Directly measured mucosal pressures in the trachea at 12 ml showed significant differences (P < 0.0001) between all three locations with the ETT and EzT, and between anterior versus  lateral and lateral versus  posterior locations with the ETC (P < 0.0001). No mucosal laceration or tear was found.

One of the major advances in ETT design was the introduction of high-volume, low-pressure cuffs intended to minimize the pressure exerted on the tracheal mucosa.1Major attention is laid upon the fact that the mucosal pressure exerted by the cuff remains below the capillary mucosal pressure of approximately 26 mmHg. This concept has been incorporated into the new EzT airway, which has been equipped with a high-volume, low-pressure distal cuff system exhibiting significantly lower tracheal pressures. In our cadaver study, we found that pharyngeal, esophageal, and tracheal mucosal pressures increased with rising cuff volumes in all tested devices.

Tracheal Pressures Using ETT, ETC, and EzT

Regarding the ETC, high tracheal mucosal pressures are observed at the maximal recommended cuff volume of 12 ml (up to 41 cm H2O). In the clinical situation, much lower volumes are usually necessary to obtain an airtight seal. Therefore, blocking of the distal ETC cuff in the trachea without a cuff-pressure gauge is only recommended in emergency use. Pressures in the trachea registered for EzT were not significantly different from those observed with a regular ETT. With respect to location, the highest values were found in the lateral location with the ETT, in the posterior location with the EzT, and in the anterior position with the ETC. Although the findings with the ETC and ETT can be explained by the fact that the posterior membranous tracheal wall is more distensible than the cartilaginous anterolateral wall, we have no explanation for the higher posterior values with the EzT. We do not have data for the long-term use of the EzT up to now.

Esophageal Pressures Using ETC and EzT

Regarding esophageal placement, both the ETC and the EzT exert similar and relatively low pressures on the esophageal mucosa (maximum around 14 cm H2O). The highest values were found with the EzT when compared with the ETC. Although the same method was used, our data show lower values as compared with the measurements of Keller et al.  1Direct esophageal injury induced by an inflated cuff can therefore be certainly prevented by strictly adhering to the manufacturer’s guidelines. Furthermore, the danger of a permanent injury to the mucosa seems limited, because the investigated devices are not intended for prolonged use. The ETC may be used for up to 8 h.9 

Pharyngeal Pressures Using ETC, EzT, LMA ™, ILMA ™, PLMA ™, and LT

The LMA-Classic ™ exerts pressures of approximately 25–35 cm H2O on the pharyngeal or hypopharyngeal mucosa, equally distributed among anterior, lateral, and posterior locations up to a filling volume of 40 ml. Interestingly, even lower pressures were observed for the PLMA ™, with mucosal pressures mainly below 20 cm H2O, at least when maximally recommended filling volumes are observed. Pharyngeal mucosal perfusion is progressively reduced when mucosal pressures exceed 34 cm H2O.3Therefore, both devices are considered safe with respect to the pharyngeal mucosa. Significantly higher pressures, especially in the anterior location, were observed for the ILMA ™ (up to 41 cm H2O). However, this device is not intended for prolonged use, and its removal is recommended after successful tracheal intubation. Interestingly, the other supraglottic airways (ETC, EzT, and LT) exert pressures lower than that with the ILMA ™ and similar to those obtained for the LMA-Classic ™ (in the range of 23–36 cm H2O). Opposite to the findings of Keller et al. ,1our data show the highest values in the anterior pharyngeal location using the ETC. Similar results were found for the LMA ™, ILMA ™, and LT. In contrast, the highest values were found in the posterior location with the PLMA ™ and EzT. The latter results may be explained because the posterior surface is adjacent to rigid anatomical structures (vertebral bodies), whereas the anterolateral surface is adjacent to compliant anatomical structures (epiglottis, tongue, pharyngeal muscles, and membranous trachea).1We have no explanation for the different behaviors of the devices, except that the PLMA ™ may exert higher pressures in the posterior portion because of the additional cuff on the posterior part of the airway. With respect to pharyngeal mucosal pressures, we could not identify any traumatizing potential of these devices with respect to pharyngeal lacerations. The LMA ™ may be used for more than 3 h.13 

A limitation of the study is that mucosal pressures in awake and anesthetized patients may not be compared with the measurements in cadavers. Therefore, the muscle tone of patients is not taken into account. However, Keller et al.  1found that pharyngeal mucosal pressures were similar for cadavers and awake volunteers, which suggests that muscle tone does not play a major role in mucosal pressure measurements. Although their findings should be interpreted cautiously, they consider them to be applicable to the anesthetized patient for several reasons. First, there is evidence that pharyngeal compliance is similar in fresh cadavers and paralyzed anesthetized patients.3,23Second, the data for tracheal mucosal pressures in cadavers closely match those of a previous ETT study in paralyzed anesthetized patients,24and the data for pharyngeal mucosal pressures closely match those of two previous studies of the cuffed oropharyngeal airway in paralyzed anesthetized patients.25 

In summary, our data demonstrate that increasing cuff volumes are directly transferred into increased mucosal pressures up to levels potentially exceeding mucosal perfusion pressure. Those mucosal pressures are seldom observed by limiting the cuff volumes following the minimal volume technique. Recommended volumes should be strictly followed because slightly overextended cuffs may exert high pressures, e.g. , inflating a size 5 LMA ™, PLMA ™, or ILMA ™ with 60 ml air can result in pharyngeal mucosal pressures of up to 60 cm H2O. The observed differences between the devices do not seem to be clinically relevant and do not preclude the use of any device in clinical routine cases, because the devices investigated (except the ETT) are not intended for prolonged use (for more than 8 h). Esophageal mucosal pressures with the ETC and EzT in the esophageal position remain low when maximal filling volumes recommended by the manufacturers are respected. We therefore recommend adherence to the manufacturers’ guidelines and the use of a pressure gauge for cuffing all devices.

1.
Keller C, Brimacombe J, Boehler M, Loeckinger A, Puehringer F: The influence of cuff volume and anatomic location on pharyngeal, esophageal, and tracheal mucosal pressures with the esophageal tracheal Combitube. Anesthesiology 2002; 96:1074–7
2.
Adams AB: Catheters for tracheal gas insufflation. Respir Care 2001; 46:177–84
3.
Brimacombe J, Keller C, Puehringer F: Pharyngeal mucosal pressure and perfusion: A fiberoptic evaluation of the posterior pharynx in anesthetized adult patient with a modified, cuffed oropharyngeal airway. Anesthesiology 1999; 91:1661–5
4.
Brain AI, Verghese C, Addy EV: The intubating laryngeal mask: II. A preliminary clinical report of a new means of intubating the trachea. Br J Anaesth 1997; 79:704–9
5.
Gaitini LA, Vaida SJ, Mostafa S, Yanovski B, Croitoru M, Capdevilla MD, Sabo E, Ben-David B: The Combitube in elective surgery: A report of 200 cases. Anesthesiology 2001; 94:79–82
6.
Frass M: The Combitube: Esophageal/tracheal double lumen airway, Airway Management: Principles and Practice. Edited by Benumof JL. St. Louis, Mosby, 1996, pp 444–54Benumof JL
St. Louis
,
Mosby
7.
Walz R, Davis S, Panning B: Is the Combitube a useful emergency airway device for anesthesiologists? (letter). Anesth Analg 1999; 88:227–34
8.
American Heart Association (AHA) Emergency Cardiac Care (ECC) Guidelines, Part 6: Advanced Cardiovascular Life Support, Section 3: Adjuncts for Oxygenation, Ventilation, and Airway Control. Airway Adjuncts. Circulation 2000; 102:I-95–104
9.
Rich JM, Mason AM, Bey TA, Krafft P, Frass M: The critical airway, rescue ventilation, and the Combitube: I and II. AANA J 2004; 72:17–27, 115–124
10.
Krafft P, Nikolic A, Frass M: Esophageal rupture associated with the use of the Combitube (letter). Anesth Analg 1998; 87:1457
11.
Thierbach AR, Piepho T, Maybauer MO: A new device for emergency airway management: The EasyTube. Resuscitation 2004; 60:347
12.
Brimacombe JR: Seal with the respiratory and gastrointestinal tract, Laryngeal Mask Anesthesia: Principles and Practice, 2nd edition. Edited by Brimacombe JR. Philadelphia, Saunders, 2005, pp 137–52Brimacombe JR
Philadelphia
,
Saunders
13.
Verghese C, Berlet J, Kapila A, Pollard R: Clinical assessment of the single use laryngeal mask airway: The LMA-Unique. Br J Anaesth 1998; 80:677–9
14.
Brimacombe J, Keller C: Laryngeal mask airway size selection in males and female Ease of insertion, oropharyngeal leak pressure, pharyngeal mucosal pressures and anatomical position. Br J Anaesth 1999; 82:703–7
15.
Keller C, Pühringer F, Brimacombe J: Influence of cuff volume on oropharyngeal leak pressure and fibreoptic position with the laryngeal mask airway. Br J Anaesth 1998; 81:186–7
16.
Minville V, N’guyen L, Coustet B, Fourcade O, Samii K: Difficult airway in obstetric using ILMA-Fastrach. Anesth Analg 2004; 99:1873
17.
Hsin ST, Chen CH, Juan CH, Tseng KW, Oh CH, Tsou MY, Tsai SK: A modified method for intubation of a patient with ankylosing spondylitis using intubating laryngeal mask airway (LMA-Fastrach): A case report. Acta Anaesthesiol Sin 2001; 39:179–82
18.
Bein B, Carstensen S, Gleim M, Claus L, Tonner PH, Steinfath M, Scholz J, Dörges V: A comparison of the ProSeal laryngeal mask airway, the laryngeal tube S and the oesophageal-tracheal Combitube during routine surgical procedures. Eur J Anaesthesiol 2005; 22:341–6
19.
Brimacombe J, Keller C: The ProSeal Laryngeal Mask Airway. Anesthesiology 2000; 93:104–9
20.
Komatsu R, Nagata O, Kamata K, Yamagata K, Sessler DI, Ozaki M: Comparison of the intubating laryngeal mask airway and laryngeal tube placement during manual in-line stabilisation of the neck. Anaesthesia 2005; 60:13–7
21.
Urtubia RM, Aguila CM, Cumsille MA: Combitube: A study for proper use. Anesth Analg 2000; 90:958–62
22.
Sachs L: Der Kolmogoroff-Smirnov-Test für die Güte der Anpassung, Angewandte Statistik. Berlin Springer Verlag, 1992, pp 426–30
Berlin
,
Springer Verlag
23.
Brimacombe J, Keller C, Gunkel AR, Puehringer F: The influence of the tonsillar gag on efficacy of seal, anatomic position, airway patency and airway protection with the flexible laryngeal mask airway: A randomized, crossover study of fresh, adult cadavers. Anesth Analg 1999; 89:181–6
24.
Brimacombe J, Keller C, Giampalmo M, Sparr HJ, Berry A: Direct measurement of mucosal pressures exerted by cuff and non-cuff portions of tracheal tubes with different cuff volumes and head and neck positions. Br J Anaesth 1999; 82:708–11
25.
Keller C, Brimacombe J: Mucosal pressures from the cuffed oropharyngeal airway versus  the laryngeal mask airway. Br J Anaesth 1999; 82:922–4