Background

The aim of this study was to evaluate the use of a glucose meter with surgical patients under general anesthesia in the operating room.

Methods

Glucose measurements were performed intraoperatively on 368 paired capillary and arterial whole blood samples using a Nova StatStrip (Nova Biomedical, USA) glucose meter and compared with 368 reference arterial whole blood glucose measurements by blood gas analyzer in 196 patients. Primary outcomes were median bias (meter minus reference), percentage of glucose meter samples meeting accuracy criteria for subcutaneous insulin dosing as defined by Parkes error grid analysis for type 1 diabetes mellitus, and accuracy criteria for intravenous insulin infusion as defined by Clinical and Laboratory Standards Institute guidelines. Time under anesthesia, patient position, diabetes status, and other variables were studied to determine whether any affected glucose meter bias.

Results

Median bias (interquartile range) was −4 mg/dl (−9 to 0 mg/dl), which did not differ from median arterial meter bias of −5 mg/dl (−9 to −1 mg/dl; P = 0.32). All of the capillary and arterial glucose meter values met acceptability criteria for subcutaneous insulin dosing, whereas only 89% (327 of 368) of capillary and 93% (344 of 368) arterial glucose meter values met accuracy criteria for intravenous insulin infusion. Time, patient position, and diabetes status were not associated with meter bias.

Conclusions

Capillary and arterial blood glucose measured using the glucose meter are acceptable for intraoperative subcutaneous insulin dosing. Whole blood glucose on the meter did not meet accuracy guidelines established specifically for more intensive (e.g., intravenous insulin) glycemic control in the acute care environment.

What We Already Know about This Topic
  • Diabetes mellitus and subsequent glycemic control are increasingly important issues in the perioperative period

  • Few studies of glucose meters have been performed, and there is a need for further study of accuracy and utility

What This Article Tells Us That Is New
  • The authors have performed a study in the operating room of new glucose meter technology with hematocrit and interference correction on capillary and arterial blood samples in comparison with reference arterial whole blood glucose measurements by a blood gas analyzer

  • The authors found that arterial and capillary blood glucose measurements with the glucose meter did not meet established guidelines for intensive blood glucose control, suggesting caution be exercised when using glucose meters for intravenous glycemic control protocols

ACCORDING to the Centers for Disease Control and Prevention, approximately 9.3% of the U.S. population has diabetes mellitus, with 27% of the population being undiagnosed.1  Although glycemic control for critically ill patients remain controversial,2–4  glycemic control using insulin in patient with diabetes mellitus undergoing surgery has become an increasingly common quality metric.5  Treatment of hyperglycemia requires monitoring of blood glucose, because even a single episode of severe hypoglycemia (glucose less than 40 mg/dl) may increase the risk of death in the hospital up to twofold.6  Guidelines recommend glucose monitoring for patients under general anesthesia if the patient receives insulin, the procedure is longer than 1 to 2 h, or if there is concern for hyperglycemia or hypoglycemia.7,8 

Glucose monitoring can be performed using laboratory serum or plasma glucose analysis, whole blood glucose by blood gas analyzer, or whole blood glucose by glucose meter. Although laboratory and blood gas analysis methods require an arterial or venous blood specimen, most glucose meters can also analyze capillary (via fingerstick) blood. Glucose analysis by laboratory methods is more time consuming and expensive than glucose meter testing and may not provide timely results for dosing insulin or treating hyperglycemia or hypoglycemia in the operating room (OR).

The correlation between results from glucose meters and laboratory glucose measurements varies between meter technologies,9  and correlation in the hypoglycemic and hyperglycemic ranges is poor for some meters.10,11  Although newer technologies may provide more accuracy,12–14  there is still concern about the use of meters in the critically ill.15,16  Studies have demonstrated potentially dangerous discrepancies in capillary glucose measurement in patients on vasopressor therapy,10,17  patients in shock18  or with poor tissue perfusion,19  and in other critically ill patient populations.20  Consequently, no glucose meter is currently approved by the U.S. Food and Drug Administration (FDA) for use with capillary samples in critically ill hospitalized patients.

Although the FDA has not provided a definition of what constitutes “critically ill” for purposes of capillary glucose testing, previous studies21  have shown that factors expected to change rapidly during anesthesia and surgery can affect glucose meter accuracy. These factors include hematocrit, blood pressure, peripheral perfusion, temperature, anesthesia technique, medications, Po2, and pH. Vasopressor therapy has been associated with glucose meter outliers in the intensive care unit. Therefore, patients receiving vasopressors intraoperatively may also be at risk of glucose meter errors or outliers. Despite these concerns, very few studies of glucose meter accuracy in the OR have been performed.

The aim of this study was to assess the accuracy of a newer glucose meter technology (Nova StatStrip, Nova Biomedical Corp., USA) with hematocrit and interference correction in the OR with patients under general anesthesia. We hypothesized that capillary and arterial whole blood samples analyzed on the glucose meter would meet established accuracy criteria for safe and effective insulin dosing. We also studied the impact of patient position, time under general anesthesia (early vs. late paired measurements), diabetes status, and other clinical and laboratory variables on the relationship between glucose meter and reference glucose concentrations.

Patient Selection

Members of the Anesthesia Clinical Research Unit performed a manual scan of the following day’s surgical listing for all patients scheduled for thoracic, vascular, or neurologic surgery from August 2014 to March 2015 (N = 234 screened). Study participants had to be 18 yr or older, have a preoperative hemoglobin greater than 10 mg/dl, and speak English. A diagnosis of diabetes mellitus was neither an inclusion or exclusion criteria. Glucose testing was performed only on patients who had an arterial catheter placed. The decision to place an arterial catheter was at the discretion of the anesthesia provider at the time of surgery, and the anesthesia provider was not aware of the patient’s study participation. The study design was approved by the Mayo Clinic Institutional Review Board, Rochester, Minnesota, and written informed consent was obtained from all of the subjects.

Sample Collection and Glucose Analysis

Sample collection was performed by the Mayo Clinic Clinical Trials Research Unit, Rochester, Minnesota. Members of Clinical Trials Research Unit nursing staff came to the OR approximately 30 min after the arterial catheter placement and obtained the first set of samples. Each set included the following: (1) a capillary fingerstick sample tested on the glucose meter; (2) an arterial whole blood sample tested on the glucose meter; and (3) an arterial whole blood sample tested in the laboratory on a blood gas analyzer (Radiometer ABL90, Radiometer America, Inc., USA). The arterial whole blood used for samples (2) and (3) were from the same blood draw and were collected in a 3-ml lithium heparin blood gas syringe. The arterial glucose result from the laboratory blood gas analyzer was considered the reference value. Arterial blood samples were obtained within 10 min of capillary sampling, and laboratory blood glucose analysis was performed within 10 min of arterial whole blood collection. Staff from the Clinical Trials Research Unit obtained a second set of samples approximately 60 min after the first set. All of the patients were under general anesthesia at the time of both glucose sample collections. Capillary samples from patients in the lateral position were obtained from a finger on the lower (dependent) hand. Staff from the Clinical Trials Research Unit were trained on glucose meter operation and capillary sampling according to manufacturer’s instructions and laboratory procedures. Blood gas analyzer testing was performed by trained technologists in the testing laboratory, and the blood gas data, including glucose, were available to providers for clinical use.

Perioperative Data

Subject data, including temperature, systolic and diastolic blood pressures, mean arterial pressure, heart rate, as well clinical variables and patient demographic information, were extracted using a customized, integrative relational research database that contains a near–real-time copy of clinical, administrative, and environmental exposure data from the Electronic Medical Record, Hospital Surgical Listing, and Mayo Clinic Life Sciences System databases.22,23  All of the intraoperative variables were electronically captured with 3-min resolution. Additional laboratory data (hemoglobin, pH, partial pressure of carbon dioxide [PCO2], and Po2) were obtained from the sample submitted for reference glucose determination on the blood gas analyzer or retrieved from clinical databases. Manual review of the data for fidelity was performed with removal of obvious errors resulting from measurement techniques.

Statistical Analysis

Primary outcomes for the study were median bias and percentage of glucose meter samples meeting accuracy criteria for subcutaneous and intravenous insulin infusion.

Median (interquartile range [IQR]) bias between capillary glucose meter and reference glucose concentration and median (IQR) bias between arterial glucose meter and reference glucose concentration were calculated (meter glucose minus reference glucose). Univariate linear regression with generalized estimating equation (GEE) analysis was used to assess for differences in median bias between capillary bias and arterial bias while accounting for the correlation between up to two observations per patient. To assess whether time under general anesthesia impacted glucose meter accuracy, linear regression with GEE was also used to determine whether glucose meter bias differed between samples collected early during surgery (within 30 min after arterial catheter placement) versus later during surgery (approximately 60 min after the initial capillary sample). We also performed linear regression with GEE to determine whether limb position during surgery (lateral vs. supine) affected capillary glucose bias.

Clinical concordance between glucose meter and reference glucose values for subcutaneous insulin dosing was assessed using the Parkes error grid for type 1 diabetes mellitus,24  developed by a consensus of endocrinologists to define glucose meter accuracy needed for safe and effective subcutaneous insulin dosing. The Parkes error grid displays the relationship between glucose meter and reference glucose by use of an error grid divided into five risk zones. Zone A represents no effect on clinical action (glucose meter and reference glucose value would lead to the same clinical action). Zone B represents altered clinical action with little or no effect on clinical outcome. Zone C represents altered clinical action likely to impact clinical outcome; zone D represents altered clinical action that could have significant medical risk; and zone E represents altered clinical action that could have dangerous consequences.24  Consistent with the ISO 15197:2013 guideline used by glucose meter manufacturers,25  we defined acceptable accuracy for subcutaneous insulin dosing as 99% of glucose meter values within zones A and B on the Parkes error grid.

We defined accuracy requirements for more intensive glucose monitoring (e.g., that required for intravenous insulin dosing) using the Clinical and Laboratory Standards Institute (CLSI) POCT12-A3 guideline.26  CLSI POCT12-A3 is a consensus guideline developed to define accuracy requirements for acute care use of glucose meters (including intravenous insulin administration). It states that 95% of glucose meter results should be within ±12 mg/dl of reference glucose value for reference glucose values less than 100 mg/dl and ±12.5% for reference glucose of 100 mg/dl or more. In addition, the guidelines require that no more than 2% of results contain error exceeding ±15 mg/dl (for reference glucose less than 75 mg/dl) or 20% (for reference glucose of 75 mg/dl or more).26 

CLSI POCT12-A3 guidelines suggest a minimum of 200 measurements from 100 samples be measured to determine accuracy in a given patient population. In previous studies we were able to differentiate glucose meter bias between capillary and arterial samples with 100 samples.27  We used the CLSI POCT12-A3 guidelines and previous experience to determine that 300 or more measurements from 150 or more patients would allow us to differentiate glucose meter bias in capillary and arterial samples.

Univariate linear regression with GEE was used to perform post hoc statistical analysis to determine whether clinical variables such as age, sex, diabetes status, disease severity (age-adjusted Charlson comorbidity index28 ), body mass index, temperature, blood pressure (systolic, diastolic, and mean arterial), or heart rate impacted the relationship (bias) between either capillary or arterial glucose meter and reference glucose concentration (secondary outcomes). For age-adjusted Charlson comorbidity index,28  we compared glucose meter bias in patients with few comorbidities (index 0 to 3) with patients with more comorbidities (index 4 to 6). For blood pressure, the average systolic, diastolic, and mean arterial pressures in the 15 min before a capillary sampling (measurements were recorded every 3 min) were used to determine whether blood pressure impacted the accuracy of capillary or arterial glucose meter measurement. A similar approach was used for heart rate and temperature (average value within 15 min of capillary sampling). Before the study we planned the use of univariate analysis with GEE to determine whether laboratory variables including pH, hemoglobin, Po2, and PCO2 affected glucose meter bias (secondary outcomes). All of the tests were two sided, and P < 0.05 was considered statistically significant. SAS version 9.4M3 (SAS Institute, Inc., USA) was used for all of the analyses.

Participants

A total of 235 subjects were assessed for study eligibility. Fifteen refused to participate, one had a starting hemoglobin less than 10 mg/dl, and one did not speak English. Of the 218 subjects who were consented, 22 subjects did not have at least one pair of capillary and arterial samples drawn (arms not accessible, arterial catheter not placed, or surgery cancelled) or were missing one or more glucose values for comparison. Thus, 196 patients had one or more complete sets of samples obtained for analysis. A total of 192 complete sets were obtained for analysis at the first blood draw (within 30 min of arterial catheter placement), whereas 176 complete second sample sets (approximately 1 h after first sample) were obtained. Characteristics of the participants studied and the surgical procedures, including positioning, fluid administration, and vasopressor use, are provided in table 1.

Median Bias and Accuracy Criteria

Median (IQR) bias among the 192 capillary samples collected within 30 min of arterial catheter placement was −4 mg/dl (−8 to −1 mg/dl), whereas median (IQR) capillary bias among the 176 samples collected approximately 60 min after the first sampling was −5 mg/dl (−10 to 0 mg/dl; P = 0.85, no difference in median bias between first and second collection). Thus, time under anesthesia (time in surgery) did not affect the accuracy of capillary sampling for glucose. Time under anesthesia also did not impact the accuracy of arterial sampling (data not shown). We therefore combined early and late measurements for all of the subsequent statistical analysis (n = 368 for all subsequent analysis).

The glucose meter slightly but systematically underestimated glucose concentrations using both capillary and arterial samples. Median (IQR) bias between capillary glucose meter and reference glucose was −4 mg/dl (−9 to 0 mg/dl), whereas median (IQR) bias between arterial glucose meter and reference glucose was −5 mg/dl (−9 to −1 mg/dl; P = 0.32, no significant difference between capillary and arterial meter bias). Parkes error grid analysis for type 1 diabetes mellitus was used to assess the clinical concordance of glucose meter values (both capillary and arterial) for subcutaneous insulin dosing. All of the capillary and arterial glucose meter values were within zones A and B of the Parkes error grid (i.e., 100% of meter results met accuracy criteria), with the majority falling into zone A (fig. 1).

Capillary glucose bias ranged from −44 to 19 mg/dl, whereas arterial glucose bias ranged from −45 to 15 mg/dl (fig. 2). A total of 327 (89%) of 368 capillary and 344 (93%) of 368 arterial glucose meter values met CLSI POCT12-A3 accuracy guidelines for intravenous insulin dosing (±12 mg/dl at reference glucose less than 100 mg/dl and ±12.5% at reference glucose of 100 mg/dl or more). Glucose meter “outlier” results, those exceeding 20% of the reference glucose value, were observed with 7 (2%) of 368 of both capillary and arterial glucose meter values.

Impact of Limb Position on Capillary and Arterial Bias

A total of 112 patients had 213 capillary meter measurements during surgery with limbs in the lateral position, whereas 78 patients had 144 capillary measurements in the supine position, and 6 patients had 11 measurements in the prone position. Median (IQR) capillary glucose bias for measurements in the lateral position was −4 mg/dl (−9 to 0 mg/dl) compared with −4 mg/dl (−9 to 0 mg/dl) for measurements in the supine position (P = 0.92). Analysis was not conducted for measurements in the prone position due to the small number of measurements. Similarly, limb position did not impact arterial glucose meter bias (data not shown).

Impact of Clinical Variables on Capillary and Arterial Glucose Meter Measurement

We performed univariate linear regression with GEE to determine whether age, sex, diabetes status, disease severity as measured by age-adjusted Charlson comorbidity index, body mass index, blood pressure, or heart rate impacted the relationship between capillary or arterial whole blood (glucose meter) and reference glucose concentration. Among clinical variables, only age-adjusted Charlson comorbidity index demonstrated a statistically significant relationship with capillary glucose meter bias (P = 0.04). However, the magnitude (slope estimate) of that effect was small (1.9 mg/dl increase in bias for patients with higher risk scores; table 2). Diabetes status did not impact either capillary or arterial glucose meter bias (table 2). For arterial whole blood glucose measured on the meter, only mean diastolic blood pressure in the 15 min before testing had a significant effect on glucose meter bias, but again the magnitude of the effect was small (0.7 mg/dl per 10-mmHg change in diastolic blood pressure; P = 0.03; table 2).

Among laboratory variables, only PCO2 significantly impacted the relationship between capillary glucose meter and reference glucose (P = 0.04). As with clinical variables, the magnitude (slope estimate) of these effects was small (1.5 mg/dl change in bias per 10-mmHg PCO2; table 3). There were no laboratory variables that impacted the relationship between arterial glucose meter and reference glucose concentration (table 3).

The use of glucose meters in critically ill patients remains controversial. A number of variables, including hematocrit, Po2, PCO2, pH, and medications, may interfere with or impact the accuracy of glucose meters.9,29–31  The accuracy of capillary (fingerstick) glucose measurement in critically ill patients is of even greater concern, because previous studies have demonstrated potentially dangerous discrepancies in capillary glucose measurement in patients on vasopressor therapy,10,17  patients in shock,18  or with poor tissue perfusion,19  as well as other critically ill patient populations.20  Studies have also found that systematic (mean or median) glucose meter bias differs for capillary glucose meter samples compared with arterial or venous whole blood samples.10,27,32 

Many of the studies demonstrating limitations to glucose meters were performed with older glucose meter technologies that did not correct for hematocrit or other interferences. Newer meter technologies may provide more accurate glucose measurement12–14  when used on critically ill patients, although studies have not specifically addressed capillary sampling or the use of meters intraoperatively. One previous study33  specifically examined glucose measurements in the OR, although few patients were under anesthesia at the time of sample collection. Their data suggest that performance of glucose meters in the OR could be much worse than in the intensive care unit. That study compared glucose meter and central laboratory measurements that were collected within 5 min of each other, using both arterial and venous blood samples.

We found that median bias did not differ between capillary and arterial blood specimens tested using the blood glucose meter. Outliers (glucose meter value greater than 20% different from reference) were also rare with both capillary and arterial glucose meter samples. Thus, we observed improved accuracy of glucose meter measurement compared with previous reports, especially for capillary whole blood glucose. It is important to interpret studies of glucose meter accuracy in the context of the generation and type of meter technology used. For both capillary and arterial samples, 100% of values were within zones A and B on the Parkes error grid for type 1 diabetes mellitus, indicating excellent clinical concordance between glucose meter and reference glucose values for subcutaneous insulin dosing. We conclude that improved accuracy (reduced systematic bias and number/percentage of outliers) allows for safe subcutaneous insulin dosing in the OR using both capillary and arterial samples on the glucose meter.

The glucose meter accuracy required for more intensive glycemic control protocols remains controversial. We assessed glucose meter accuracy for more intensive glycemic control protocols using the CLSI POCT12-A3 guidelines. Although Parkes error grid analysis was developed to reflect accuracy required for self-monitoring of blood glucose (subcutaneous insulin dosing), CLSI POCT12-A3 guidelines were developed to reflect accuracy requirements for various uses of glucose meters in the acute care environment. Specifically, the CLSI POCT12-A3 guidelines comment on required accuracy for intravenous insulin glycemic control protocols.26  Compared with glucose correction using subcutaneous insulin, glycemic control protocols using intravenous insulin increase the rate of hypoglycemia in the intensive care unit up to fivefold.34  Error simulation models suggest that glucose meters that exceed 10 to 15% total error allow for large insulin dosing errors (those most likely to result in hypoglycemia) during intravenous insulin therapy using common protocols.35–37  Finally, both error simulation models and empiric data demonstrate that, as glucose meter error increases above 10 to 15%, there is a decrease in the efficacy of glycemic control (increased rate of hypoglycemia and hyperglycemia and increased glycemic variability) during intravenous insulin infusion.38,39  These and other justifications have been used to define more stringent accuracy requirements for use of glucose meters for intravenous insulin infusion in the acute care setting.

A total of 344 (93%) of 368 of arterial whole blood glucose values met CLSI POCT12-A3 guidelines. Thus, with arterial samples, the glucose meter just failed to meet CLSI POCT12-A3 accuracy requirements (95%). Only 368 (89%) of capillary whole blood glucose measurements met these same accuracy requirements. The large number of glucose meter measurements that underestimated true glucose by greater than 12.5% would be expected to impact glycemic control efficacy (rates of hyperglycemia and glycemic variability) in the context of intravenous glycemic control protocols. Thus, caution should be exercised when using glucose meters for this purpose intraoperatively, and capillary sampling is best avoided when using meters for more intensive glycemic control protocols.

A secondary aim of our study was to identify clinical and laboratory variables associated with glucose meter bias and/or outliers. We found a statistically significant relationship between disease severity (age-adjusted Charlson comorbidity index) and capillary glucose meter bias, although the magnitude of this effect was small. PCO2 also had a statistically significant but small impact on capillary glucose bias. For arterial glucose meter samples, only mean diastolic blood pressure in the 15 min before sampling was associated with glucose meter bias, but again the magnitude of this effect was small. Patient position, time under general anesthesia (early vs. late measurement), diabetes status, and other clinical and laboratory variables were not significantly associated with glucose meter bias. Likely because we observed less systematic bias and fewer outliers compared with previous studies, we were not able to identify clinical or laboratory variables that had a major effect on glucose meter performance.

The Nova StatStrip device FDA labeling cautions against the use of capillary samples for some intensively treated patients but does not specify exactly which patients should not have capillary sampling performed. The FDA has not given glucose meter manufacturers clear guidance on the number of different patient populations (e.g., surgical, surgical intensive care unit, medical intensive care unit) or number of patients who should be evaluated to determine the effectiveness of capillary glucose sampling for intensively treated patients. Accuracy criteria have also not been clearly defined by the FDA, although criteria close to those defined in CLSI POCT12-A3 have been suggested.

Patients in the OR share many characteristics with other intensively treated patient populations, including acute physiologic alterations to blood pressure, Po2, PCO2, and pH that are associated with glucose meter error. Surgical patients often receive vasopressors and other intravenous fluids and medications commonly used among critically ill or intensively treated patients. These factors, along with the paucity of published data on intraoperative glucose meter accuracy, make it unclear whether capillary sampling is appropriate in the OR. Many patients in our study received vasopressor therapy, and patients had a wide range of comorbidities (age-adjusted Charlson risk score; table 1). The results of this study will therefore help users evaluate the safety and efficacy of capillary sampling for either subcutaneous or intravenous insulin dosing during surgery. The results may also prove useful for glucose meter manufacturers in demonstrating the accuracy of capillary glucose meter testing in one intensively treated patient population.

Our study was limited in that we only studied adult patients during thoracic, vascular, and neurologic surgery and studied only a single capillary sampling site (finger). Patients were selected for the study based on availability of an arterial catheter for sampling, rather than a specific comorbidity or condition associated with glucose meter inaccuracy. In addition, no patients had hypoglycemia (reference glucose less than 70 mg/dl). We have observed similar bias for glucose values less than 70 mg/dl in previous studies using this device.37  All of the patients were under general anesthesia at the time of capillary sampling, which may have led to more accurate capillary glucose measurement due to the vasodilating effects of general anesthesia. Finally, we tested only one point of care glucose meter model. The results obtained in this study may not be generalizable to all glucose meters, especially older technologies. Additional studies are necessary to determine whether glucose meter accuracy may differ in other types of surgery or other acute care patient populations, such as patients in intensive care units.

Conclusions

The median bias for arterial whole blood glucose meter samples collected intraoperatively was –5 mg/dl, which did not differ from a median bias of −4 mg/dl for capillary glucose meter samples. In contrast to findings of previous studies that used older glucose meter technologies, we found that neither the systematic (median) bias nor the number/percentage of outlier results differed between capillary and arterial glucose meter samples. One hundred percent of both capillary and arterial glucose values were within zones A and B on the Parkes error grid for type 1 diabetes mellitus, demonstrating that both capillary and arterial whole blood glucose can be used to safely dose subcutaneous insulin in the OR. Neither arterial nor capillary whole blood glucose met CSLI POCT12-A3 guidelines, suggesting that caution must be exercised when using glucose meter values for intravenous or more intensive glycemic control protocols.

The authors acknowledge Mayo Clinic Anesthesia Clinical Research Unit study coordinators Wayne Weber, R.R.T., and Melissa Passe, R.R.T., for their help with data extraction; Man Li, Lavonne M. Liedl, R.R.T., and the staff of the Mayo Clinic Anesthesia Clinical Research Unit for their efforts in identifying, recruiting, and consenting participants; Sarah Wolhart, R.N., for help in preparing the protocol; and the staff of the Mayo Clinic Clinical Research and Trials Unit for performing point of care measurements.

Supported by grant No. UL1 TR000135 from the National Center for Advancing Translational Sciences, Bethesda, Maryland, the Mayo Clinic Department of Laboratory Medicine and Pathology, Rochester, Minnesota, and the Mayo Clinic Department of Anesthesiology, Rochester, Minnesota.

Dr. Karon has received travel support from Nova Biomedical Corporation, Waltham, Massachusetts. The other authors declare no competing interests.

1.
Centers for Disease Control and Prevention
:
2014 National Diabetes Statistics Report.
2.
NICE-SUGAR Study Investigators
:
Intensive versus conventional glucose control in critically ill patients.
N Engl J Med
2009
;
360
:
1283
97
3.
Lewis
KS
,
Kane-Gill
SL
,
Bobek
MB
,
Dasta
JF
:
Intensive insulin therapy for critically ill patients.
Ann Pharmacother
2004
;
38
:
1243
51
4.
van den Berghe
G
,
Wouters
P
,
Weekers
F
,
Verwaest
C
,
Bruyninckx
F
,
Schetz
M
,
Vlasselaers
D
,
Ferdinande
P
,
Lauwers
P
,
Bouillon
R
:
Intensive insulin therapy in critically ill patients.
N Engl J Med
2001
;
345
:
1359
67
5.
Kwon
S
,
Thompson
R
,
Dellinger
P
,
Yanez
D
,
Farrohki
E
,
Flum
D
:
Importance of perioperative glycemic control in general surgery: A report from the Surgical Care and Outcomes Assessment Program.
Ann Surg
2013
;
257
:
8
14
6.
Krinsley
JS
,
Grover
A
:
Severe hypoglycemia in critically ill patients: Risk factors and outcomes.
Crit Care Med
2007
;
35
:
2262
7
7.
Dhatariya
K
,
Levy
N
,
Kilvert
A
,
Watson
B
,
Cousins
D
,
Flanagan
D
,
Hilton
L
,
Jairam
C
,
Leyden
K
,
Lipp
A
,
Lobo
D
,
Sinclair-Hammersley
M
,
Rayman
G
;
Joint British Diabetes Societies
:
NHS diabetes guideline for the perioperative management of the adult patient with diabetes.
Diabet Med
2012
;
29
:
420
33
8.
Joshi
GP
,
Chung
F
,
Vann
MA
,
Ahmad
S
,
Gan
TJ
,
Goulson
DT
,
Merrill
DG
,
Twersky
R
;
Society for Ambulatory Anesthesia
:
Society for Ambulatory Anesthesia consensus statement on perioperative blood glucose management in diabetic patients undergoing ambulatory surgery.
Anesth Analg
2010
;
111
:
1378
87
9.
Karon
BS
,
Griesmann
L
,
Scott
R
,
Bryant
SC
,
Dubois
JA
,
Shirey
TL
,
Presti
S
,
Santrach
PJ
:
Evaluation of the impact of hematocrit and other interference on the accuracy of hospital-based glucose meters.
Diabetes Technol Ther
2008
;
10
:
111
20
10.
Kanji
S
,
Buffie
J
,
Hutton
B
,
Bunting
PS
,
Singh
A
,
McDonald
K
,
Fergusson
D
,
McIntyre
LA
,
Hebert
PC
:
Reliability of point-of-care testing for glucose measurement in critically ill adults.
Crit Care Med
2005
;
33
:
2778
85
11.
Khan
AI
,
Vasquez
Y
,
Gray
J
,
Wians
FH
Jr
,
Kroll
MH
:
The variability of results between point-of-care testing glucose meters and the central laboratory analyzer.
Arch Pathol Lab Med
2006
;
130
:
1527
32
12.
Karon
BS
,
Blanshan
CT
,
Deobald
GR
,
Wockenfus
AM
:
Retrospective evaluation of the accuracy of Roche AccuChek Inform and Nova StatStrip glucose meters when used on critically ill patients.
Diabetes Technol Ther
2014
;
16
:
828
32
13.
Louie
R
,
Curtis
C
,
Toffaletti
J
,
Handel
E
,
Slingerland
R
,
Fokkert
M
,
Muller
W
,
Weinert
S
,
Lee
D
,
Kotagiri
S
:
Performance evaluation of a glucose monitoring system for point-of-care testing with the critically ill patient population.
Point of Care
2015
;
14
:
37
41
14.
Mitsios
JV
,
Ashby
LA
,
Haverstick
DM
,
Bruns
DE
,
Scott
MG
:
Analytic evaluation of a new glucose meter system in 15 different critical care settings.
J Diabetes Sci Technol
2013
;
7
:
1282
7
15.
Dungan
K
,
Chapman
J
,
Braithwaite
SS
,
Buse
J
:
Glucose measurement: Confounding issues in setting targets for inpatient management.
Diabetes Care
2007
;
30
:
403
9
16.
Scott
M
,
Bruns
D
,
Boyd
J
,
Sacks
D
:
Tight glucose control in the intensive care unit: Are glucose meters up to the task?
Clin Chem
2008
;
55
:
18
20
17.
Ellis
MF
,
Benjamin
K
,
Cornell
M
,
Decker
K
,
Farrell
D
,
McGugan
L
,
Porter
GP
,
Shearin
H
,
Zhao
Y
,
Granger
BB
:
Suitability of capillary blood glucose analysis in patients receiving vasopressors.
Am J Crit Care
2013
;
22
:
423
9
18.
Sylvain
HF
,
Pokorny
ME
,
English
SM
,
Benson
NH
,
Whitley
TW
,
Ferenczy
CJ
,
Harrison
JG
:
Accuracy of fingerstick glucose values in shock patients.
Am J Crit Care
1995
;
4
:
44
8
19.
Desachy
A
,
Vuagnat
AC
,
Ghazali
AD
,
Baudin
OT
,
Longuet
OH
,
Calvat
SN
,
Gissot
V
:
Accuracy of bedside glucometry in critically ill patients: Influence of clinical characteristics and perfusion index.
Mayo Clin Proc
2008
;
83
:
400
5
20.
Critchell
CD
,
Savarese
V
,
Callahan
A
,
Aboud
C
,
Jabbour
S
,
Marik
P
:
Accuracy of bedside capillary blood glucose measurements in critically ill patients.
Intensive Care Med
2007
;
33
:
2079
84
21.
Rice
MJ
,
Pitkin
AD
,
Coursin
DB
:
Review article: Glucose measurement in the operating room: More complicated than it seems.
Anesth Analg
2010
;
110
:
1056
65
22.
Herasevich
V
,
Kor
DJ
,
Li
M
,
Pickering
BW
:
ICU data mart: a non-iT approach: A team of clinicians, researchers and informatics personnel at the Mayo Clinic have taken a homegrown approach to building an ICU data mart.
Healthc Inform
2011
;
28
:
42, 44
5
23.
Holtby
H
,
Skowno
JJ
,
Kor
DJ
,
Flick
RP
,
Uezono
S
:
New technologies in pediatric anesthesia.
Paediatr Anaesth
2012
;
22
:
952
61
24.
Parkes
JL
,
Slatin
SL
,
Pardo
S
,
Ginsberg
BH
:
A new consensus error grid to evaluate the clinical significance of inaccuracies in the measurement of blood glucose.
Diabetes Care
2000
;
23
:
1143
8
25.
International Organization of Standardization
:
In vitro diagnostic test systems: Requirements for blood glucose monitoring systems for self-testing in managing diabetes mellitus.
Report No. ISO 15197:2003, 2003
26.
CLSI
:
Point of care blood glucose testing in acute and chronic care facilities:
in
Approved guideline, 3rd edition, CLSI Document POCT12-A3
.
Wayne, PA
,
Clinical and Laboratory Standards Institute
,
2013
27.
Karon
BS
,
Gandhi
GY
,
Nuttall
GA
,
Bryant
SC
,
Schaff
HV
,
McMahon
MM
,
Santrach
PJ
:
Accuracy of roche accu-chek inform whole blood capillary, arterial, and venous glucose values in patients receiving intensive intravenous insulin therapy after cardiac surgery.
Am J Clin Pathol
2007
;
127
:
919
26
28.
Charlson
M
,
Szatrowski
TP
,
Peterson
J
,
Gold
J
:
Validation of a combined comorbidity index.
J Clin Epidemiol
1994
;
47
:
1245
51
29.
Tang
Z
,
Du
X
,
Louie
RF
,
Kost
GJ
:
Effects of drugs on glucose measurements with handheld glucose meters and a portable glucose analyzer.
Am J Clin Pathol
2000
;
113
:
75
86
30.
Tang
Z
,
Louie
RF
,
Lee
JH
,
Lee
DM
,
Miller
EE
,
Kost
GJ
:
Oxygen effects on glucose meter measurements with glucose dehydrogenase- and oxidase-based test strips for point-of-care testing.
Crit Care Med
2001
;
29
:
1062
70
31.
Watkinson
PJ
,
Barber
VS
,
Amira
E
,
James
T
,
Taylor
R
,
Young
JD
:
The effects of precision, haematocrit, pH and oxygen tension on point-of-care glucose measurement in critically ill patients: a prospective study.
Ann Clin Biochem
2012
;
49
(
pt 2
):
144
51
32.
Petersen
JR
,
Graves
DF
,
Tacker
DH
,
Okorodudu
AO
,
Mohammad
AA
,
Cardenas
VJ
Jr
:
Comparison of POCT and central laboratory blood glucose results using arterial, capillary, and venous samples from MICU patients on a tight glycemic protocol.
Clin Chim Acta
2008
;
396
:
10
3
33.
Mraovic
B
,
Schwenk
ES
,
Epstein
RH
:
Intraoperative accuracy of a point-of-care glucose meter compared with simultaneous central laboratory measurements.
J Diabetes Sci Technol
2012
;
6
:
541
6
34.
Wiener
RS
,
Wiener
DC
,
Larson
RJ
:
Benefits and risks of tight glucose control in critically ill adults: A meta-analysis.
JAMA
2008
;
300
:
933
44
35.
Boyd
JC
,
Bruns
DE
:
Monte Carlo simulation in establishing analytical quality requirements for clinical laboratory tests meeting clinical needs.
Methods Enzymol
2009
;
467
:
411
33
36.
Karon
BS
,
Boyd
JC
,
Klee
GG
:
Glucose meter performance criteria for tight glycemic control estimated by simulation modeling.
Clin Chem
2010
;
56
:
1091
7
37.
Karon
BS
,
Boyd
JC
,
Klee
GG
:
Empiric validation of simulation models for estimating glucose meter performance criteria for moderate levels of glycemic control.
Diabetes Technol Ther
2013
;
15
:
996
1003
38.
Boyd
JC
,
Bruns
DE
:
Effects of measurement frequency on analytical quality required for glucose measurements in intensive care units: Assessments by simulation models.
Clin Chem
2014
;
60
:
644
50
39.
Karon
BS
,
Meeusen
JW
,
Bryant
SC
:
Impact of glucose meter error on glycemic variability and time in target range during glycemic control after cardiovascular surgery.
J Diabetes Sci Technol
2015
;
10
:
336
42